Severe wear and fracture of zirconia heads against alumina inserts in hip simulator studies with microseparation

被引:57
|
作者
Stewart, TD
Tipper, JL
Insley, G
Streicher, RM
Ingham, E
Fisher, J
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Div Microbiol, Leeds LS2 9JT, W Yorkshire, England
[3] Stryker SA, Sci & Clin Affairs, Thalwil, Switzerland
来源
JOURNAL OF ARTHROPLASTY | 2003年 / 18卷 / 06期
基金
英国工程与自然科学研究理事会;
关键词
hip prostheses; zirconia; alumina; wear; microseparation; simulator;
D O I
10.1016/S0883-5403(03)00204-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The wear of zirconia femoral heads against alumina acetabular inserts with swing-phase microseparation was investigated in a hip joint simulator. Under mild microseparation conditions, the wear was very low, with an average wear rate of 0.05 mm(3)/million cycles reported over 5 million cycles of testing. However, under severe microseparation conditions representative of greater joint laxity, the wear rate of zirconia against alumina increased by 2 orders of magnitude, producing severe wear and, in one case, femoral head fracture. The adverse results of this study indicate that the combination of a zirconia femoral head articulating against an alumina acetabular insert is not recommended for clinical use. The results further raise concerns over the suitability of conventional simulators in evaluating the wear of ceramic hip prostheses. (C) 2003 Elsevier Inc. All rights reserved.
引用
收藏
页码:726 / 734
页数:9
相关论文
共 50 条
  • [1] Polyethylene wear against alumina and zirconia heads in cemented total hip arthroplasty
    Liang, Bojian
    Kawanabe, Keiichi
    Ise, Kentaro
    Iida, Hirokazu
    Nakamura, Takashi
    [J]. JOURNAL OF ARTHROPLASTY, 2007, 22 (02): : 251 - 257
  • [2] Comparison of polyethylene wear against alumina and zirconia heads in cemented total hip arthroplasty
    Kawanabe, K.
    Liang, B.
    Ise, K.
    Nakamura, T.
    [J]. Bioceramics and Alternative Bearings in Joint Arthroplasty, 2007, : 83 - 87
  • [3] Three-body wear of UHMWPE acetabular cups by PMMA particles against CoCr, alumina and zirconia heads in a hip joint simulator
    Wang, A
    Essner, A
    [J]. WEAR, 2001, 250 : 212 - 216
  • [4] Polyethylene wear against alumina and zirconia heads in K-max total hip arthroplasty
    Liang, B
    Kawanabe, K
    Ise, K
    Shimizu, M
    Iida, H
    Nakamura, T
    [J]. BIOCERAMICS, VOL 17, 2005, 284-286 : 971 - 974
  • [5] Microseparation and stripe wear in alumina-on-alumina hip implants
    Affatato, Saverio
    Traina, Francesco
    Toni, Aldo
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2011, 34 (06): : 506 - 512
  • [6] WEAR OF POLYETHYLENE ACETABULAR CUPS AGAINST ZIRCONIA FEMORAL HEADS STUDIED WITH A HIP-JOINT SIMULATOR
    SAIKKO, V
    [J]. WEAR, 1994, 176 (02) : 207 - 212
  • [7] Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns
    Nevelos, J
    Ingham, E
    Doyle, C
    Streicher, R
    Nevelos, A
    Walter, W
    Fisher, J
    [J]. JOURNAL OF ARTHROPLASTY, 2000, 15 (06): : 793 - 795
  • [8] Investigation of the wear properties of ceramic total hip prostheses made of alumina and zirconia combination: A hip simulator study
    Ueno, M
    Ikeuchi, K
    [J]. BIOCERAMICS 15, 2003, 240-2 : 809 - 812
  • [9] Stripe wear rates in alumina THR - Comparison of microseparation simulator study with retrieved implants
    Manaka, M
    Clarke, IC
    Yamamoto, K
    Shishido, T
    Gustafson, A
    Imakiire, A
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2004, 69B (02) : 149 - 157
  • [10] WEAR OF POLYETHYLENE ACETABULAR CUPS AGAINST ALUMINA FEMORAL HEADS - 5 PROSTHESES COMPARED IN A HIP SIMULATOR FOR 35 MILLION WALKING CYCLES
    SAIKKO, VO
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (05): : 507 - 512