Monte Carlo study of the neutron ambient dose equivalent at the heavy ion medical machine in Wuwei

被引:0
|
作者
Huang, Sheng-Cong [1 ,2 ]
Zhang, Hui [1 ,2 ]
Bai, Kai [2 ,3 ]
Luo, Ying [1 ,2 ]
Mao, Hai-Jun [2 ,3 ]
Dai, Zhong-Ying [1 ,2 ]
机构
[1] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
[2] Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
[3] Lanzhou Univ, Sch Nucl Sci & Technol, Lanzhou 730000, Peoples R China
基金
中国国家自然科学基金;
关键词
Carbon-ion radiotherapy; Passive beam delivery; Secondary neutrons; Monte Carlo method; PASSIVE CARBON-ION; SECONDARY NEUTRON; TUMOR-THERAPY; PROTON; RADIOTHERAPY; SIMULATIONS; EXPOSURES;
D O I
10.1007/s41365-022-01093-z
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The use of carbon-ion radiotherapy (CIRT) is gradually increasing. Owing to the generation of high-energy secondary neutrons during CIRT, its use presents new challenges in radiation protection. Thus, secondary neutron dose distributions must be explored and evaluated under clinical scenarios based on different treatment configurations. However, neutron dose and energy spectrum measurements are often difficult. This can be primarily attributed to the inherent limitations of most neutron detectors, such as their unsuitability for spectral measurements and inaccurate responses to neutrons with energies above 20 MeV. Numerical calculation methods based on probabilistic statistical theory are fast and convenient for neutron dose evaluation. In this study, external secondary neutron doses at the heavy ion medical machine in Wuwei, which is equipped with a passive beam delivery system, were calculated using the Monte Carlo method. The dependence of neutron doses on various treatment parameters (incident carbon-ion beam energy, spatial location, field size, and spread-out Bragg peak (SOBP) width) was investigated. Furthermore, the feasibility of applying an analytical model to predict the ambient dose equivalent was verified. For the combination involving an energy of 400 MeV/u and SOBP width of 6 cm, the ambient dose equivalent per therapeutic dose (H/D) at the isocenter was 79.87 mSv/Gy. The H/D value decreased rapidly with increasing spatial distance and slightly with increasing aperture size and SOBP width. The H/D values derived from the Monte Carlo simulations were in good agreement with the results reported in the literature. The analytical model could be used to quickly predict the H/D value along the incidence direction of the beam with an error of less than 20%. Thus, our study contributes to the understanding of the relationship between neutron radiation and treatment configuration parameters, which establishes a basis for predicting non-therapeutic radiation doses in CIRT.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Monte Carlo study of the neutron ambient dose equivalent at the heavy ion medical machine in Wuwei
    Sheng-Cong Huang
    Hui Zhang
    Kai Bai
    Ying Luo
    Hai-Jun Mao
    Zhong-Ying Dai
    [J]. Nuclear Science and Techniques, 2022, 33 (09) : 113 - 127
  • [2] Monte Carlo study of the neutron ambient dose equivalent at the heavy ion medical machine in Wuwei
    Sheng-Cong Huang
    Hui Zhang
    Kai Bai
    Ying Luo
    Hai-Jun Mao
    Zhong-Ying Dai
    [J]. Nuclear Science and Techniques, 2022, 33
  • [3] Monte Carlo study of neutron-ambient dose equivalent to patient in treatment room
    Mohammadi, A.
    Afarideh, H.
    Davani, F. Abbasi
    Ghergherehchi, M.
    Arbabi, A.
    [J]. APPLIED RADIATION AND ISOTOPES, 2016, 118 : 140 - 148
  • [4] Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design
    Titt, Uwe
    Pera, Enzo
    Gillin, Michael T.
    [J]. INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2020, 6 (04) : 29 - 37
  • [5] Monte Carlo Study of Neutron Dose Equivalent for a Compact Proton Therapy Unit
    Zheng, Y.
    Klein, E.
    Chen, K.
    Liu, Y.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [6] The use of a Monte Carlo method for uncertainty calculation, with an application to the measurement of neutron ambient dose equivalent rate
    Cox, Maurice
    Harris, Peter
    Nam, Gyeonghee
    Thomas, David
    [J]. RADIATION PROTECTION DOSIMETRY, 2006, 121 (01) : 12 - 23
  • [7] Spatial fractionation of the dose in heavy ion therapy: a Monte Carlo study
    Peucelle, C.
    Martinez-Rovira, I.
    Prezado, Y.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S37 - S38
  • [8] Monte Carlo study of neutron dose equivalent during passive scattering proton therapy
    Zheng, Yuanshui
    Newhauser, Wayne
    Fontenot, Jonas
    Taddei, Phil
    Mohan, Radhe
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15): : 4481 - 4496
  • [9] Neutron dose equivalent rate for heavy ion bombardment
    Li Gui Sheng
    [J]. Nuclear Science and Techniques, 1998, (03) : 17 - 19
  • [10] Different methods of organ equivalent dose scoring in Monte Carlo neutron dose calculations
    Jarlskog, C. Zacharatou
    Lee, C.
    Bolch, W.
    Xu, X.
    Paganetti, H.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2437 - 2437