Posterior tilting of the tibial component decreases femoral rollback in posterior-substituting knee replacement: A computer simulation study

被引:57
|
作者
Piazza, SJ
Delp, SL
Stulberg, SD
Stern, SH
机构
[1] Rehabil Inst Chicago, Sensory Motor Performance Program, Chicago, IL 60611 USA
[2] Northwestern Univ, Dept Mech Engn, Chicago, IL 60611 USA
[3] Northwestern Univ, Dept Biomed Engn, Chicago, IL 60611 USA
[4] Northwestern Univ, Dept Phys Med & Rehabil, Chicago, IL 60611 USA
[5] NW Mem Hosp, Dept Orthopaed, Chicago, IL 60611 USA
关键词
D O I
10.1002/jor.1100160214
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Posterior tilting of the tibial component is thought to increase the range of motion in posterior cruciate-retaining total knee replacement, but its effect on implant motion in posterior cruciate-substituting total knee replacement is unknown. This issue has become of interest recently because manufacturers have introduced instrumentation that produces a posteriorly tilted tibial cut for both implant types. The purpose of this study was to investigate how motion of posterior cruciate-substituting total knee replacement is affected when the tibial component is installed with posterior tilt. Sagittal plane implant motions were predicted from prosthesis geometry with use of a computer simulation in which the femoral condyles were assumed to sit in the bottoms of the tibial condylar wells when the knee was in extension. Rollback of the femoral component was produced by a cam-spine mechanism at higher angles of flexion. The simulations revealed that even small degrees of posterior tilt reduced rollback by limiting the interaction between the cam and spine. Tilting the component posteriorly by 5 degrees caused the cam to contact the spine at a knee flexion angle that was 18 degrees higher than with the untilted component. The results suggest that posterior tilting of the tibial component in posterior cruciate-substituting knee replacement may not produce the same beneficial effects that have been reported for the tilting of tibial components in posterior cruciate-retaining knee replacement.
引用
收藏
页码:264 / 270
页数:7
相关论文
共 30 条
  • [1] Biomechanics of posterior-substituting total knee arthroplasty: An in vitro study
    Li, GA
    Otterberg, E
    Sabbag, K
    Zayontz, S
    Johnson, T
    Rubash, H
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (404) : 214 - 225
  • [2] Interactive effect of femoral posterior condylar offset and tibial posterior slope on knee flexion in posterior cruciate ligament-substituting total knee arthroplasty
    Han, Hyuk-Soo
    Kang, Seung-Baik
    KNEE, 2018, 25 (02): : 335 - 340
  • [3] Optimisation of the posterior stabilised tibial post for greater femoral rollback after total knee arthroplasty—a finite element analysis
    Nagarajan Chandran
    Farid Amirouche
    Mark H. Gonzalez
    Kevin M. Hilton
    Riad Barmada
    Wayne Goldstein
    International Orthopaedics, 2009, 33 : 687 - 693
  • [4] The effect of the posterior slope of the tibial plateau osteotomy with a rotational error on tibial component malalignment in total knee replacement
    Tsukeoka, T.
    Tsuneizumi, Y.
    Lee, T. H.
    BONE & JOINT JOURNAL, 2013, 95B (09): : 1201 - 1203
  • [5] Optimisation of the posterior stabilised tibial post for greater femoral rollback after total knee arthroplasty-a finite element analysis
    Chandran, Nagarajan
    Amirouche, Farid
    Gonzalez, Mark H.
    Hilton, Kevin M.
    Barmada, Riad
    Goldstein, Wayne
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (03) : 687 - 693
  • [6] Fracture of femoral component in an Oxford® unicompartmental knee replacement with displaced posterior fragment: a case report
    Kiliyanpilakkil, V. B.
    Subramanian, S.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2020, 102 (07) : 148 - 151
  • [7] Contact Kinematics Correlates to Tibial Component Migration Following Single Radius Posterior Stabilized Knee Replacement
    Teeter, Matthew G.
    Perry, Kevin I.
    Yuan, Xunhua
    Howard, James L.
    Lanting, Brent A.
    JOURNAL OF ARTHROPLASTY, 2018, 33 (03): : 740 - 745
  • [8] Inserting the best knee: A short-term analysis of posterior stabilized versus cruciate substituting ultracongruent tibial inserts for total knee replacement
    V. Rajguhan, T.
    Vanchi, Pravin K.
    Prashanth, K. R. Tarun
    Veeraraghavan, Raghav R.
    Murugesan, Mohan K.
    JOURNAL OF MUSCULOSKELETAL SURGERY AND RESEARCH, 2023, 7 (03) : 180 - 185
  • [9] High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement
    Han, H. S.
    Kang, S. -B.
    Yoon, K. S.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (11): : 1457 - 1461
  • [10] A computer simulation study for preserving the tibial posterior slope in open-wedge high tibial osteotomy
    Mitsuru Hanada
    Kensuke Hotta
    Yukihiro Matsuyama
    European Journal of Orthopaedic Surgery & Traumatology, 2020, 30 : 1285 - 1291