Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis comparison with echocardiography

被引:185
|
作者
Caruthers, SD
Lin, SJ
Brown, P
Watkins, MP
Williams, TA
Lehr, KA
Wickline, SA
机构
[1] Washington Univ, Sch Med, Barnes Jewish Hosp, Cardiovasc Magnet Resonance Labs, St Louis, MO 63110 USA
[2] Washington Univ, Dept Biomed Engn, St Louis, MO 63110 USA
[3] Philips Med Syst, Best, Netherlands
关键词
magnetic resonance imaging; aorta; stenosis; echocardiography; valves;
D O I
10.1161/01.CIR.0000095268.47282.A1
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-Valvular pathology can be analyzed quickly and accurately through the use of Doppler ultrasound. For aortic stenosis, the continuity equation approach with Doppler velocity-time integral (VTI) data is by far the most commonly used clinical method of quantification. In view of the emerging popularity of cardiac magnetic resonance (CMR) as a routine clinical imaging tool, the purposes of this study were to define the reliability of velocity-encoded CMR as a routine method for quantifying stenotic aortic valve area, to compare this method with the accepted standard, and to evaluate its reproducibility. Methods and Results-Patients (n=24) with aortic stenosis (ranging from 0.5 to 1.8 cm(2)) were imaged with CMR and echocardiography. Velocity-encoded CMR was used to obtain velocity information in the aorta and left ventricular outflow tract. From this flow data, pressure gradients were estimated by means of the modified Bernoulli equation, and VTIs were calculated to estimate aortic valve orifice dimensions by means of the continuity equation. The correlation coefficients between modalities for pressure gradients were r=0.83 for peak and r=0.87 for mean. The measurements of VTI correlated well, leading to an overall strong correlation between modalities for the estimation of valve dimension (r=0.83, by means of the identified best approach). For 5 patients, the CMR examination was repeated using the best approach. The repeat calculations of valve size correlated well (r=0.94). Conclusions-Velocity-encoded CMR can be used as a reliable, user-friendly tool to evaluate stenotic aortic valves. The measurements of pressure gradients, VTIs, and the valve dimension correlate well with the accepted standard of Doppler ultrasound.
引用
收藏
页码:2236 / 2243
页数:8
相关论文
共 50 条
  • [1] Quantification of aortic valve stenosis by a rapid cardiac magnetic resonance protocol: Comparison with echocardiography
    Anand, D. V.
    Burchell, T. R.
    Mohiddin, S. A.
    Davies, L. C.
    Petersen, S. E.
    Westwood, M. A.
    EUROPEAN HEART JOURNAL, 2010, 31 : 143 - 143
  • [2] A hybrid approach for quantification of aortic valve stenosis using cardiac magnetic resonance imaging and echocardiography
    Haghi, D
    Papavassiliu, T
    Kalmar, G
    Schroder, M
    Neff, W
    Kaden, JJ
    Muller, U
    Haase, KK
    Borggrefe, M
    Suselbeck, T
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2005, 7 (03) : 581 - 586
  • [3] Quantification of Congenital Aortic Valve Stenosis in Pediatric Patients: Comparison Between Cardiac Magnetic Resonance Imaging and Transthoracic Echocardiography
    Sirin, Selma
    Nassenstein, Kai
    Neudorf, Ulrich
    Jensen, Christoph J.
    Mikat, Christian
    Schlosser, Thomas
    PEDIATRIC CARDIOLOGY, 2014, 35 (05) : 771 - 777
  • [4] Quantification of Congenital Aortic Valve Stenosis in Pediatric Patients: Comparison Between Cardiac Magnetic Resonance Imaging and Transthoracic Echocardiography
    Selma Sirin
    Kai Nassenstein
    Ulrich Neudorf
    Christoph J. Jensen
    Christian Mikat
    Thomas Schlosser
    Pediatric Cardiology, 2014, 35 : 771 - 777
  • [5] Is cardiac magnetic resonance imaging as accurate as echocardiography in the assessment of aortic valve stenosis?
    Wong, Sophia
    Spina, Roberto
    Toemoe, Sianne
    Dhital, Kumud
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2016, 22 (04) : 480 - 486
  • [6] A hybrid approach for quantification of aortic valve stenosis using cardiac magnetic resonance imaging and echocardiography: Comparison to right heart catheterization and standard echocardiography
    Haghi, D
    Suselbeck, T
    Fluechter, S
    Kalmar, G
    Schroder, M
    Kaden, JJ
    Poerner, T
    Borggrefe, M
    Papavassiliu, T
    CLINICAL RESEARCH IN CARDIOLOGY, 2006, 95 (03) : 162 - 167
  • [7] A hybrid approach for quantification of aortic valve stenosis using cardiac magnetic resonance imaging and echocardiography:Comparison to right heart catheterization and standard echocardiography
    Dariusch Haghi
    T. Suselbeck
    S. Fluechter
    G. Kalmar
    M. Schroder
    J. J. Kaden
    T. Poerner
    M. Borggrefe
    T. Papavassiliu
    Clinical Research in Cardiology, 2006, 95 : 162 - 167
  • [8] A hybrid approach for quantification of aortic valve stenosis using cardiac magnetic resonance imaging and echocardiography: comparison to right heart catheterisation and standard echocardiography
    Haghi, D.
    Suselbeck, T.
    Fluechter, S.
    Kalmar, G.
    Kaden, J. J.
    Poerner, T.
    Borggrefe, M.
    Papavassiliu, T.
    EUROPEAN HEART JOURNAL, 2005, 26 : 298 - 298
  • [9] Evaluation of aortic stenosis - A comparison of cardiac catheterisation, magnetic resonance imaging and echocardiography
    Kupfahl, C
    Mahrholdt, H
    Honold, M
    Vogelsberg, H
    Meinhardt, G
    Wagner, A
    Sechtem, U
    CIRCULATION, 2003, 108 (17) : 431 - 431
  • [10] Assessment of aortic valve area in aortic stenosis using cardiac magnetic resonance tomography: Comparison with echocardiography
    Malyar, Nasser M.
    Schlosser, Thomas
    Barkhausen, Joerg
    Gutersohn, Achim
    Buck, Thomas
    Bartel, Thomas
    Erbel, Raimund
    CARDIOLOGY, 2008, 109 (02) : 126 - 134