An accurate continuous calibration system for high voltage current transformer

被引:9
|
作者
Tong, Yue [1 ]
Li, Bin Hong [1 ]
机构
[1] Huazhong Univ Sci & Technol, Coll Elect & Elect Engn, Wuhan 430074, Hubei Province, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2011年 / 82卷 / 02期
关键词
D O I
10.1063/1.3523341
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A continuous calibration system for high voltage current transformers is presented in this paper. The sensor of this system is based on a kind of electronic instrument current transformer, which is a clamp-shape air core coil. This system uses an optical fiber transmission system for its signal transmission and power supply. Finally the digital integrator and fourth-order convolution window algorithm as error calculation methods are realized by the virtual instrument with a personal computer. It is found that this system can calibrate a high voltage current transformer while energized, which means avoiding a long calibrating period in the power system and the loss of power metering expense. At the same time, it has a wide dynamic range and frequency band, and it can achieve a high accuracy measurement in a complex electromagnetic field environment. The experimental results and the on-site operation results presented in the last part of the paper, prove that it can reach the 0.05 accuracy class and is easy to operate on site. (C) 2011 American Institute of Physics. [doi:10.1063/1.3523341]
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Research on the Accurate Continuous Calibration System for Current Transformer
    Fan, Yonghong
    Zhao, Jiandong
    Li, Huaguang
    [J]. Proceedings of the 2016 6th International Conference on Applied Science, Engineering and Technology (ICASET), 2016, 77 : 45 - 48
  • [2] Voltage Transformer Calibration System Based on a Digital Current Comparator
    Trigo, Leonardo
    Brehm, Marcelo
    Slomovitz, Daniel
    [J]. 2018 IEEE 9TH POWER, INSTRUMENTATION AND MEASUREMENT MEETING (EPIM), 2018,
  • [3] Harmonic Mitigation of an Improved High Voltage Current Source for On-site System Calibration of High Voltage Transformer Loss Measurement Systems
    Said, Amr
    So, Eddy
    Angelo, Dave
    [J]. 2018 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS (CPEM 2018), 2018,
  • [4] An accurate online calibration system based on combined clamp-shape coil for high voltage electronic current transformers
    Li, Zhen-hua
    Li, Hong-bin
    Zhang, Zhi
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (07):
  • [5] A high-current calibration system based on indirect comparison of current transformer and Rogowski coil
    Luo, Pandian
    Li, Zhenhua
    Li, Hongbin
    Li, Hongfeng
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2013, 24 (12)
  • [6] Continuous and Accurate PCRAM Current-voltage Model
    Jung, Chul-Moon
    Lee, Eun-Sub
    Min, Kyeong-Sik
    [J]. JOURNAL OF SEMICONDUCTOR TECHNOLOGY AND SCIENCE, 2011, 11 (03) : 162 - 168
  • [7] GATED, HIGH-VOLTAGE IONTOPHORESIS SYSTEM WITH ACCURATE CURRENT MONITORING
    SMITH, BM
    HOFFER, BJ
    [J]. ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1978, 44 (03): : 398 - 402
  • [8] Voltage Dependence of the Reference System in Medium- and High-Voltage Current Transformer Calibrations
    van den Brom, Helko E.
    van Leeuwen, Ronald
    Rietveld, Gert
    Houtzager, Ernest
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2021, 70
  • [9] Electromagnetic modeling of a high voltage current transformer
    Collett, Walter
    Mahajan, Satish
    [J]. 2007 39TH NORTH AMERICAN POWER SYMPOSIUM, VOLS 1 AND 2, 2007, : 80 - +
  • [10] NRC-PTB intercomparison of voltage transformer calibration systems for high voltage
    So, E
    Latzel, HG
    [J]. 2000 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2000, : 329 - 330