Determination of organophosphate pesticides at a carbon nanotube/organophosphorus hydrolase electrochemical biosensor

被引:190
|
作者
Deo, RP
Wang, J
Block, I
Mulchandani, A
Joshi, KA
Trojanowicz, M
Scholz, F
Chen, W
Lin, YH
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
[2] Ernst Moritz Arndt Univ Greifswald, Inst Chem & Biochem, D-17489 Greifswald, Germany
[3] Univ Calif Riverside, Dept Chem & Environm Engn, Riverside, CA 92521 USA
[4] Warsaw Univ, Dept Chem, PL-02093 Warsaw, Poland
[5] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
OPH; CNT; biosensors; oganophosphorus compounds; nitrophenol;
D O I
10.1016/j.aca.2004.09.072
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric biosensor for oganophosphorus (OP) pesticides based on a carbon nanotube (CNT)-modified transducer and an organophosphorus hydrolase (OPH) biocatalyst is described. A bilayer approach with the OPH layer atop of the CNT film was used for preparing the CNT/OPH biosensor. The CNT layer leads to a greatly improved anodic detection of the enzymatically generated p-nitrophenol product, including higher sensitivity and stability. The sensor performance was optimized with respect to the surface modification and operating conditions. Under the optimal conditions the biosensor was used to measure as low as 0.15 muM paraoxon and 0.8 muM methyl parathion with sensitivities of 25 and 6 nA/muLM, respectively. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 50 条
  • [1] Amperometric microbial biosensor for direct determination of organophosphate pesticides using recombinant microorganism with surface expressed organophosphorus hydrolase
    Mulchandani, P
    Chen, W
    Mulchandani, A
    Wang, J
    Chen, L
    [J]. BIOSENSORS & BIOELECTRONICS, 2001, 16 (7-8): : 433 - 437
  • [2] Organophosphate Hydrolase in Conductometric Biosensor for the Detection of Organophosphate Pesticides
    Mulyasuryani, Ani
    Prasetyawan, Sasangka
    [J]. ANALYTICAL CHEMISTRY INSIGHTS, 2015, 10 : 23 - 27
  • [3] Organophosphorus hydrolase-based assay for organophosphate pesticides
    Rogers, KR
    Wang, Y
    Mulchandani, A
    Mulchandani, P
    Chen, W
    [J]. BIOTECHNOLOGY PROGRESS, 1999, 15 (03) : 517 - 521
  • [4] DETERMINATION OF ORGANOPHOSPHORUS AND CARBAMIC PESTICIDES WITH A CHOLINE AND ACETYLCHOLINE ELECTROCHEMICAL BIOSENSOR
    BERNABEI, M
    CREMISINI, C
    MASCINI, M
    PALLESCHI, G
    [J]. ANALYTICAL LETTERS, 1991, 24 (08) : 1317 - 1331
  • [5] An amperometric biosensor for the determination of organophosphorus pesticides
    Telefoncu, A
    [J]. TURKISH JOURNAL OF CHEMISTRY, 1996, 20 (03): : 253 - 257
  • [6] CONDUCTOMETRIC BIOSENSOR FOR DETERMINATION OF ORGANOPHOSPHORUS PESTICIDES
    DZYADEVICH, SV
    SOLDATKIN, AP
    SHULGA, AA
    STRIKHA, VI
    ELSKAYA, AV
    [J]. JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 49 (08) : 789 - 792
  • [7] Hybrid electrochemical biosensor for organophosphorus pesticides quantification
    Stoytcheva, Margarita
    Zlatev, Roumen
    Velkova, Zdravka
    Valdez, Benjamin
    Ovalle, Marcela
    Petkov, Lubomir
    [J]. ELECTROCHIMICA ACTA, 2009, 54 (06) : 1721 - 1727
  • [8] Preparation of Electrochemical Biosensor for Detection of Organophosphorus Pesticides
    Gothwal, Ashish
    Beniwal, Puneet
    Dhull, Vikas
    Hooda, Vikas
    [J]. INTERNATIONAL JOURNAL OF ANALYTICAL CHEMISTRY, 2014, 2014
  • [9] Development of a versatile organophosphorus hydrolase-based assay for organophosphate pesticides
    Rogers, KR
    Wang, Y
    Mulchandani, A
    Mulchandani, P
    Chen, W
    [J]. ENVIRONMENTAL MONITORING AND REMEDIATION TECHNOLOGIES, 1999, 3534 : 420 - 428
  • [10] Biosensor for direct determination of organophosphorus pesticides.
    Lei, Y
    Mulchandani, P
    Chen, W
    Wang, J
    Mulchandani, A
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2003, 225 : U129 - U129