Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courier

被引:17
|
作者
Peter, John Victor [2 ]
Patole, Shalom [3 ]
Fleming, Jude Joseph [4 ]
Selvakumar, Ratnasamy [4 ]
Graham, Petra L. [1 ]
机构
[1] Macquarie Univ, Dept Stat, Sydney, NSW 2109, Australia
[2] Christian Med Coll & Hosp, Med ICU, Vellore 632004, Tamil Nadu, India
[3] Christian Med Coll & Hosp, Dept Med, Vellore 632004, Tamil Nadu, India
[4] Christian Med Coll & Hosp, Dept Clin Biochem, Vellore 632004, Tamil Nadu, India
关键词
agreement; arterial blood gas; Bland-Altman; concordance; pneumatic tube system; point of care testing; TUBE SYSTEM; AIR BUBBLES; OXYGEN-TENSION; TIME; TEMPERATURE; SPECIMENS;
D O I
10.1515/CCLM.2011.611
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Rapid accurate assessment of metabolic derangements is crucial in the critically ill. We evaluated if arterial blood gas (ABG) samples transported through a pneumatic tube system (PTS) agreed with values transported by a human courier. Methods: In this prospective study of 50-paired ABG samples, the couriered reference ABG was compared with those transported by PTS. Agreement was summarised by the mean difference with 95% limits of agreement (LOA) and Lin's concordance correlation (pc). Results: The mean (+/- SD) time from sampling to analysis was 35.7 +/- 23.2 (courier) and 38.6 +/- 22.1 (PTS) minutes. Agreement was good between courier and PTS for pH, PaCO(2), bicarbonate, oxygen saturation and PaO(2) values (pc > 0.97). Although the mean difference in PaO(2) values between PTS and courier was small (-0.9 mm Hg) and the agreement was good, individual differences were clinically significant (95% LOA -40.8 to 39.0). For PaO(2) < 160 mm Hg, analysis of PTS samples yielded erroneously high PaO(2) values and vice versa for PaO(2) > 160 mm Hg compared to manual courier. This suggested exaggerated oxygen movement between the blood sample and air in the PTS. Conclusions: In this study, analysis of samples transported through the PTS resulted in clinically unacceptable PaO(2) values. Delay in transport and analysis of ABG samples should be avoided and samples transported manually if they cannot be assessed on-site.
引用
收藏
页码:1303 / 1309
页数:7
相关论文
共 16 条
  • [1] IMMUNOLOGICAL STUDIES OF BLOOD-SAMPLES TRANSPORTED BY A PNEUMATIC TUBE SYSTEM
    GREENDYK, RM
    BANZHAF, JC
    PELYSKO, S
    BAUMAN, B
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1977, 68 (04) : 508 - 510
  • [2] Transport of blood gas samples - Is the pneumatic tube system safe?
    Zanner, R.
    Moser, N.
    Blobner, M.
    Luppa, P. B.
    ANAESTHESIST, 2006, 55 (10): : 1099 - 1104
  • [3] Changes in blood gas samples produced by a pneumatic tube system
    Collinson, PO
    John, CM
    Gaze, DC
    Ferrigan, LF
    Cramp, DG
    JOURNAL OF CLINICAL PATHOLOGY, 2002, 55 (02) : 105 - 107
  • [4] The comparison of blood gas analysis (BGA) parameters in samples transported by the pneumatic tube system and manual transport at Dr. Soetomo Hospital, Surabaya, Indonesia
    Gunawan, Hantoro
    Marpaung, Ferdy Royland
    BALI MEDICAL JOURNAL, 2019, 8 (02) : 400 - 403
  • [5] Delivery of blood samples in an airtight carrier through a pneumatic tube system is acceptable for blood gas analysis.
    Yu, Y
    Liu, M
    Yang, Y
    Lin, J
    Wang, B
    CLINICAL CHEMISTRY, 2004, 50 (06) : A173 - A173
  • [6] PNEUMATIC TUBE SYSTEM EXACERBATES THE EFFECT OF ROOM AIR CONTAMINATION IN BLOOD-GAS SAMPLES
    ASTLES, R
    LUBARSKY, D
    LOUN, B
    SEDOR, F
    TOFFALETTI, J
    CLINICAL CHEMISTRY, 1995, 41 (06) : S155 - S155
  • [7] Blood gas analysis: POCT versus central laboratory on samples sent by a pneumatic tube system
    Zaman, Z
    Demedts, M
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 101 - 106
  • [8] Impact of Pneumatic Tube System Transport of Whole Blood Samples on Free DNA Concentration in Human Plasma
    Kriegshaeuser, Gernot
    Wagner, Carina
    Mangge, Harald
    Halwachs-Baumann, Gabriele
    Enko, Dietmar
    CLINICAL CHEMISTRY, 2016, 62 (12) : 1672 - 1673
  • [9] Agreement between arterial and non-arterialised fingertip capillary blood gas and acid-base values
    Kongstad, Heidi K.
    Rosendal, Christian A. H.
    Rasmussen, Bodil S.
    Weinreich, Ulla M.
    EUROPEAN CLINICAL RESPIRATORY JOURNAL, 2019, 6 (01):
  • [10] Profiles of neonicotinoid insecticides and characteristic metabolites in paired urine and blood samples: Partitioning between urine and blood and implications for human exposure
    Xu, Miaomiao
    Zhang, Zhanpeng
    Li, Zhiyong
    Kan, Shunyan
    Liu, Zhaoxiang
    Wang, Desheng
    Liu, Qihui
    Zhang, Hua
    SCIENCE OF THE TOTAL ENVIRONMENT, 2021, 773