A New Model of Centrifugal Blood Pump for Cardiopulmonary Bypass: Design Improvement, Performance, and Hemolysis Tests

被引:11
|
作者
Leme, Juliana
Fonseca, Jeison
Bock, Eduardo [1 ]
da Silva, Cibele
da Silva, Bruno Utiyama
dos Santos, Alex Eugenio [2 ]
Dinkhuysen, Jarbas
Andrade, Aron
Biscegli, Jose F.
机构
[1] Univ Estadual Campinas, Sao Paulo, Brazil
[2] UNIPAC Div, Pompeia, Brazil
关键词
Blood pump; Centrifugal pump; Hemolysis; Cardiopulmonary bypass; Axial pump; SPIRAL PUMP; SYSTEM;
D O I
10.1111/j.1525-1594.2011.01254.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new model of blood pump for cardiopulmonary bypass (CPB) application has been developed and evaluated in our laboratories. Inside the pump housing is a spiral impeller that is conically shaped and has threads on its surface. Worm gears provide an axial motion of the blood column. Rotational motion of the conical shape generates a centrifugal pumping effect and improves pumping performance. One annular magnet with six poles is inside the impeller, providing magnetic coupling to a brushless direct current motor. In order to study the pumping performance, a mock loop system was assembled. Mock loop was composed of Tygon tubes (Saint-Gobain Corporation, Courbevoie, France), oxygenator, digital flowmeter, pressure monitor, electronic driver, and adjustable clamp for flow control. Experiments were performed on six prototypes with small differences in their design. Each prototype was tested and flow and pressure data were obtained for rotational speed of 1000, 1500, 2000, 2500, and 3000 rpm. Hemolysis was studied using pumps with different internal gap sizes (1.35, 1.45, 1.55, and 1.7 mm). Hemolysis tests simulated CPB application with flow rate of 5 L/min against total pressure head of 350 mm Hg. The results from six prototypes were satisfactory, compared to the results from the literature. However, prototype #6 showed the best results. Best hemolysis results were observed with a gap of 1.45 mm, and showed a normalized index of hemolysis of 0.013 g/100 L. When combined, axial and centrifugal pumping principles produce better hydrodynamic performance without increasing hemolysis.
引用
收藏
页码:443 / 447
页数:5
相关论文
共 50 条
  • [1] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Gobel, C
    Eilers, R
    Reul, H
    Schwindke, P
    Jorger, M
    Rau, G
    [J]. ARTIFICIAL ORGANS, 1997, 21 (07) : 841 - 845
  • [2] Comparison of a centrifugal pump with a peristaltic roller pump during cardiopulmonary bypass: Effects on blood hemolysis and coagulation
    Lieberman, JA
    Tammineedi, S
    Aukerman, J
    [J]. ANESTHESIA AND ANALGESIA, 1999, 88 (02): : U70 - U70
  • [3] Improvement of two-stage centrifugal blood pump for cardiopulmonary support system and evaluation of anti-hemolysis performance
    Horiguchi, Hironori
    Tsukiya, Tomonori
    Takemika, Toratarou
    Nomoto, Takeshi
    Tsujimoto, Yoshinobu
    [J]. Horiguchi, H. (horiguti@me.es.osaka-u.ac.jp), 1600, Japan Society of Mechanical Engineers (79): : 492 - 504
  • [4] Improvement of two-stage centrifugal blood pump for cardiopulmonary support system and evaluation of anti-hemolysis performance
    Horiguchi, Hironori
    Tsukiya, Tomonori
    Takemika, Toratarou
    Nomoto, Takeshi
    Tsujimoto, Yoshinobu
    [J]. International Journal of Fluid Machinery and Systems, 2015, 8 (01) : 1882 - 9554
  • [5] Hydrodynamic design and performance analysis of a centrifugal blood pump for cardiopulmonary circulation
    Oh, HW
    Yoon, ES
    Park, MR
    Sun, K
    Hwang, CM
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART A-JOURNAL OF POWER AND ENERGY, 2005, 219 (A7) : 525 - 532
  • [6] Improvement of hemolysis performance in a hydrodynamically levitated centrifugal blood pump by optimizing a shroud size
    Kosaka, Ryo
    Sakota, Daisuke
    Nishida, Masahiro
    Maruyama, Osamu
    Yamane, Takashi
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2021, 24 (02) : 157 - 163
  • [7] Improvement of hemolysis performance in a hydrodynamically levitated centrifugal blood pump by optimizing a shroud size
    Ryo Kosaka
    Daisuke Sakota
    Masahiro Nishida
    Osamu Maruyama
    Takashi Yamane
    [J]. Journal of Artificial Organs, 2021, 24 : 157 - 163
  • [8] In Vivo Evaluation of Centrifugal Blood Pump for Cardiopulmonary Bypass-Spiral Pump
    da Silva, Cibele
    da Silva, Bruno Utiyama
    Leme, Juliana
    Uebelhart, Beatriz
    Dinkhuysen, Jarbas
    Biscegli, Jose F.
    Andrade, Aron
    Zavaglia, Cecilia
    [J]. ARTIFICIAL ORGANS, 2013, 37 (11) : 954 - 957
  • [9] Development of design methods for a centrifugal blood pump with a fluid dynamic approach: Results in hemolysis tests
    Masuzawa, T
    Tsukiya, T
    Endo, S
    Tatsumi, E
    Taenaka, Y
    Takano, H
    Yamane, T
    Nishida, M
    Asztalos, B
    Miyazoe, Y
    Ito, K
    Sawairi, T
    Konishi, Y
    [J]. ARTIFICIAL ORGANS, 1999, 23 (08) : 757 - 761
  • [10] A COMPACT CENTRIFUGAL PUMP FOR CARDIOPULMONARY BYPASS
    SASAKI, T
    JIKUYA, T
    AIZAWA, T
    SHIONO, M
    SAKUMA, I
    TAKATANI, S
    GLUECK, J
    NOON, GP
    NOSE, Y
    DEBAKEY, ME
    [J]. ARTIFICIAL ORGANS, 1992, 16 (06) : 592 - 598