Feasibility and safety of magnetic-end double-J ureteral stent insertion and removal in children

被引:10
|
作者
Chalhoub, Marc [1 ,2 ]
Kohaut, Jules [1 ,2 ]
Vinit, Nicolas [1 ,2 ]
Botto, Nathalie [1 ]
Aigrain, Yves [1 ,2 ]
Heloury, Yves [1 ,2 ]
Lottmann, Henri [1 ]
Blanc, Thomas [1 ,2 ,3 ]
机构
[1] Hop Necker Enfants Malad, AP HP, Serv Chirurg Viscerale & Urol Pediat, Paris, France
[2] Univ Paris, Paris, France
[3] Univ Paris 05, Inst Necker Enfants Malad, Dept Croissance & Signalisat, Hop Necker Enfants Malad,INSERM,U1151,CNRS,UMR 82, Paris, France
关键词
Children; Urology; Ureteral stent; Magnetic stent; Removal; EXPERIENCE; CYSTOSCOPY;
D O I
10.1007/s00345-020-03339-0
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose The need for surgical removal of a double-J ureteral stent (DJUS) is considered one of its disadvantages. Apart from increased cost, repeated exposure to general anesthesia is a concern in children. Alternative techniques have been described, all failing to become integrated into mainstream practice. Stents with a distal magnetic end, although introduced in the early 1980s, have only recently gained acceptance. We report the feasibility and safety of insertion and removal of a magnetic-end double-J ureteral stent (MEDJUS) in a pediatric population. Materials and methods We retrospectively analyzed the use of the Magnetic Black-Star Urotech(R)MEDJUS between 11/2016 and 12/2019 in children. Stents were removed in the outpatient clinic using a transurethral catheter with a magnetic tip. Results MEDJUS insertion was attempted in 100 patients (65 boys). Mean age was 7.8 years (0.5-18). The stent was placed in an antegrade procedure (n = 47), by a retrograde route (n = 10), and during open surgery (n = 43). Stent insertion was successful in 84 cases (84%). All 16 failures occurred during the antegrade approach in laparoscopic pyeloplasty, with inability to push the stent and its magnet through the ureterovesical junction in 14. Magnetic removal was attempted in 83 patients, successful in 81 (98%). There was no added morbidity with the MEDJUS. Conclusions The use of MEDJUS is a safe and effective strategy that obviates the need for additional general anesthesia in children. Its insertion is similar to that with regular DJUS, and its easy and less time-consuming removal benefits both the patient and the hospital and validates its clinical use.
引用
下载
收藏
页码:1649 / 1655
页数:7
相关论文
共 50 条
  • [1] Feasibility and safety of magnetic-end double-J ureteral stent insertion and removal in children
    Marc Chalhoub
    Jules Kohaut
    Nicolas Vinit
    Nathalie Botto
    Yves Aigrain
    Yves Héloury
    Henri Lottmann
    Thomas Blanc
    World Journal of Urology, 2021, 39 : 1649 - 1655
  • [2] Strong parental satisfaction with the use of magnetic-end double-j ureteral stent in children
    Glenisson, M.
    Vinit, N.
    Marquant, F.
    Party, P.
    Nefzy, L.
    Broch, A.
    Elie, C.
    Lottmann, H.
    Botto, N.
    Blanc, T.
    EUROPEAN UROLOGY, 2024, 85 : S28 - S28
  • [3] Noncystoscopic Removal of Double-J Stent in Women: Feasibility and Safety
    Chen, Qiaolin
    Sun, Xianjun
    Guo, Lizhen
    Lin, Houwei
    UROLOGY, 2024, 184 : 40 - 44
  • [4] Percutaneous insertion of double-J ureteral stent in children with ureteral obstruction: Our experiences
    Kljucevsek, Damjana
    Kljucevsek, Tomaz
    JOURNAL OF PEDIATRIC UROLOGY, 2013, 9 (02) : 188 - 192
  • [5] TECHNIQUE FOR REMOVAL OF MIGRATED DOUBLE-J URETERAL STENT
    KATSKE, FA
    CELIS, P
    UROLOGY, 1991, 37 (06) : 579 - 579
  • [6] Forgotten Double-J Ureteral Stent
    Lopes Neto, Antonio Correa
    INTERNATIONAL BRAZ J UROL, 2019, 45 (06): : 1087 - 1089
  • [7] EXTENDED DOUBLE-J URETERAL STENT
    KLUTKE, CG
    GOLOMB, J
    RAZ, S
    UROLOGY, 1991, 37 (05) : 444 - 445
  • [8] Insertion of a single double-J stent for bilateral open ureteral reimplantation: introducing a novel technique and assessment of feasibility
    Nabavizadeh, Behnam
    Keihani, Sorena
    Sharifi, Seyed Hossein Hosseini
    Kajbafzadeh, Abdol-Mohammad
    INTERNATIONAL UROLOGY AND NEPHROLOGY, 2016, 48 (07) : 1015 - 1019
  • [9] Insertion of a single double-J stent for bilateral open ureteral reimplantation: introducing a novel technique and assessment of feasibility
    Behnam Nabavizadeh
    Sorena Keihani
    Seyed Hossein Hosseini Sharifi
    Abdol-Mohammad Kajbafzadeh
    International Urology and Nephrology, 2016, 48 : 1015 - 1019
  • [10] Evaluation of pain perception associated with use of the magnetic-end ureteric double-J stent for short-term ureteric stenting
    S. Sevcenco
    K. Eredics
    L. Lusuardi
    Hans Christoph Klingler
    World Journal of Urology, 2018, 36 : 475 - 479