A test-method for determination of molybdenum(VI) using phenylfluorone on a solid phase of fibrous sorbents

被引:0
|
作者
Dedkova, VP [1 ]
Shvoeva, OP [1 ]
Savvin, SB [1 ]
机构
[1] Russian Acad Sci, VI Vernadskii Inst Geochem & Analyt Chem, Moscow, Russia
来源
INDUSTRIAL LABORATORY | 2000年 / 66卷 / 05期
关键词
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The system of phenylfluorone - fibrous sorbent filled with AV-17 anion exchanger is studied and a test method for visual determination of molybdenum is proposed. Mo(VI) is sorbed from a solution of ascorbic acid on a solid phase and detected using a solution of phenylfluorone in 0.5 M HCl. Determination of Mo(VI) is done using a color scale from orange to crimson with a concentration range of 0.025 - 0.4 mug/ml. The test method is used to analyze river water and tap water. The detection limit for molybdenum is 0.02 mg/liter. The relative error does not exceed 25%. The time of analysis is 10 min.