Simple modification of arm position improves B1+ and signal homogeneity in the thoracolumbar spine at 3T

被引:2
|
作者
Ishizaka, Kinya [1 ]
Kudo, Kohsuke [2 ]
Harada, Kuniaki [3 ]
Shirai, Toru [4 ]
Fujiwara, Taro [1 ]
Aoike, Suzuko [1 ]
Takamori, Sayaka [1 ]
Shirato, Hiroki [5 ,6 ]
机构
[1] Hokkaido Univ Hosp, Dept Radiol Technol, Sapporo, Hokkaido, Japan
[2] Hokkaido Univ Hosp, Div Diagnost & Intervent Radiol, Sapporo, Hokkaido, Japan
[3] Hitachi Ltd, Healthcare Business Unit, Tokyo, Japan
[4] Hitachi Ltd, Res & Dev Grp, Tokyo, Japan
[5] Hokkaido Univ, Grad Sch Med, Dept Radiat Med, Sapporo, Hokkaido, Japan
[6] Hokkaido Univ, Global Stn Quantum Med Sci & Engn, Global Inst Collaborat Res & Educ, Sapporo, Hokkaido, Japan
关键词
thoracolumbar spine MRI; arms position; B-1(+) homogeneity; 3.0; T; MRI; EXCITATION; FIELD; ARTIFACTS; ARRAY; PADS; CORD;
D O I
10.1002/jmri.25767
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo evaluate the homogeneity of the radiofrequency magnetic field (B-1(+)) and signal intensity using different arm positions during 3T thoracolumbar spinal imaging. Materials and MethodsTwenty volunteers were scanned with a four-channel radiofrequency (RF) transmit coil at 3T, with arms on the bed (conventional), arms elevated by 100 mm (arm lift), or with the arms-up position (elevated arm). Axial B-1(+) maps and sagittal T-1-weighted image (T1WI)-performed RF shimming were obtained for each arm position. The mean and standard deviation (SD) of the flip angle (FA) at the center of the vertebra on each B-1(+) map, and contrast noise ratios (CNRs) between the spinal cord and cerebrospinal fluid of sagittal T1WI, were calculated and compared among the different arm positions. ResultsMean FA values (degrees) for the arm lift and elevated arm positions were significantly larger than for the conventional position (P < 0.001 for both) at the twelfth thoracic vertebra (Th12). FA SD values for the arm lift and elevated arm position were significantly smaller than for the conventional position (P < 0.001 for both) at Th12. CNR for the arm lift and elevated arm position were significantly higher than for the conventional position (P = 0.007 and 0.002, respectively). The mean and SD of the FA and the CNR did not differ significantly for the arm lift and elevated arm positions (P = 0.591, 0.958, and 0.927, respectively). ConclusionInhomogeneities of B-1(+) and signal intensities were improved by simply changing the arm position in 3T thoracolumbar spinal imaging. Level of Evidence: 3 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2018;47:123-130.
引用
收藏
页码:123 / 130
页数:8
相关论文
共 12 条
  • [1] Optimal flip angles for in vivo liver 3D T1 mapping and B1+ mapping at 3T
    Belsley G.
    Tyler D.J.
    Robson M.D.
    Tunnicliffe E.M.
    Magnetic Resonance in Medicine, 2023, 90 (03) : 950 - 962
  • [2] Accuracy and precision of electrical permittivity mapping at 3T: the impact of three B1+ mapping techniques
    Gavazzi, Soraya
    van den Berg, Cornelis A. T.
    Sbrizzi, Alessandro
    Kok, H. Petra
    Stalpers, Lukas J. A.
    Lagendijk, Jan J. W.
    Crezee, Hans
    van Lier, Astrid L. H. M. W.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (06) : 3628 - 3642
  • [3] In vivo electric conductivity of cervical cancer patients based on B1+ maps at 3T MRI
    Balidemaj, E.
    de Boer, P.
    van Lier, A. L. H. M. W.
    Remis, R. F.
    Stalpers, L. J. A.
    Westerveld, G. H.
    Nederveen, A. J.
    van den Berg, C. A. T.
    Crezee, J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (04): : 1596 - 1607
  • [4] B1+ compensation in 3T cardiac Imaging using short 2DRF pulses
    Sung, Kyunghyun
    Nayak, Krishna S.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 59 (03) : 441 - 446
  • [5] Motion-robust cardiac B1+ mapping at 3T using interleaved bloch-siegert shifts
    Weingartner, Sebastian
    Zimmer, Fabian
    Metzger, Gregory J.
    Ugurbil, Kamil
    Van de Moortele, Pierre-Francois
    Akcakaya, Mehmet
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (02) : 670 - 677
  • [6] Optimization and Evaluation of Reference Region Variable Flip Angle (RR-VFA) B1+ and T1 Mapping in the Prostate at 3T
    Rangwala, Novena A.
    Dregely, Isabel
    Wu, Holden H.
    Sung, Kyunghyun
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 45 (03) : 751 - 760
  • [7] Simulations of a birdcage coil B1+ field on a human body model for designing a 3T multichannel TMS/MRI head coil array
    de Lara, Lucia I. Navarro
    Rad, Laleh Golestani
    Makarov, Sergey N.
    Stockmann, Jason
    Wald, Lawrence L.
    Nummenmaa, Aapo
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 4752 - 4755
  • [8] Improving phase-based conductivity reconstruction by means of deep learning based denoising of B1+ phase data for 3T MRI
    Jung, Kyu-Jin
    Mandija, Stefano
    Kim, Jun-Hyeong
    Ryu, Kanghyun
    Jung, Soozy
    Cui, Chuanjiang
    Kim, Soo-Yeon
    Park, Mina
    van den Berg, Cornelis A. T.
    Kim, Dong-Hyun
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (04) : 2084 - 2094
  • [9] Improvement in B1+-homogeneity of 3T cardiac MRI in swine with dual-source parallel RF excitation
    D Herzka
    H Ding
    M Schär
    Journal of Cardiovascular Magnetic Resonance, 15 (Suppl 1)
  • [10] Evaluation of a Subject specific dual-transmit approach for improving B1 field homogeneity in cardiovascular magnetic resonance at 3T
    Krishnamurthy, Ramkumar
    Pednekar, Amol
    Kouwenhoven, Marc
    Cheong, Benjamin
    Muthupillai, Raja
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2013, 15