An X-Band Crossed-Loop EPR Resonator

被引:4
|
作者
Rinard, George A. [2 ,3 ]
Quine, Richard W. [2 ,3 ]
McPeak, Joseph [1 ,2 ]
Buchanan, Laura [1 ,2 ]
Eaton, Sandra S. [1 ,2 ]
Eaton, Gareth R. [1 ,2 ]
机构
[1] Univ Denver, Dept Chem & Biochem, 2101 E Wesley Ave, Denver, CO 80208 USA
[2] Univ Denver, Ctr EPR Imaging Vivo Physiol, 2101 E Wesley Ave, Denver, CO 80208 USA
[3] Univ Denver, Sch Engn & Comp Sci, Denver, CO 80208 USA
基金
美国国家科学基金会;
关键词
ELECTRON DOUBLE-RESONANCE; GAP RESONATOR; SPECTROMETER;
D O I
10.1007/s00723-017-0945-2
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A copper X-band (9.22 GHz) cross-loop resonator has been constructed for use with 4 mm sample tubes. The Q for the two resonators is 380 and 350, respectively. The resonator efficiency is about 1 G per square root of watt. Operation has been demonstrated with measurement of T (1) by saturation recovery for samples of coal and an immobilized nitroxide.
引用
收藏
页码:1219 / 1226
页数:8
相关论文
共 50 条
  • [1] An X-Band Crossed-Loop EPR Resonator
    George A. Rinard
    Richard W. Quine
    Joseph McPeak
    Laura Buchanan
    Sandra S. Eaton
    Gareth R. Eaton
    [J]. Applied Magnetic Resonance, 2017, 48 : 1219 - 1226
  • [2] An L-band crossed-loop (bimodal) EPR resonator
    Rinard, GA
    Quine, RW
    Eaton, GR
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2000, 144 (01) : 85 - 88
  • [3] Easily tunable crossed-loop (bimodal) EPR resonator
    Rinard, GA
    Quine, RW
    Ghim, BT
    Eaton, SS
    Eaton, GR
    [J]. JOURNAL OF MAGNETIC RESONANCE SERIES A, 1996, 122 (01) : 50 - 57
  • [4] A novel loop-gap resonator probehead for EPR and ENDOR at X-band
    Grishin, Y
    Kay, CWM
    Doubinskii, AA
    Mobius, K
    [J]. APPLIED MAGNETIC RESONANCE, 1997, 13 (3-4) : 387 - 392
  • [5] A Novel Loop-Gap Resonator Probehead for EPR and ENDOR at X-Band
    Grishin, Y.
    Kay, C. W. M.
    Doubinskii, A. A.
    Moebius, K.
    [J]. Applied Magnetic Resonance, 13 (3-4):
  • [6] A novel loop-gap resonator probehead for EPR and ENDOR at X-band
    Yu. Grishin
    C. W. M. Kay
    A. A. Doubinskii
    K. Möbius
    [J]. Applied Magnetic Resonance, 1997, 13 : 387 - 392
  • [7] Design and Realization of an Open EPR Resonator at X-Band Frequencies
    Pisa, Stefano
    Della Monaca, Sara
    Di Trocchio, Luigi
    Ficcadenti, Luca
    Mostacci, Andrea
    Pittella, Erika
    Sacco, Giulia
    Fattibene, Paola
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 2019, 55 (02)
  • [8] 250 MHz crossed-loop electron paramagnetic resonator for pulsed resonance
    Rinard, GA
    Ouine, RW
    Eaton, GR
    Eaton, SS
    [J]. CONCEPTS IN MAGNETIC RESONANCE, 2002, 15 (01): : 37 - 46
  • [9] Rutile dielectric loop-gap resonator for X-band EPR spectroscopy of small aqueous samples
    Mett, Richard R.
    Sidabras, Jason W.
    Anderson, James R.
    Klug, Candice S.
    Hyde, James S.
    [J]. JOURNAL OF MAGNETIC RESONANCE, 2019, 307
  • [10] A Wire-Crossed-Loop Resonator for Rapid Scan EPR
    Rinard, George A.
    Quine, Richard W.
    Biller, Joshua R.
    Eaton, Gareth R.
    [J]. CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2010, 37B (02) : 86 - 91