Performance analysis of a film dosimetric quality assurance procedure for IMRT with regard to the employment of quantitative evaluation methods

被引:27
|
作者
Winkler, P
Zurl, B
Guss, H
Kindl, P
Stuecklschweiger, G
机构
[1] Univ Med Sch, Dept Radiat Oncol, A-8036 Graz, Austria
[2] Graz Univ Technol, Inst Mat Phys, A-8010 Graz, Austria
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 04期
关键词
D O I
10.1088/0031-9155/50/4/006
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A system for dosimetric verification of intensity-modulated radiotherapy (IMRT) treatment plans using absolute calibrated radiographic films is presented. At our institution this verification procedure is performed for all IMRT treatment plans prior to patient irradiation. Therefore clinical treatment plans are transferred to a phantom and recalculated. Composite treatment plans are irradiated to a single film. Film density to absolute dose conversion is performed automatically based on a single calibration film. A software application encompassing film calibration, 2D registration of measurement and calculated distributions, image fusion, and a number of visual and quantitative evaluation utilities was developed. The main topic of this paper is a performance analysis for this quality assurance procedure, with regard to the specification of tolerance levels for quantitative evaluations. Spatial and dosimetric precision and accuracy were determined for the entire procedure, comprising all possible sources of error. The overall dosimetric and spatial measurement uncertainties obtained thereby were 1.9% and 0.8 mm respectively. Based on these results, we specified 5% dose difference and 3 mm distance-to-agreement as our tolerance levels for patient-specific quality assurance for IMRT treatments.
引用
收藏
页码:643 / 654
页数:12
相关论文
共 50 条
  • [1] A novel method for IMRT quality assurance: Quantitative biological evaluation of IMRT film dosimetry
    Gibbons, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1265 - 1265
  • [2] IMRT quality assurance: Dosimetric assessment of three current methods
    Hauger, J.
    Beach, M.
    Diez, D.
    Duggan, D.
    Durant, J.
    Ding, G.
    Coffey, C.
    MEDICAL PHYSICS, 2006, 33 (06) : 1985 - 1985
  • [3] Dosimetric Evaluation of the Compass Program for Patient Dose Analysis in IMRT Delivery Quality Assurance
    Song, J.
    MEDICAL PHYSICS, 2019, 46 (06) : E607 - E607
  • [4] Dosimetric evaluation of the compass program for patient dose analysis in IMRT delivery quality assurance
    Song, Ju-Young
    Ahn, Sung-Ja
    PLOS ONE, 2018, 13 (12):
  • [5] Dosimetric Comparison of Radiochromic Film and Diode Array Measurements for Quality Assurance of Small IMRT Fields
    Ivanova, T.
    Kassaee, A.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [6] Evaluation and Implementation of An IMRT Quality Assurance Procedure to Include Patient-Specific Volumetric Dose Analysis
    Mathews, B.
    Grant, E.
    Maricle, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Dosimetric evaluation of MapCHECK 2 and 3DVH in the IMRT delivery quality assurance process
    Song, Ju-Young
    Kim, Yong-Hyeob
    Jeong, Jae-Uk
    Yoon, Mee Sun
    Ahn, Sung-Ja
    Chung, Woong-Ki
    Nam, Taek-Keun
    MEDICAL DOSIMETRY, 2014, 39 (02) : 134 - 138
  • [8] Patient Specific Quality Assurance of IMRT: Quantitative Approach Using Film Dosimetry and Optimization
    Shin, Kyung Hwan
    Park, Sung-Yong
    Park, Dong Hyun
    Shin, Dongho
    Park, Dahl
    Kim, Tae Hyun
    Pyo, Hongryull
    Kim, Joo-Young
    Kim, Dae Yong
    Cho, Kwan Ho
    Huh, Sun Nyung
    Kim, Il Han
    Park, Charn Il
    RADIATION ONCOLOGY JOURNAL, 2005, 23 (03): : 176 - 185
  • [9] Analysis of the sources of uncertainty for EDR2 film-based IMRT quality assurance
    Shi, Chengyu
    Papanikolaou, Nikos
    Yan, Yulong
    Weng, Xuejun
    Jiang, Hongyu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (02): : 1 - 8
  • [10] Quality assurance in IMRT: film-dosimetry evaluation of transmission under two different collimation systems
    Fiandra, C
    Rizzi, E
    Ragona, R
    Anglesio, S
    Giglioli, FR
    Ricardi, U
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S105 - S105