Sensitivity and specificity analysis of 2D small field measurement array: Patient-specific quality assurance of small target treatments and spatially fractionated radiotherapy

被引:7
|
作者
Carmen Banos-Capilla, Maria [1 ,2 ]
Domingo Lago-Martin, Jose [3 ]
Gil, Patricia [1 ]
Maria Larrea, Luis [1 ]
机构
[1] Hosp Vithas Consuelo, Radiat Oncol Dept, Valencia, Spain
[2] Univ Nacl Educ Distancia, Fac Sci, Math & Fluid Phys Dept, Madrid, Spain
[3] Fdn Inst Valenciano Oncol, Radiat Oncol Dept, Valencia, Spain
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2021年 / 22卷 / 10期
关键词
diode arrays; dosimetry; IMRT QA; small-field dosimetry; SRS-resolution arrays; IMRT; RESOLUTION; DOSIMETRY; SEGMENTS; CHAMBER;
D O I
10.1002/acm2.13402
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose"The aim of this paper is to describe the tests carried out on a SRSMapCheck array, to verify its reliability and sensitivity for quality assurance (QA) of high gradient treatments as an alternative system to the use of high spatial resolution detectors, such as gafchromic film, whose processing requires meticulous and time-consuming procedures. Methods: In an initial step, general functionality tests were carried out to verify that the equipment meets the manufacturer's specifications. A study of the accuracy of the application of correction factors to compensate for variation in detector response due to dose rate, field size and beam angle incidence has been included. Besides, to assess the ability of the array to detect inaccurately delivered treatments, systematic errors corresponding to the deviation in the position of the leaves and the accuracy of the gantry position, have been introduced. Based on these results, an estimate of sensitivity and specificity values of the device has been completed. The final step included a study applied to high gradient treatment for real cases of spatially fractionated radiotherapy, where the results of SRSMapCheck measurements have been compared with gafchromic films. Results: General commissioning tests meet the manufacturer's specifications. dose rate (DR) response variation is better than 1.5% and for DR above 50 MU/min better than 1%. The results for beam incidences are better than 1% for all gantry angles, including beam incidences parallel to the array. Field size response differences are within the range of +/- 1% for sizes up to 2 x 2 cm(2), with a maximum value obtained of 3.5%, for 1 x 1 cm(2). From the systematic error study, using a Gamma function Gamma (2%, 2 mm), the detector presents a high specificity with a value greater than 90% at its lower limit, while its sensitivity has a moderate mean value of 81%. Sensitivity values increase above 86% when we apply a Gamma function Gamma (2%, 1 mm) is applied. Finally, the study of actual cases comprises 17 patients, distributed into 11 lung tumors, 3 gynecological and 3 soft tissue tumors. The gafchromic film showed a lower passing rate with an average value of Gamma (2%, 2 mm) = 94.1% compared to Gamma (2%, 2 mm) = 98.6% reached by the measurements with the array. Conclusions: Gamma function obtained with the SRSMapCheck array always presented a higher value than gafchromic film measurements, resulting in a greater number of plans considered correct. This fact, together with the sensitivity and specificity study carried out, allows us to conclude the recommendation that a restrictive metric must be established, in this way we will improve sensitivity, and therefore we will reduce the rate of incorrect plans qualified as correct. The characteristics of the equipment together with the correction factors applied, led to reliably performing acquisitions for complex treatments with multiple small targets in oblique rotational incidences. The spatial resolution of detectors allows the verification of high gradient dose plans such as those achieved in spatially fractionated radiotherapy (SFRT).
引用
收藏
页码:104 / 119
页数:16
相关论文
共 22 条
  • [1] A 2D silicon detector array for quality assurance in small field dosimetry: DUO
    Al Shukaili, Khalsa
    Petasecca, Marco
    Newall, Matthew
    Espinoza, Anthony
    Perevertaylo, Vladimir L.
    Corde, Stephanie
    Lerch, Michael
    Rosenfeld, Anatoly B.
    MEDICAL PHYSICS, 2017, 44 (02) : 628 - 636
  • [2] Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy
    Stathakis, Sotirios
    Myers, Pamela
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    MEDICAL PHYSICS, 2013, 40 (07)
  • [3] Evaluation of Plastic Scintillator Detector for Small Field Stereotactic Patient-Specific Quality Assurance
    Qin, Y.
    Gardner, S.
    Huang, Y.
    Kim, J.
    Wen, N.
    Chetty, I.
    MEDICAL PHYSICS, 2016, 43 (06) : 3664 - 3664
  • [4] Field size analysis of patient-specific quality assurance in scanned carbon ion radiotherapy
    Deng, Yu
    Wang, Qianxia
    Huang, Zhijie
    MEDICAL PHYSICS, 2021, 48 (11) : 6627 - 6633
  • [5] Investigation of 2D radiophotoluminescence films in radiotherapy: Multileaf collimator Quality Assurance and small field dosimetry
    Nascimento, Luana de Freitas
    Castellano, Fabio
    Goossens, Jo
    van Dijk, Anke
    Leblans, Paul
    De Saint-Hubert, Marijke
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1028
  • [6] Technical Note: Evaluation of plastic scintillator detector for small field stereotactic patient-specific quality assurance
    Qin, Yujiao
    Gardner, Stephen J.
    Kim, Joshua
    Huang, Yimei
    Wen, Ning
    Doemer, Anthony
    Chetty, Indrin J.
    MEDICAL PHYSICS, 2017, 44 (10) : 5509 - 5516
  • [7] Correlation analysis between 2D and 3D patient-specific quality assurance for volumetric modulated arc therapy
    Sharma, S.
    Sharma, D.
    Subramani, V.
    Gopishankar, N.
    Bhaskar, S.
    Pathy, S.
    Kumar, P.
    Thulkar, S.
    Chander, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (04): : 685 - 691
  • [8] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Diaz Moreno, Rogelio Manuel
    Venencia, Daniel
    Garrigo, Edgardo
    Pipman, Yakov
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 145 - 151
  • [9] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Rogelio Manuel Diaz Moreno
    Daniel Venencia
    Edgardo Garrigo
    Yakov Pipman
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 145 - 151
  • [10] A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT
    Rajasekaran, Dhanabalan
    Jeevanandam, Prakash
    Sukumar, Prabakar
    Ranganathan, Arulpandiyan
    Johnjothi, Samdevakumar
    Nagarajan, Vivekanandan
    MEDICAL DOSIMETRY, 2014, 39 (04) : 300 - 308