Comparative evaluation of the human whole blood and human peripheral blood monocyte tests for pyrogens

被引:13
|
作者
Andrade, SS [1 ]
Silveira, RL [1 ]
Schmidt, CA [1 ]
Brum, L [1 ]
Dalmora, SL [1 ]
机构
[1] Univ Fed Santa Maria, Hlth Sci Ctr, Ind Pharm Dept, BR-97105900 Santa Maria, RS, Brazil
关键词
blood; IL-6; in vitro test; pyrogens; monocytes; quality control;
D O I
10.1016/j.ijpharm.2003.07.005
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Two different in vitro tests for pyrogens, using human peripheral blood monocytes (PBMNC) and diluted whole blood (WBC), respectively, were applied to different classes of parenteral medicinal products. Many of these products did not have a specified endotoxin limit concentration that was established as the maximum valid dilution to comply with the test. The results of the in vitro tests for pyrogens were compared with the results from the Limulus amoebocyte lysate (LAL) and rabbit pyrogen tests. The Second International Standard for endotoxin was used to calibrate all of the assays and the International Standard for IL-6 was used to calibrate the IL-6 ELISA which provided the readout for the in vitro tests for pyrogens. Preparatory tests were conducted to ensure that the "criteria for validity and precision of the standard curve" were satisfied and that the drugs being tested did not interfere in the tests. The PBMNC/IL-6 test had a detection limit of 0.06 EU/ml and spike recoveries were 62-165%. The whole blood/IL-6 test also had a detection limit of 0.06 EU/ml and spike recoveries were 58-132%. The application to the detection of non-endotoxin pyrogens needs to be evaluated in more detail, but the two in vitro tests for pyrogens showed good agreement overall, both with each other and with the LAL test and the rabbit pyrogen test for the detection of endotoxins. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:115 / 124
页数:10
相关论文
共 50 条
  • [1] Development, Validation and Applications of the Monocyte Activation Test for Pyrogens Based on Human Whole Blood
    Schindler, Stefanie
    von Aulock, Sonja
    Daneshian, Mardas
    Hartung, Thomas
    [J]. ALTEX-ALTERNATIVEN ZU TIEREXPERIMENTEN, 2009, 26 (04): : 265 - 277
  • [2] The detection of pyrogens with a method involving human whole blood
    Hartung, T
    Wendel, A
    [J]. ANIMAL ALTERNATIVES, WELFARE, AND ETHICS, 1997, 27 : 683 - 686
  • [3] A COMPARATIVE EVALUATION OF ORTHOPAEDIC CEMENTS IN HUMAN WHOLE BLOOD
    Axen, N.
    Ahnfelt, N. -O.
    Persson, T.
    Hermansson, L.
    Sanchez, J.
    Larsson, R.
    [J]. ADVANCES IN BIOCERAMICS AND BIOCOMPOSITES, 2005, 26 (06): : 71 - 77
  • [4] Heterogeneity of human peripheral blood monocyte subsets
    Grage-Griebenow, E
    Flad, HD
    Ernst, M
    [J]. JOURNAL OF LEUKOCYTE BIOLOGY, 2001, 69 (01) : 11 - 20
  • [5] First Results in the prevaluation of the human whole blood assay for pyrogens in biological pharmaceuticals
    Fischer, M
    Hartzsch, K
    Hartung, T
    Montag-Lessing, T
    [J]. ALTEX-ALTERNATIVEN ZU TIEREXPERIMENTEN, 1998, 15 : 10 - 13
  • [6] Monocyte precursors in the peripheral blood of normal human subjects
    Bigi, G
    Vanoli, M
    Fabio, G
    CantuRajnoldi, A
    [J]. EXPERIMENTAL HEMATOLOGY, 1997, 25 (08) : 543 - 543
  • [7] A simple method for human peripheral blood monocyte isolation
    de Almeida, MC
    Silva, AC
    Barral, A
    Netto, MB
    [J]. MEMORIAS DO INSTITUTO OSWALDO CRUZ, 2000, 95 (02): : 221 - 223
  • [8] MONOCYTE NUMBERS IN NORMAL HUMAN PERIPHERAL-BLOOD
    FERNANDEZ, LA
    MACSWEEN, JM
    [J]. CLINICAL RESEARCH, 1976, 24 (05): : A670 - A670
  • [9] Xenon preserves neutrophil and monocyte function in human whole blood
    de Rossi, L
    Gott, K
    Horn, N
    Hecker, K
    Hutschenreuter, G
    Rossaint, R
    [J]. CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2002, 49 (09): : 942 - 945
  • [10] Inflammatory response to pyrogens determined by a novel ELISA method using human whole blood
    Banerjee, Siddharth
    Mohanan, P. V.
    [J]. JOURNAL OF IMMUNOLOGICAL METHODS, 2011, 369 (1-2) : 146 - 153