Efficiency and safety increases after the implementation of a multi-institutional automated plan check tool at our institution

被引:9
|
作者
Berry, Sean L. [1 ]
Zhou, Ying [1 ]
Pham, Hai [1 ]
Elguindi, Sharif [1 ]
Mechalakos, James G. [1 ]
Hunt, Margie [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
来源
关键词
automation; plan checking; quality and safety; treatment planning;
D O I
10.1002/acm2.12845
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The plan check tool (PCT) is the result of a multi-institutional collaboration to jointly develop a flexible automated plan checking framework designed with the versatility to be shared across collaborating facilities while supporting the individual differences between practices. We analyze the effect that PCT has had on the efficiency and effectiveness of initial chart checks at our institution. Methods and Materials Data on errors identified during initial chart checks were acquired during two time periods: before the introduction of PCT in the clinic (6/24/2015 to 7/31/2015, 187 checks) and post-clinical release (4/14/2016 to 5/2/2016, 186 checks). During each time period, human plan checkers were asked to record all issues that they either manually detected or that were detected by PCT as well as the amount of time, less breaks, or interruptions, it took to check each plan. Results After the clinical release of PCT, there was a statistically significant decrease in the number of issues recorded by the human plan checkers both related to checks explicitly performed by PCT (13 vs 50, P < 0.001) and in issues identified overall (127 vs 200, P < 0.001). The mean and medium time for a plan check decreased by 20%. Conclusions The use of a multi-institutional, configurable, automated plan checking tool has resulted in both substantial gains in efficiency and moving error detection to earlier points in the planning process, decreasing their likelihood that they reach the patient. The sizeable startup effort needed to create this tool from scratch was mitigated by the sharing, and subsequent co-development, of software code from a peer institution.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 34 条
  • [1] A multi-institutional study on VMAT delivery accuracy and efficiency for a benchmark plan
    Pasler, M.
    Kaas, J.
    Perik, T.
    Geuze, J.
    Wittkamper, F.
    Dreindl, R.
    Kuenzler, T.
    Georg, D.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S125 - S126
  • [2] Effects of Implementation of WHO Surgical Safety Check List: Our Institutional Analysis
    Haridarshan, S. J.
    Girish, C. S.
    Rajagopalan, S.
    INDIAN JOURNAL OF SURGERY, 2018, 80 (05) : 465 - 469
  • [3] Fully automated VMAT plan generation - an international multi-institutional validation study
    Heijmen, B.
    Voet, P.
    Fransen, D.
    Akhiat, H.
    Bonomo, P.
    Casati, M.
    Georg, D.
    Goldner, G.
    Henry, A.
    Lilley, J.
    Lohr, F.
    Marrazzo, L.
    Milder, M.
    Pallotta, S.
    Penninkhof, J.
    Seppenwoolde, Y.
    Simontacchi, G.
    Steil, V.
    Stieler, F.
    Wilson, S.
    Pellegrini, R.
    Breedveld, S.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S125 - S125
  • [4] Optimizing Efficiency and Safety in External Beam Radiotherapy Using Automated Plan Check (APC) Tool and Six Sigma Methodology
    Liu, S.
    Bush, K.
    Bertini, J.
    Fu, Y.
    Lewis, J. M.
    Pham, D.
    Yang, Y.
    Niedermayr, T.
    Skinner, L.
    Xing, L.
    Beadle, B. M.
    Hsu, A.
    Kovalchuk, N.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : S22 - S22
  • [5] Optimizing efficiency and safety in external beam radiotherapy using automated plan check (APC) tool and six sigma methodology
    Liu, Shi
    Bush, Karl K.
    Bertini, Julian
    Fu, Yabo
    Lewis, Jonathan M.
    Pham, Daniel J.
    Yang, Yong
    Niedermayr, Thomas R.
    Skinner, Lawrie
    Xing, Lei
    Beadle, Beth M.
    Hsu, Annie
    Koyalchuk, Nataliya
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (08): : 56 - 64
  • [6] Six Sigma-Driven Automated Plan Check (APC) Tool Enhances Safety and Efficiency in External Beam Radiation Therapy
    Kovalchuk, N.
    Liu, S.
    Lim, S.
    Bertini, J.
    Bush, K.
    Lewis, J.
    Pham, D.
    Yang, Y.
    Niedermayr, T.
    Skinner, L.
    Beadle, B.
    Xing, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E476 - E476
  • [7] Six Sigma-driven Automated Plan Check (APC) Tool Enhances Safety and Efficiency in External Beam Radiation Therapy
    Kovalchuk, N.
    Liu, S.
    Lim, S.
    Bertini, J.
    Bush, K.
    Lewis, J. M.
    Pham, D.
    Yang, Y.
    Niedermayr, T.
    Skinner, L.
    Beadle, B. M.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : E199 - E199
  • [8] Multi-Institutional Study of a Bayesian Network-Based Assistive Tool for Physics Plan Review
    Luk, S.
    Kalendralis, P.
    Zegers, C.
    Bermejo, I.
    Kalet, A.
    MEDICAL PHYSICS, 2022, 49 (06) : E155 - E156
  • [9] Effects of Implementation of W.H.O Surgical Safety Check List: Our Institutional Analysis
    S. J. Haridarshan
    C. S. Girish
    S. Rajagopalan
    Indian Journal of Surgery, 2018, 80 : 465 - 469
  • [10] A Multi-Institutional Plan Quality Checking Tool Built On Oncospace: A Shared Radiation Oncology Database System
    Bowers, M.
    Robertson, S.
    Moore, J.
    Wong, J.
    Phillips, M.
    Hendrickson, K.
    Evans, K.
    McNutt, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3366 - 3366