Treatment Planning System for Electron FLASH Radiation Therapy: Open-Source for Clinical Implementation

被引:9
|
作者
Rahman, Mahbubur [1 ]
Ashraf, M. Ramish [1 ]
Gladstone, David J. [1 ,2 ,3 ]
Bruza, Petr [1 ]
Jarvis, Lesley A. [2 ,3 ]
Schaner, Philip E. [2 ,3 ]
Cao, Xu [1 ]
Pogue, Brian W. [1 ,2 ,4 ]
Hoopes, P. Jack [1 ,2 ,3 ,4 ]
Zhang, Rongxiao [1 ,2 ,3 ]
机构
[1] Dartmouth Coll, Thayer Sch Engn, Hanover, NH 03755 USA
[2] Dartmouth Coll, Dept Med, Radiat Oncol, Hanover, NH 03755 USA
[3] Dartmouth Hitchcock Med Ctr, Norris Cotton Canc Ctr, Lebanon, NH 03766 USA
[4] Dartmouth Coll, Dept Surg, Hanover, NH 03755 USA
关键词
DOSE-RATE; IRRADIATION; PHOTON; RADIOTHERAPY;
D O I
10.1016/j.ijrobp.2021.10.148
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To present a Monte Carlo (MC) beam model and its implementation in a clinical treatment planning system (TPS, Varian Eclipse) for a modified ultrahigh dose-rate electron FLASH radiation therapy (eFLASH-RT) linear accelerator (LINAC) using clinical accessories and geometry. Methods and Materials: The gantry head without scattering foils or targets, representative of the LINAC modifications, was modeled in the Geant4-based GAMOS MC toolkit. The energy spectrum (sigma(E)) and beam source emittance cone angle (theta(cone)) were varied to match the calculated open-field central-axis percent depth dose (PDD) and lateral profiles with Gafchromic film measurements. The beam model and its Eclipse configuration were validated with measured profiles of the open field and nominal fields for clinical applicators. An MC forward dose calculation was conducted for a mouse whole-brain treatment, and an eFLASH-RT plan was compared with a conventional (Conv-) RT electron plan in Eclipse for a human patient with metastatic renal cell carcinoma. Results: The eFLASH beam model agreed best with measurements at sigma(E) = 0.5 MeV and theta(cone) = 3.9 degrees +/- 0.2 degrees. The model and its Eclipse configuration were validated to clinically acceptable accuracy (the absolute average error was within 1.5% for in-water lateral, 3% for in-air lateral, and 2% for PDDs). The forward calculation showed adequate dose delivery to the entire mouse brain while sparing the organ at risk (lung). The human patient case demonstrated the planning capability with routine accessories to achieve an acceptable plan (90% of the tumor volume receiving 95% and 90% of the prescribed dose for eFLASH and Conv-RT, respectively). Conclusions: To our knowledge, this is the first functional beam model commissioned in a clinical TPS for eFLASH-RT enabling planning and evaluation with minimal deviation from the Conv-RT workflow. It facilitates the clinical translation because eFLASH-RT and Conv-RT plan quality were comparable for a human patient involving complex geometries and tissue heterogeneity. The methods can be expanded to model other eFLASH irradiators with different beam characteristics. (C) 2021 Elsevier Inc. All rights reserved.
引用
收藏
页码:1023 / 1032
页数:10
相关论文
共 50 条
  • [1] Clinical Application of the Histogram Analysis in Radiation Therapy (HART): An Open-Source Software System in Radiation Oncology
    Pyakuryal, A. P.
    Bacchus, I.
    Narayanasamy, G.
    Jang, S.
    Perez-Andujar, A.
    Sio, T. T.
    Kudrimoti, M.
    Mittal, B. B.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S626 - S627
  • [2] Development of a radiation therapy open-source platform
    Gruenwald, Oxana
    Ruhlmann, Juergen
    Buzug, Thorsten M.
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1887 - 1892
  • [3] DosePredict: An Open-source Software to Support Treatment Planning for Personalised Radiopharmaceutical Therapy
    Ferreira, C. Gomes
    Warschun, V. R.
    Hu, J.
    Xue, S.
    Mercolli, L.
    Vidal, T. W.
    Navab, N.
    Afshar-Oromieh, A.
    Rominger, A.
    Shi, K.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S749 - S750
  • [4] Treatment Planning System for Clinical Translation of Electron FLASH Radiotherapy
    Rahman, M.
    Ashraf, M. R.
    Gladstone, D. J.
    Bruza, P.
    Jarvis, L. A.
    Schaner, P. E.
    Cao, X.
    Pogue, B. W.
    Hoopes, P. J.
    Zhang, R.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 111 (03): : S31 - S31
  • [5] Clinical Application of Radiobiological Modeling in the Histogram Analysis in Radiation Therapy (HART): An Open-source Software System in Radiation Oncology
    Pyakuryal, A.
    Sio, T. T. W.
    Narayanasamy, G.
    Lee, C.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E537 - E537
  • [6] Design and clinical implementation of an open-source bionic leg
    Azocar, Alejandro F.
    Mooney, Luke M.
    Duval, Jean-Francois
    Simon, Ann M.
    Hargrove, Levi J.
    Rouse, Elliott J.
    [J]. NATURE BIOMEDICAL ENGINEERING, 2020, 4 (10) : 941 - 953
  • [7] Design and clinical implementation of an open-source bionic leg
    Alejandro F. Azocar
    Luke M. Mooney
    Jean-François Duval
    Ann M. Simon
    Levi J. Hargrove
    Elliott J. Rouse
    [J]. Nature Biomedical Engineering, 2020, 4 : 941 - 953
  • [8] Improvements to the Histogram Analysis in Radiation Therapy ( HART) Open-Source Software System
    Pyakuryal, A.
    Myint, K.
    Gopalakrishnan, M.
    Jang, S.
    Sathiaseelan, V.
    Logemann, J.
    Mittal, B.
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 2547 - +
  • [9] Feasibility and Clinical Implementation of Electron FLASH Radiation Therapy in the Yorkshire Swine Model
    Abana, C. O.
    Palmiero, A. N.
    Velasquez, B. D.
    Liu, K.
    Koong, A. C.
    Beddar, S.
    Mitra, D.
    Schueler, E.
    Lin, S. H.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2023, 117 (02): : E637 - E638
  • [10] Current Status of the Histogram Analysis in Radiation Therapy (HART): An Open-Source Software System
    Pyakuryal, A.
    Kepka, A.
    Gopalakrishnan, M.
    Jang, S.
    Sathiaseelan, V.
    Mittal, B.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)