Evaluation of the Bio-Rad VARIANT™ II HbA2/HbA1C Dual Program for measurement of hemoglobin concentrations and detection of variants

被引:8
|
作者
Riou, J
Godart, C
Mathis, M
Hurtrel, D
Wajcman, H
Préhu, C
Bardakdjian, J
机构
[1] Hop Henri Mondor, Serv Biochem Genet, F-94010 Creteil, France
[2] Hop Henri Mondor, INSERM, U468, F-94010 Creteil, France
关键词
hemoglobin A(2); hemoglobin F; sickle cell disease; thalassemia;
D O I
10.1515/CCLM.2005.040
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
In this work data obtained on the VARIANT(TM) II hemoglobin analyzer using the Dual Kit elution system were compared to those obtained with the beta-Thalassemia Short Program. Since many laboratories still use an earlier model of the hemoglobin analyzer, the Variant(TM) 1, these data were also compared to those obtained with the latter instrument. Our study is divided into two parts. The first is an evaluation of the precision of the VARIANT(TM) II for determining the levels of hemoglobin (Hb)A(2) and Hb F. This was carried out for normal subjects, thalassemic patients and patients with HbS. The second part concerns the potential of this instrument in the presumptive identification of Hb variants.
引用
收藏
页码:237 / 243
页数:7
相关论文
共 50 条
  • [1] Evaluation and comparison of hemoglobin HbA1c on the Bio-Rad Variant II to total glycated hemoglobin on the Bio-Rad Variant Hemoglobin Testing Systems.
    Toney, LJ
    McLawhon, RW
    Zibrat, SJ
    CLINICAL CHEMISTRY, 2001, 47 (06) : A4 - A4
  • [2] Analytical evaluation of the Bio-Rad Variant II automated HbA1c analyzer
    Higgins, TN
    Blakney, GB
    Dayton, J
    CLINICAL BIOCHEMISTRY, 2001, 34 (05) : 361 - 365
  • [3] Evaluation of an Improved Bio-Rad VARIANT™ II TURBO HbA1c Kit
    Tapply, C. B.
    Yu, F. H.
    Lee, H. K.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A175 - A175
  • [4] Silent haemoglobin variants and determination of HbA1c with the HPLC Bio-Rad Variant II
    Lahousen, T
    Roller, RE
    Lipp, RW
    Schnedl, WJ
    JOURNAL OF CLINICAL PATHOLOGY, 2002, 55 (09) : 699 - 703
  • [5] Evaluation of the new kit HbA1c Analyzer 2.0 Variant II Turbo (Bio-Rad)
    Meurice, Julie
    Guillard, Emmanuelle
    Jaisson, Stephane
    Leroy, Nathalie
    Gillery, Philippe
    ANNALES DE BIOLOGIE CLINIQUE, 2011, 69 (05) : 561 - 569
  • [6] Evaluation of the Bio-Rad Variant II® analyzer for hemoglobin A1c measurement
    Maquart, FX
    Leroy, N
    Volondat, A
    Gillery, P
    ANNALES DE BIOLOGIE CLINIQUE, 2000, 58 (02) : 199 - 203
  • [7] Determination of HbA1c in the presence of unusual hemoglobin variants:: A comparison of results from the Bayer DCA-2000 and the Bio-Rad Variant II.
    Lloyd, Kelly A.
    Doersch, Marcia
    Mamiya, Linda Y.
    Sabath, Daniel E.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2006, 126 (03) : 473 - 473
  • [8] Falsely elevated hemoglobin A1c due to S-β+-thalassemia interference in Bio-Rad Variant II Turbo HbA1c assay
    Zhu, Yusheng
    Williams, Laurie M.
    CLINICA CHIMICA ACTA, 2009, 409 (1-2) : 18 - 20
  • [9] Evaluation of Variant II® analyzer equipped with the new 270-2101 NU kit (Bio-Rad) for HbA1c assay
    Lavalard, E.
    Szymezak, J.
    Leroy, N.
    Gillery, P.
    ANNALES DE BIOLOGIE CLINIQUE, 2009, 67 (01) : 55 - 65
  • [10] Incidental Detection of Hemoglobin Variants During Evaluation of HbA1c
    Kulkarni, Jayashree D.
    Shivashanker, Sweta
    INDIAN JOURNAL OF CLINICAL BIOCHEMISTRY, 2022, 37 (02) : 242 - 246