Effect of Ethylene Oxide Sterilization on Polyvinyl Alcohol Hydrogel Compared with Gamma Radiation

被引:1
|
作者
Pohan, Grace [1 ]
Mattiassi, Sabrina [1 ]
Yao, Yuan [1 ]
Zaw, Aung Moe [1 ]
Anderson, Deirdre E. J. [2 ]
Cutiongco, Marie F. A. [3 ]
Hinds, Monica T. [2 ]
Yim, Evelyn K. F. [1 ,4 ,5 ]
机构
[1] Univ Waterloo, Dept Chem Engn, 200 Univ Ave W,Engn 6-3024, Waterloo, ON N2L 3E9, Canada
[2] Oregon Hlth & Sci Univ, Dept Biomed Engn, Portland, OR 97201 USA
[3] Natl Univ Singapore, Mechanobiol Inst Singapore, Singapore, Singapore
[4] Univ Waterloo, Ctr Biotechnol & Bioengn, Waterloo, ON, Canada
[5] Univ Waterloo, Waterloo Inst Nanotechnol, Waterloo, ON, Canada
基金
加拿大自然科学与工程研究理事会; 新加坡国家研究基金会; 美国国家卫生研究院;
关键词
small-diameter vascular grafts; terminal sterilization; irradiation; hemocompatibility; cell adhesion; HIGH-ENERGY RADIATION; POLY(VINYL ALCOHOL); MECHANICAL-PROPERTIES; PVA; DELIVERY; IRRADIATION; CARTILAGE; POLYMERS; GRAFTS; MICRO;
D O I
10.1089/ten.tea.2020.0002
中图分类号
Q813 [细胞工程];
学科分类号
摘要
This study investigated the effects of terminal sterilization of polyvinyl alcohol (PVA) biomaterials using clinically translatable techniques, specifically ethylene oxide (EtO) and gamma (gamma) irradiation. While a few studies have reported the possibility of sterilizing PVA with gamma-radiation, the use of EtO sterilization of PVA requires additional study. PVA solutions were chemically crosslinked with trisodium trimetaphosphate and sodium hydroxide. The three experimental groups included untreated control, EtO, and gamma-irradiation, which were tested for the degree of swelling and water content, and mechanical properties such as radial compliance, longitudinal tensile, minimum bend radius, burst pressure, and suture retention strength. In addition, samples were characterized with scanning electron microscopy, differential scanning calorimetry, X-ray photoelectron spectroscopy, and water contact angle measurements. Cell attachment was assessed using the endothelial cell line EA.hy926, and the sterilized PVA cytotoxicity was studied with a live/dead stain. Platelet and fibrin accumulation was measured using an ex vivo shunt baboon model. Finally, the immune responses of PVA implants were analyzed after a 21-day subcutaneous implantation in rats and a 30-day implantation in baboon. EtO sterilization reduced the PVA graft wall thickness, its degree of swelling, and water content compared with both gamma-irradiated and untreated PVA. Moreover, EtO sterilization significantly reduced the radial compliance and increased Young's modulus. EtO did not change PVA hydrophilicity, while gamma-irradiation increased the water contact angle of the PVA. Consequently, endothelial cell attachment on the EtO-sterilized PVA showed similar results to the untreated PVA, while cell attachment significantly improved on the gamma-irradiated PVA. When exposing the PVA grafts to circulating whole blood, fibrin accumulation of EtO-sterilized PVA was found to be significantly lower than gamma-irradiated PVA. The immune responses of gamma-irradiated PVA, EtO-treated PVA, and untreated PVA were compared. Implanted EtO-treated PVA showed the least MAC387 reaction. The terminal sterilization methods in this study changed PVA hydrogel properties; nevertheless, based on the characterizations performed, both sterilization methods were suitable for sterilizing PVA. We concluded that EtO can be used as an alternative method to sterilize PVA hydrogel material.
引用
收藏
页码:1077 / 1090
页数:14
相关论文
共 50 条
  • [1] Sterilization of Chemically Cross-linked Polyvinyl Alcohol Hydrogel
    Pohan, G.
    Mattiassi, S.
    Yim, E. K.
    TISSUE ENGINEERING PART A, 2017, 23 : S56 - S57
  • [2] Mechanical strength of cortical allografts with gamma radiation versus ethylene oxide sterilization
    Zhou, Zongke
    Qin, Tingwu
    Yang, Jing
    Shen, Bin
    Kang, Pengde
    Pei, Fuxing
    ACTA ORTHOPAEDICA BELGICA, 2011, 77 (05): : 670 - 675
  • [3] Boron containing polyvinyl alcohol/polyethylene oxide/polyvinyl pyrrolidone composites: Preparation, characterization, gamma radiation shielding and gamma radiation effect on it's thermal properties
    Saltan, Fehmi
    Sirin, Kamil
    Aydin, Selim
    Taskopru, Caner
    Yildirim, Yeliz
    RADIATION PHYSICS AND CHEMISTRY, 2024, 214
  • [4] Boron containing polyvinyl alcohol/ polyethylene oxide/polyvinyl pyrrolidone composites: Preparation, characterization, gamma radiation shielding and gamma radiation effect on it's thermal properties
    Saltan, Fehmi
    Şirin, Kamil
    Aydın, Selim
    Taşköprü, Caner
    Yıldırım, Yeliz
    Radiation Physics and Chemistry, 2024, 214
  • [5] EFFECT OF GAMMA-RADIATION ON GELATION IN POLYVINYL-ALCOHOL SOLUTIONS
    CHEN, WX
    BAO, HY
    ZHANG, MW
    RADIATION PHYSICS AND CHEMISTRY, 1985, 26 (01): : 43 - 47
  • [6] The effect of Gamma and Ethylene Oxide Sterilization on a Selection of Active Pharmaceutical Ingredients for Ophthalmics
    Van Cauwenbergh, Tim
    Theys, Emily
    Stroeykens, Dagmar
    Croonenborghs, Bart
    Gillet, Annick
    DeMent, Aaron
    Van Schepdael, Ann
    Haghedooren, Erik
    JOURNAL OF PHARMACEUTICAL SCIENCES, 2022, 111 (07) : 2011 - 2017
  • [7] GAMMA-RAYS AND ETHYLENE-OXIDE STERILIZATION
    STAR, EG
    ZENTRALBLATT FUR BAKTERIOLOGIE MIKROBIOLOGIE UND HYGIENE SERIE B-UMWELTHYGIENE KRANKENHAUSHYGIENE ARBEITSHYGIENE PRAVENTIVE MEDIZIN, 1980, 171 (01): : 33 - 41
  • [8] EFFECT OF MOISTURE ON ETHYLENE OXIDE STERILIZATION
    GILBERT, GL
    HOFFMAN, RK
    SPINER, DR
    PHILLIPS, CR
    GAMBILL, VM
    APPLIED MICROBIOLOGY, 1964, 12 (06) : 496 - &
  • [9] EFFECT OF ETHYLENE-OXIDE STERILIZATION
    UCHIDA, K
    JAPANESE JOURNAL OF VETERINARY SCIENCE, 1971, 33 : 298 - &
  • [10] ELIMINATION OF TOXICITY FROM POLYVINYL TRAYS AFTER STERILIZATION WITH ETHYLENE OXIDE
    RUETER, A
    SCHLEICHER, JB
    APPLIED MICROBIOLOGY, 1969, 18 (06) : 1057 - +