A simple, accurate, and effective capillary electrophoresis method with ultraviolet absorbance detection was developed and validated for the quantitation of the antihistamine fexofenadine in capsules. The separation was performed with an uncoated fused-silica capillary (47 cm x 75 Lm id) and was operated at 20 W potential. Temperature was maintained at 25 degrees C. The run buffer was prepared with 20mM Na2B4O7 X 10 H2O- Software was used for system control, data acquisition, and analysis. Method validation was performed by evaluation of the analytical parameters linearity, precision, accuracy, limits of detection and quantitation, and specificity. The method was linear (r = 0.9999) at concentrations ranging from 20 to 100 mu g/mL, precise (relative standard deviation intra-assay = 1.2,1.6, and 1.8% and interassay = 1.5%); accurate (recovery = 98.1 %); and specific. The limits of detection and quantitation were 0.69 and 2.09 mu g/mL, respectively. The method was compared to the liquid chromatography method developed previously by the authors for the same drug, and no significant difference was found between the 2 methods in fexofenadine hydrochloride quantitation.