Calibration of a precision current measurement system for high AC voltages using an AC quantum voltmeter

被引:2
|
作者
Ilic, Damir [1 ]
Heinrich, Alexander [2 ]
Meisner, Johann [2 ]
Behr, Ralf [2 ]
机构
[1] Univ Zagreb, Primary Electromagnet Lab FER PEL, Fac Elect Engn & Comp, Zagreb, Croatia
[2] Phys Tech Bundesanstalt PTB, Braunschweig, Germany
关键词
AC quantum voltmeter; AC Josephson metrology; current measurement; high-voltage techniques; CURRENT SHUNTS; WAVE;
D O I
10.1088/1681-7575/ac8482
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper describes the calibration of a high precision AC current measurement device which is required in a system to measure AC voltages up to 800 kV in the frequency range of 10 Hz to 400 Hz. The AC current measurement unit of such a system requires the calibration of the input current in the milliampere range at a frequency of approximately 50 Hz, and all measurements have been made at a suitable frequency of 62.5 Hz. An AC quantum voltmeter (AC-QVM) is used to achieve a traceability chain from DC resistance calibration, when it works as DC quantum standard, and as AC quantum standard for the AC current measurement. All DC and AC measurements needed for one calibration point (one current level) can be done in a rapid sequence within 30 min using the same AC-QVM, such that the combined uncertainty is at a level of 1 mu A A(-1) (k = 1).
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Calibration of a precision current measurement system for high AC voltages using an AC Quantum Voltmeter
    Ilic, D.
    Luddecke, C.
    Heinrich, A.
    Meisner, J.
    Behr, R.
    [J]. 2020 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM), 2020,
  • [2] The AC Quantum Voltmeter used for AC Current Calibrations
    Starkloff, M.
    Bauer, M.
    Schubert, M.
    Lee, J.
    Behr, R.
    Palafox, L.
    Schaidhammer, L.
    Boeck, A. C.
    Fleischmann, P. M.
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [3] Improvements of the AC Quantum Voltmeter
    Behr, R.
    Palafox, L.
    Kohlmann, J.
    [J]. 2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 44 - 45
  • [4] Methods for precision measurement of ac voltages in the millivolt ranges
    Pogliano, U.
    Bosco, G. C.
    Lanzillotti, M.
    [J]. MEASUREMENT, 2008, 41 (02) : 205 - 210
  • [5] A Dual-Channel Calibration System for AC Currents and Small AC Voltages
    Djokic, Branislav
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2015, 64 (06) : 1773 - 1778
  • [6] From AC Quantum Voltmeter to Quantum Calibrator
    Lee, J.
    Behr, R.
    Schumacher, B.
    Palafox, L.
    Schubert, M.
    Starkloff, M.
    Boeck, A. C.
    Fleischmann, P. M.
    [J]. 2016 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2016), 2016,
  • [7] On the Development of an AC-Quantum Voltmeter
    Iuzzolino, Ricardo
    Bierzychudek, Marcos E.
    Palafox, Luis
    Behr, Ralf
    Tedesco, Agustin
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [8] Direct comparison of Josephson waveforms using an AC quantum voltmeter
    Behr, Ralf
    Palafox, Luis
    Ramm, Guenther
    Moser, Harald
    Melcher, Juergen
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2007, 56 (02) : 235 - 238
  • [9] Guarded vector voltmeter for ac ratio standard calibration
    Callegaro, L
    D'Elia, V
    [J]. IMTC/2001: PROCEEDINGS OF THE 18TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1-3: REDISCOVERING MEASUREMENT IN THE AGE OF INFORMATICS, 2001, : 1441 - 1444
  • [10] OPERATION OF THE AC QUANTUM VOLTMETER AT KHZ FREQUENCY
    Behr, Ralf
    Henderson, Dale
    Williams, Jonathan
    Palafox, Luis
    Lee, Jinni
    Pickering, John
    [J]. 2010 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS CPEM, 2010, : 93 - +