Simplified Method for FROC Observer Study to Evaluate the Diagnostic Accuracy of a Digital Breast Imaging System by Using a CDMAM Phantom

被引:0
|
作者
Tanaka, Rie [1 ]
Akita, Fujiyo [2 ]
Fukuoka, Daisuke [3 ]
Bamba, Yusuke [4 ]
Shiraishi, Junji [5 ]
机构
[1] Kanazawa Univ, Coll Med Pharmaceut & Hlth Sci, Sch Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Shizuoka Canc Ctr, Dept Diagnost Imaging, Shizuoka, Japan
[3] Gifu Univ, Fac Educ, Dept Technol Educ, Gifu, Japan
[4] EIZO Corp, Visual Technol ASIC, R&D, Haku San, Japan
[5] Kumamoto Univ, Fac Life Sci, Kumamoto, Japan
来源
BREAST IMAGING, IWDM 2016 | 2016年 / 9699卷
关键词
FROC observer study; CDMAM; Digital mammography; Jackknife free-response receiver operating characteristic (JAFROC);
D O I
10.1007/978-3-319-41546-8_11
中图分类号
TP31 [计算机软件];
学科分类号
081202 ; 0835 ;
摘要
We propose a simplified method for performing an FROC observer study by using regions of interest (ROIs) of CDMAM phantom images as case sample images. This new method would help researchers to plan and perform FROC observer studies in shorter time, with smaller errors and reduced complexity. As examples of experimental procedure, the digital images obtained by a direct-conversion flat-panel detector (FPD) system in three imaging conditions with different levels of patient dose were used to compare the diagnostic accuracies for low-contrast signal detection. There was a high correlation between the average figure of merit (FOM) obtained by our proposed method and inverse image quality figure (inv. IQF) calculated by CDMAM Analyser 1.55 in each dose level (r = 0.98).
引用
收藏
页码:83 / 88
页数:6
相关论文
共 22 条
  • [1] Dose reduction and its influence on diagnostic accuracy and radiation risk in digital mammography: an observer performance study using an anthropomorphic breast phantom
    Svahn, T.
    Hemdal, B.
    Ruschin, M.
    Chakraborty, D. P.
    Andersson, I.
    Tingberg, A.
    Mattsson, S.
    BRITISH JOURNAL OF RADIOLOGY, 2007, 80 (955): : 557 - 562
  • [2] A simplified scoring protocol to improve diagnostic accuracy with the breast imaging reporting and data system in breast magnetic resonance imaging
    Zhong, Yuting
    Li, Menglu
    Zhu, Jingjin
    Zhang, Boya
    Liu, Mei
    Wang, Zhili
    Wang, Jiandong
    Zheng, Yiqiong
    Cheng, Liuquan
    Li, Xiru
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (07) : 3860 - 3872
  • [3] Experimental phantom lesion detectability study using a digital breast tomosynthesis prototype system
    Schulz-Wendtland, R.
    Wenkel, E.
    Lell, M.
    Boehner, C.
    Bautz, W. A.
    Mertelmeier, T.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (12): : 1219 - 1223
  • [4] Evaluation of Image Quality in Digital Mammography and Digital Breast Tomosynthesis: Phantom Observer Study From American College of Radiology Imaging Network PA 4006
    Thomas, M.
    Matsutani, Y.
    Conant, E.
    Maidment, A.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [5] Using Maximum Intensity Projection in the Evaluation of Digital Breast Tomosynthesis: A Phantom Observer Study From the American College of Radiology Imaging Network PA 4006 Trial
    Thomas, M.
    Matsutani, Y.
    Conant, E.
    Maidment, A.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [6] Molecular Imaging Using Light-Absorbing Imaging Agents and a Clinical Optical Breast Imaging System—a Phantom Study
    Stephanie M. W. Y. van de Ven
    Niculae Mincu
    Jean Brunette
    Guobin Ma
    Mario Khayat
    Debra M. Ikeda
    Sanjiv S. Gambhir
    Molecular Imaging and Biology, 2011, 13 : 232 - 238
  • [7] Molecular Imaging Using Light-Absorbing Imaging Agents and a Clinical Optical Breast Imaging System-a Phantom Study
    van de Ven, Stephanie M. W. Y.
    Mincu, Niculae
    Brunette, Jean
    Ma, Guobin
    Khayat, Mario
    Ikeda, Debra M.
    Gambhir, Sanjiv S.
    MOLECULAR IMAGING AND BIOLOGY, 2011, 13 (02) : 232 - 238
  • [8] A Phantom Study of the Spatial Precision and Accuracy of Stereotactic Localization Using Computed Tomography Imaging with the Leksell Stereotactic System
    Tafreshi, Ali R.
    Peng, Terrance
    Yu, Cheng
    Kramer, Daniel R.
    Gogia, Angad S.
    Lee, Morgan B.
    Barbaro, Michael F.
    Sebastian, Rinu
    Del Campo-Vera, Roberto Martin
    Chen, Kuang-Hsuan
    Kellis, Spencer S.
    Lee, Brian
    WORLD NEUROSURGERY, 2020, 139 : E297 - E307
  • [9] A METHOD TO STUDY THE EFFECT OF CONTROLLED CHANGES OF BREAST IMAGE TOTAL SYSTEM MTFS ON DIAGNOSTIC-ACCURACY
    BENCOMO, JA
    HAUS, AG
    PAULUS, DD
    HILL, CE
    LOGAN, WW
    JOHNSTON, DA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1982, 139 (02) : 426 - 426
  • [10] An in vitro study to evaluate the accuracy of stereotactic localization using magnetic resonance imaging by means of the Leksell Stereotactic System
    Gliemroth, J
    Gaebel, C
    Kehler, U
    Grande-Nagel, I
    Missler, U
    Arnold, H
    MINIMALLY INVASIVE NEUROSURGERY, 2002, 45 (01) : 1 - 5