Automatic flow titrator based on a multicommutated unsegmented flow system for alkalinity monitoring in wastewaters

被引:11
|
作者
Almeida, CMNV [1 ]
Lapa, RAS [1 ]
Lima, JLFC [1 ]
机构
[1] Univ Porto, Fac Farm, CEQUP, Dept Quim Fis, P-4050047 Porto, Portugal
关键词
automatic flow titrator; alkalinity; monitoring; potentiometry; anaerobic digestion; wastewaters;
D O I
10.1016/S0003-2670(00)01341-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A full automatic flow system based on potentiometric titration for alkalinity monitoring in wastewater treating plants is presented. Titration to an end-point of pH 5.75 partial alkalinity CPA) and then to pH 4.3 intermediate alkalinity CIA) allows to distinguish the relative buffering contributions of both bicarbonate and volatile acids in anaerobic digesters and, thus, the attainment of a IA:PA ratio which is a sensitive parameter of digestion monitoring that increases rapidly with the process upset. The titration approach is based on a time-based sequential introduction of increasing titrant and decreasing titrand volumes into a mixing chamber The theoretical model for the titration process already presented and discussed was used to determine the titrand concentration without prior calibration. The interface between the analytical system and the anaerobic digester was also developed, allowing completely automated on-line monitoring of alkalinity. The results obtained were reproducible (R.S.D. lower than 2.4 and 5.1% for repeatability and reproducibility, respectively) and in good agreement with those given by the comparative method. (C) 2001 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:291 / 298
页数:8
相关论文
共 50 条
  • [1] Precipitation titrations using an automatic titrator based on a multicommutated unsegmented flow system
    Almeida, CMNV
    Araújo, MCU
    Lapa, RAS
    Lima, JLFC
    Reis, BF
    Zagatto, EAG
    ANALYST, 2000, 125 (02) : 333 - 340
  • [2] Oxidability determination in waste waters using an automatic titrator based on a multicommutated unsegmented flow system
    Almeida, CMNV
    Lapa, RAS
    Lima, JLFC
    INTERNATIONAL JOURNAL OF ENVIRONMENTAL ANALYTICAL CHEMISTRY, 2000, 78 (3-4) : 315 - 332
  • [3] An automatic titrator based on a multicommutated unsegmented flow system -: Its application to acid-base titrations
    Almeida, CMNV
    Lapa, RAS
    Lima, JLFC
    Zagatto, EAG
    Araújo, MCU
    ANALYTICA CHIMICA ACTA, 2000, 407 (1-2) : 213 - 223
  • [4] Automatic multicommutated flow system for ethanol determination in alcoholic beverages by spectrophotometry
    Comitre, ALD
    Reis, BF
    LABORATORY ROBOTICS AND AUTOMATION, 2000, 12 (01) : 31 - 36
  • [5] AUTOMATIC CALIBRATION AND DILUTION IN UNSEGMENTED FLOW SYSTEMS
    AGUDO, M
    RIOS, A
    VALCARCEL, M
    ANALYTICA CHIMICA ACTA, 1992, 264 (02) : 265 - 273
  • [6] A MULTICOMMUTATED FLOW SYSTEM FOR PERACETIC ACID DETERMINATION
    Campos Barbosa da Silva, Fabiola Soraia Vital
    da Silva Junior, Joilton Jose
    Lavorante, Andre Fernando
    Silveira Paim, Ana Paula
    BRAZILIAN JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 1 (03): : 158 - 163
  • [7] Automated potentiometric titrator based in monosegmented flow system.
    Ganzarolli, EM
    Lehmkuhl, A
    de Queiróz, RRR
    de Souza, IG
    QUIMICA NOVA, 1999, 22 (01): : 53 - 57
  • [8] A multicommutated flow system based on an opened-loop with micropump propulsion
    Rodrigues, Eunice R. G. O.
    Lapa, Rui A. S.
    Lima, Jose L. F. C.
    ANALYTICAL LETTERS, 2007, 40 (08) : 1632 - 1645
  • [9] Bianalyte multicommutated flow analysis system for microproteinuria diagnostics
    Strzelak, Kamil
    Misztal, Jagoda
    Tymecki, Lukasz
    Koncki, Robert
    TALANTA, 2016, 148 : 707 - 711
  • [10] AUTOMATIC IMPLEMENTATION OF THE METHOD OF STANDARD ADDITIONS IN UNSEGMENTED FLOW SYSTEMS
    AGUDO, M
    RIOS, A
    VALCARCEL, M
    ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) : 77 - 84