Necessity of Hysterosalpingography After Essure Microinsert Placement for Contraception

被引:9
|
作者
Lazarus, Elizabeth [1 ]
Lourenco, Ana P. [1 ]
Casper, Susan [1 ]
Allen, Rebecca H. [2 ]
机构
[1] Brown Univ, Alpert Med Sch, Dept Diagnost Imaging, Women & Infants Hosp, Providence, RI 02903 USA
[2] Brown Univ, Sch Med, Dept Obstet Gynecol, Women & Infants Hosp, Providence, RI 02903 USA
关键词
fallopian tubes; hysterosalpingography; sterilization;
D O I
10.2214/AJR.11.7667
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBECTIVE. The objective of our study was to determine whether hysterosalpingography is necessary after Essure microinsert placement by evaluating the rates of appropriate placement and of satisfactory tubal occlusion in a general population. MATERIALS AND METHODS. We identified all patients who underwent hysterosalpingography after Essure microinsert placement for desired sterility between January 1, 2008, and August 1, 2010. We recorded demographic information and the hysterosalpingographic results. The images and operative reports of all cases with abnormal hysterosalpingographic findings were reviewed. The hysterosalpingographic results were reviewed for appropriate placement of the microinsert, for successful tubal occlusion, and for any additional abnormalities. We also reviewed the medical records for documentation of subsequent pregnancies. RESULTS. Two hundred forty hysterosalpingographic examinations were performed after 237 hysteroscopic microinsert placement procedures in 235 women. The mean age of the subjects was 35 years (range, 20-50 years). Twenty-two examinations (9.2%) were abnormal. Fourteen (5.8%) revealed inappropriate placement: six with tubal occlusion, seven without tubal occlusion, and one with an indeterminate finding for tubal occlusion. Fifteen examinations (6.3%) showed tubal nonocclusion: Microinsert placement was inappropriate in seven cases and appropriate in eight. Of the 22 abnormal hysterosalpingographic examinations, 20 had operative reports available. Eleven (55%) described difficulties with device insertion. Forty-two endometrial abnormalities were described in hysterosalpingographic reports of 38 patients. One subsequent pregnancy was documented in a patient with satisfactory device placement and tubal occlusion on hysterosalpingography. CONCLUSION. Hysterosalpingography after Essure microinsert placement is necessary because 6.3% of examinations showed abnormalities requiring an alternative form of contraception.
引用
收藏
页码:1460 / 1463
页数:4
相关论文
共 50 条
  • [1] Hysterosalpingography for assessing efficacy of essure microinsert permanent birth control device
    Wittmer, Michael H.
    Famuyide, Abimbola O.
    Creedon, Douglas J.
    Hartman, Robert P.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (04) : 955 - 958
  • [2] Contrast infusion sonography to assess microinsert placement and tubal occlusion after Essure®
    Connor, VF
    [J]. FERTILITY AND STERILITY, 2006, 85 (06) : 1791 - 1793
  • [3] Dislodged Essure microinsert
    Booker, Corenthian J.
    Yarwood, Robert L.
    Dodson, William C.
    [J]. FERTILITY AND STERILITY, 2008, 89 (04) : 964 - 965
  • [4] Tubal perforation by ESSURE microinsert
    Thoma, V
    Chua, I
    Garbin, O
    Hummel, M
    Wattiez, A
    [J]. JOURNAL OF MINIMALLY INVASIVE GYNECOLOGY, 2006, 13 (02) : 161 - 163
  • [5] Surgical aspects of removal of Essure microinsert
    Albright, Catherine M.
    Frishman, Gary N.
    Bhagavath, Bala
    [J]. CONTRACEPTION, 2013, 88 (03) : 334 - 336
  • [6] Computed tomographic confirmation of distally placed Essure* microinsert
    Palmer, Andrea B.
    Siegle, John C.
    [J]. FERTILITY AND STERILITY, 2009, 91 (04) : 1294.e5 - 1294.e7
  • [7] Analysis of Tubal Patency After Essure Placement
    Rodriguez, Ana M.
    Kilic, Gokhan S.
    Vu, Thao P.
    Kuo, Yong-Fang
    Breitkopf, Daniel
    Snyder, Russell R.
    [J]. JOURNAL OF MINIMALLY INVASIVE GYNECOLOGY, 2013, 20 (04) : 468 - 472
  • [8] Confirmation of Essure® microinsert tubal coil placement with conventional and volume-contrast imaging three-dimensional ultrasound
    Thiel, JA
    Suchet, IB
    Lortie, K
    [J]. FERTILITY AND STERILITY, 2005, 84 (02) : 504 - 508
  • [9] Risk of Essure microinsert abdominal migration: case report and review of literature
    Ricci, Giuseppe
    Restaino, Stefano
    Di Lorenzo, Giovanni
    Fanfani, Francesco
    Scrimin, Federica
    Mangino, Francesco P.
    [J]. THERAPEUTICS AND CLINICAL RISK MANAGEMENT, 2014, 10 : 963 - 968
  • [10] Pregnancy after Essure® placement: report of two cases
    Moses, A. Whitney
    Burgis, Judith T.
    Bacon, Janice L.
    Risinger, Jennifer
    [J]. FERTILITY AND STERILITY, 2008, 89 (03) : 724.e9 - 724.e11