Statistical process control for radiotherapy quality assurance

被引:108
|
作者
Pawlicki, T [1 ]
Whitaker, M
Boyer, AL
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
[2] Radiol Imaging Technol Inc, Colorado Springs, CO 80907 USA
关键词
quality assurance; statistical process control; process behavior charts;
D O I
10.1118/1.2001209
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Every quality assurance process uncovers random and systematic errors. These errors typically consist of many small random errors and a very few number of large errors that dominate the result. Quality assurance practices in radiotherapy do not adequately differentiate between these two sources of error. The ability to separate these types of errors would allow the dominant source(s) of error to be efficiently detected and addressed. In this work, statistical process control is applied to quality assurance in radiotherapy for the purpose of setting action thresholds that differentiate between random and systematic errors. The theoretical development and implementation of process behavior charts are described. We report on a pilot project is which these techniques are applied to daily output and flatness/symmetry quality assurance for a 10 MV photon beam in our department. This clinical case was followed over 52 days. As part of our investigation, we found that action thresholds set using process behavior charts were able to identify systematic changes in our daily quality assurance process. This is in contrast to action thresholds set using the standard deviation, which did not identify the same systematic changes in the process. The process behavior thresholds calculated from a subset of the data detected a 2% change in the process whereas with a standard deviation calculation, no change was detected. Medical physicists must make decisions on quality assurance data as it is acquired. Process behavior charts help decide when to take action and when to acquire more data before making a change in the process. (C) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:2777 / 2786
页数:10
相关论文
共 50 条
  • [1] Radiotherapy Quality Assurance Using Statistical Process Control
    Binny, Diana
    Lancaster, Craig M.
    Kairn, Tanya
    Trapp, Jamie V.
    Crowe, Scott B.
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2018, VOL 3, 2019, 68 (03): : 437 - 442
  • [2] Statistical Process Control in Monitoring Radiotherapy Quality Assurance Program: An Institutional Experience
    Vysakh, R.
    Ganapathi Raman, R.
    Niyas, P.
    Aflah, P.
    [J]. Iranian Journal of Medical Physics, 2022, 19 (03) : 189 - 198
  • [3] Feasibility study of using statistical process control to optimize quality assurance in radiotherapy
    Gerard, Karine
    Grandhaye, Jean-Pierre
    Marchesi, Vincent
    Aletti, Pierre
    Husson, Francois
    Noel, Alain
    Kafrouni, Hanna
    [J]. JOURNAL OF QUALITY IN MAINTENANCE ENGINEERING, 2009, 15 (04) : 331 - +
  • [4] Moving towards process-based radiotherapy quality assurance using statistical process control
    Raveendran, Vysakh
    Raman, R. Ganapathi
    Anjana, P. T.
    Bhasi, Saju
    Ranjith, C. P.
    Kinhikar, Rajesh Ashok
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 112
  • [5] Statistical process control for IMRT quality assurance
    Pawlicki, T
    Whitaker, M
    Price, RA
    Boyer, AL
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1351 - 1351
  • [6] Statistical process control for quality assurance of patient positioning during head-and-neck radiotherapy
    Louwe, R.
    Moore, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S163 - S164
  • [7] Statistical process control (SPC) to optimize quality assurance in intensity-modulated radiotherapy (IMRT)
    Gerard, K.
    Marchesi, V.
    Grandhaye, JP.
    Franceries, X.
    Husson, F.
    Noel, A.
    Kafrouni, H.
    Aletti, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S97 - S97
  • [8] Application of statistical process control to IMRT quality assurance
    Fang, B.
    Shoale, J.
    Pawlicki, T.
    Boyer, A.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2071 - 2071
  • [9] ONLINE QUALITY ASSURANCE BY STATISTICAL PROCESS-CONTROL
    MAETKE, H
    RUHLE, W
    [J]. TECHNISCHES MESSEN, 1989, 56 (09): : 332 - 340
  • [10] Statistical Process Control for EPID Dosimetry in the Quality Assurance of IMRT
    Matsumoto, K.
    Okumura, M.
    Asai, Y.
    Shimomura, K.
    Tamura, M.
    Nishimura, Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S867 - S867