Scatter radiation dose profile evaluation in computed tomography using Monte Carlo simulation

被引:0
|
作者
Mohammadi, Gh R. Fallah [1 ]
Hesamnezhad, L. [2 ]
Mahdavi, M. [2 ]
机构
[1] Mazandaran Univ Med Sci, Fac Allied Med, Dept Radiol, Sari, Iran
[2] Univ Mazandaran, Dept Phys, Babolsar, Iran
来源
关键词
Dose profile; CT scan; scatter radiation; CTDI phantom; MC simulation; CONE-BEAM CT; DOSIMETRY;
D O I
10.29242/ijrr.19.4.813
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Conventional radiation dosimetry methods in computed tomography (CT) are not able to measure the dose distribution along the patient's longitudinal axis. To calculate the dose index on a CT scan, the dose distribution from the center of the radiation field must be calculated. In this study, the most appropriate integral interval for calculating the CT dose index in the axial mode was determined using the Monte Carlo (MC) method based on X-ray photon energy and slice thickness. Materials and Methods: The computed tomography dose index (CTDI) phantom was simulated in the EGSnrc/BEAMnrcMC system and was irradiated with several X-ray energies and several slice thicknesses and dose profiles in phantom were investigated. The area under the dose profile and the scatter to primary radiation dose ratio (SPR) were calculated. Results: The range of scattered beams from the center of the radiation field reaches 450 mm in 140 kV and a 40 mm slice thickness. The SPR value for all levels of X-ray photon energy (between 80 and 140 kV) significantly decreases as slice thickness increases. CT scan imaging technical factors greater than 310 mm from the center of the slice thickness have no effect on the behavior of the scattered radiation. Conclusion: The primary beams are more affected by the energy of the photons, and the scatter beams are more strongly affected by the slice thickness. For 64-slice scanners, the polymethyl methacrylate (PMMA) phantom length should be between 700 mm and 900 mm to yield accurate CTDI estimations.
引用
收藏
页码:813 / 818
页数:6
相关论文
共 50 条
  • [1] Monte Carlo Simulation of Coherent Scatter Computed Tomography (CSCT)
    Lewis, K.
    Donnelly, E.
    Pickens, D.
    Price, R.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Evaluation of radiation dose to organs during kilovoltage cone-beam computed tomography using Monte Carlo simulation
    Son, Kihong
    Cho, Seungryong
    Kim, Jin Sung
    Han, Youngyih
    Ju, Sang Gyu
    Choi, Doo Ho
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (02): : 295 - 302
  • [3] Estimation of Computed Tomography Dose Index (CTDI) Using Monte Carlo Simulation
    Talbi, Mohammed
    Aabid, Mohammed
    El Mansouri, M'hamed
    Nhila, Oussama
    Semghouli, Slimane
    Sebihi, Rajaa
    Choukri, Abdelmajid
    Khalis, Mohammed
    [J]. MOSCOW UNIVERSITY PHYSICS BULLETIN, 2021, 76 (SUPPL 1) : S103 - S109
  • [4] Estimation of Computed Tomography Dose Index (CTDI) Using Monte Carlo Simulation
    Mohammed Talbi
    Mohammed Aabid
    M’hamed El Mansouri
    Oussama Nhila
    Slimane Semghouli
    Rajaa Sebihi
    Abdelmajid Choukri
    Mohammed Khalis
    [J]. Moscow University Physics Bulletin, 2021, 76 : S103 - S109
  • [5] Monte Carlo Simulation of the Computed Tomography Dose Index (CTDI) Using GATE
    Mkimel, M.
    El Baydaoui, R.
    Mesradi, M. R.
    Tahiri, Z.
    Talasmat, K.
    Halimi, A.
    Krim, M.
    Saad, E.
    Hilali, A.
    [J]. PHYSICS OF PARTICLES AND NUCLEI LETTERS, 2020, 17 (06) : 900 - 907
  • [6] Monte Carlo Simulation of the Computed Tomography Dose Index (CTDI) Using GATE
    M. Mkimel
    R. El Baydaoui
    M. R. Mesradi
    Z. Tahiri
    K. Talasmat
    A. Halimi
    M. Krim
    E. Saad
    A. Hilali
    [J]. Physics of Particles and Nuclei Letters, 2020, 17 : 900 - 907
  • [7] CTmod - A toolkit for Monte Carlo simulation of projections including scatter in computed tomography
    Malusek, Alexandr
    Sandborg, Michael
    Carlsson, Gudrun Alm
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2008, 90 (02) : 167 - 178
  • [8] Estimating organ dose in computed tomography using tube current modulation: A Monte Carlo simulation
    Hosseinzadeh, V
    Ghaffari, H.
    Rezaeyan, A.
    Deilami, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 575 - 581
  • [9] Combined Organ Dose for Digital Subtraction Angiography and Computed Tomography Using Monte Carlo Simulation
    Sakabe, D.
    Ohno, T.
    Araki, F.
    Hashida, M.
    Funama, Y.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3777 - 3777
  • [10] Evaluation of the effect of patient dose from cone beam computed tomography on prostate IMRT using Monte Carlo simulation
    Chow, James C. L.
    Leung, Michael K. K.
    Islam, Mohammad K.
    Norrlinger, Bernhard D.
    Jaffray, David A.
    [J]. MEDICAL PHYSICS, 2008, 35 (01) : 52 - 60