Development of an Analytical Protocol for Determination of Cyanide in Human Biological Samples Based on Application of Ion Chromatography with Pulsed Amperometric Detection

被引:14
|
作者
Jaszczak, Ewa [1 ]
Ruman, Marek [2 ]
Narkowicz, Sylwia [1 ]
Namiesnik, Jacek [1 ]
Polkowska, Zaneta [1 ]
机构
[1] Gdansk Univ Technol, Fac Chem, Dept Analyt Chem, Narutowicza Str 2-12, PL-80233 Gdansk, Poland
[2] Univ Silesia, Fac Earth Sci, Bedzinska 60, PL-41200 Sosnowiec, Poland
关键词
GAS-CHROMATOGRAPHY; THIOCYANATE; BLOOD; VALIDATION; WATER;
D O I
10.1155/2017/7157953
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and accurate ion chromatography (IC) method with pulsed amperometric detection (PAD) was proposed for the determination of cyanide ion in urine, sweat, and saliva samples. The sample pretreatment relies on alkaline digestion and application of Dionex OnGuard II H cartridge. Under the optimized conditions, the method showed good linearity in the range of 1-100 mu g/L for urine, 5-100 mu g/L for saliva, and 3-100 mu g/L for sweat samples with determination coefficients (R) > 0.992. Low detection limits (LODs) in the range of 1.8 mu g/L, 5.1 mu g/L, and 5.8 mu g/L for urine, saliva, and sweat samples, respectively, and good repeatability (CV < 3%, n = 3) were obtained. The proposed method has been successfully applied to the analysis of human biological samples.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Application of ion chromatography with pulsed amperometric detection for the determination of trace cyanide in biological samples, including breast milk
    Jaszczak, Ewa
    Koziol, Krystyna
    Kielbratowska, Bogumila
    Polkowska, Zaneta
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2019, 1110 : 36 - 42
  • [2] Determination of cyanide in urine and saliva samples by ion chromatography with pulsed amperometric detection
    Ewa Jaszczak
    Sylwia Narkowicz
    Jacek Namieśnik
    Żaneta Polkowska
    [J]. Monatshefte für Chemie - Chemical Monthly, 2017, 148 : 1645 - 1649
  • [3] Determination of cyanide ion in urine samples using ion chromatography with pulsed amperometric detection
    Jaszczak, Ewa
    Narkowicz, Sylwia
    Namiesnik, Jacek
    Polkowska, Aneta
    [J]. PROCEEDINGS OF THE 12TH INTERNATIONAL STUDENTS CONFERENCE MODERN ANALYTICAL CHEMISTRY, 2016, : 1 - 5
  • [4] Determination of cyanide in urine and saliva samples by ion chromatography with pulsed amperometric detection
    Jaszczak, Ewa
    Narkowicz, Sylwia
    Namiesnik, Jacek
    Polkowska, Zaneta
    [J]. MONATSHEFTE FUR CHEMIE, 2017, 148 (09): : 1645 - 1649
  • [5] Ion Chromatography with Pulsed Amperometric Detection for Determining Cyanide in Urine and Meconium Samples
    Jaszczak-Wilke, Ewa
    Koziol, Krystyna
    Kielbratowska, Bogumila
    Polkowska, Zaneta
    [J]. MOLECULES, 2021, 26 (15):
  • [6] Direct determination of free cyanide in drinking water by ion chromatography with pulsed amperometric detection
    Christison, Terri T.
    Rohrer, Jeffrey S.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2007, 1155 (01) : 31 - 39
  • [7] Direct determination of hydrogen cyanide in cigarette mainstream smoke by ion chromatography with pulsed amperometric detection
    Zhang, Zi-Wei
    Xu, Ying-Bo
    Wang, Cheng-Hui
    Chen, Kai-Bo
    Tong, Hong-Wu
    Liu, Shao-Min
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (07) : 1016 - 1019
  • [8] Determination of cyanide in bamboo shoots by microdiffusion combined with ion chromatography-pulsed amperometric detection
    Ding, Ming
    Wang, Kailiang
    [J]. ROYAL SOCIETY OPEN SCIENCE, 2018, 5 (04):
  • [9] Rapid measurement of free cyanide in liquor by ion chromatography with pulsed amperometric detection
    Wu, Wenlin
    Xiao, Quanwei
    Zhang, Ping
    Ye, Mei
    Wan, Yuping
    Liang, Hengxing
    [J]. FOOD CHEMISTRY, 2015, 172 : 681 - 684
  • [10] Determination of Carbohydrates in Foodstuffs by Ion Chromatography with Pulsed Amperometric Detection
    Steinbach, Alfred
    Wille, Andrea
    [J]. LC GC EUROPE, 2008, : 15 - 16