Electroanalytical detection of pindolol: comparison of unmodified and reduced graphene oxide modified screen-printed graphite electrodes

被引:37
|
作者
Cumba, Loanda R. [1 ,2 ]
Smith, Jamie P. [1 ]
Brownson, Dale A. C. [1 ]
Iniesta, Jesus [3 ,4 ]
Metters, Jonathan P. [1 ]
do Carmo, Devaney R. [2 ]
Banks, Craig E. [1 ]
机构
[1] Manchester Metropolitan Univ, Div Chem & Environm Sci, Sch Chem & Environm, Fac Sci & Engn, Manchester M1 5GD, Lancs, England
[2] Univ Estadual Paulista, UNESP, Fac Engn Ilha Solteira, Dept Quim & Fis, BR-15385000 Solteira, SP, Brazil
[3] Univ Alicante, Dept Phys Chem, Alicante 03690, Spain
[4] Univ Alicante, Inst Electrochem, Alicante 03690, Spain
关键词
ABSORPTION SPECTROMETRIC DETERMINATION; PERFORMANCE LIQUID-CHROMATOGRAPHY; SPECTROPHOTOMETRIC DETERMINATION; FORENSIC ELECTROCHEMISTRY; URIC-ACID; SOLUBILITY; PROPRANOLOL; ENANTIOMERS; SENSORS; SERUM;
D O I
10.1039/c4an02005g
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Recent work has reported the first electroanalytical detection of pindolol using reduced graphene oxide (RGO) modified glassy carbon electrodes [S. Smarzewska and W. Ciesielski, Anal. Methods, 2014, 6, 5038] where it was reported that the use of RGO provided significant improvements in the electroanalytical signal in comparison to a bare (unmodified) glassy carbon electrode. We demonstrate, for the first time, that the electroanalytical quantification of pindolol is actually possible using bare (unmodified) screen-printed graphite electrodes (SPEs). This paper addresses the electroanalytical determination of pindolol utilising RGO modified SPEs. Surprisingly, it is found that bare (unmodified) SPEs provide superior electrochemical signatures over that of RGO modified SPEs. Consequently the electroanalytical sensing of pindolol is explored at bare unmodified SPEs where a linear range between 0.1 mu M-10.0 mu M is found to be possible whilst offering a limit of detection (3 sigma) corresponding to 0.097 mu M. This provides a convenient yet analytically sensitive method for sensing pindolol. The optimised electroanalytical protocol using the unmodified SPEs, which requires no pre-treatment (electrode polishing) or electrode modification step (such as with the use of RGO), was then further applied to the determination of pindolol in urine samples. This work demonstrates that the use of RGO modified SPEs have no significant benefits when compared to the bare (unmodified) alternative and that the RGO free electrode surface can provide electro-analytically useful performances.
引用
收藏
页码:1543 / 1550
页数:8
相关论文
共 50 条
  • [1] Graphene Oxide Bulk-Modified Screen-Printed Electrodes Provide Beneficial Electroanalytical Sensing Capabilities
    Rowley-Neale, Samuel J.
    Brownson, Dale A. C.
    Smith, Graham
    Banks, Craig E.
    BIOSENSORS-BASEL, 2020, 10 (03):
  • [2] Graphene oxide bulk-modified screen-printed electrodes provide beneficial electroanalytical sensing capabilities
    Rowley-Neale, Samuel J.
    Brownson, Dale A.C.
    Smith, Graham
    Banks, Craig E.
    Biosensors, 2020, 10 (03):
  • [3] Graphene Quantum Dots Modified Screen-printed Electrodes as Electroanalytical Sensing Platform for Diethylstilbestrol
    Gevaerd, Ava
    Banks, Craig E.
    Bergamini, Marcio F.
    Marcolino-Junior, Luiz Humberto
    ELECTROANALYSIS, 2019, 31 (05) : 838 - 843
  • [4] Electrochemically Reduced Graphene Oxide Modified Screen-Printed Electrodes for Sensitive Determination of Acetylsalicylic Acid
    Zhao, Chunzhi
    Ling, Jihong
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2017, 12 (11): : 10177 - 10186
  • [5] Fat taste detection with odorant-binding proteins (OBPs) on screen-printed electrodes modified by reduced graphene oxide
    Lu, Yanli
    Huang, Yixuan
    Li, Shuang
    Zhang, Qian
    Wu, Jiajia
    Xiong, Zhongyan
    Xiong, Lingxi
    Wan, Qianqian
    Liu, Qingjun
    SENSORS AND ACTUATORS B-CHEMICAL, 2017, 252 : 973 - 982
  • [6] Biosensors Based on Graphene Modified Screen-Printed Electrodes for The Detection of Catecholamines
    Apetrei, I. M.
    Popa , C. V.
    Apetrei, C.
    Tutunaru, D.
    ROMANIAN BIOTECHNOLOGICAL LETTERS, 2014, 19 (05): : 9801 - 9809
  • [7] Carbon Black-Modified Screen-Printed Electrodes as Electroanalytical Tools
    Arduini, Fabiana
    Di Nardo, Fabio
    Amine, Aziz
    Micheli, Laura
    Palleschi, Giuseppe
    Moscone, Danila
    ELECTROANALYSIS, 2012, 24 (04) : 743 - 751
  • [8] Electrocatalytic Oxidation of NADH on Graphene Oxide and Reduced Graphene Oxide Modified Screen-Printed Electrode
    Zhang, Lei
    Li, Yang
    Zhang, Li
    Li, Da-Wei
    Karpuzov, Dimitre
    Long, Yi-Tao
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2011, 6 (03): : 819 - 829
  • [9] An Electrochemical Enzyme Biosensor for 3-Hydroxybutyrate Detection Using Screen-Printed Electrodes Modified by Reduced Graphene Oxide and Thionine
    Martinez-Garcia, Gonzalo
    Perez-Julian, Elena
    Agui, Lourdes
    Cabre, Naomi
    Joven, Jorge
    Yanez-Sedeno, Paloma
    Manuel Pingarron, Jose
    BIOSENSORS-BASEL, 2017, 7 (04):
  • [10] Comparative immobilization of antibodies on modified screen-printed graphite electrodes
    Presnova G.V.
    Rubtsova M.Yu.
    Shumyantseva V.V.
    Bulko T.V.
    Egorov A.M.
    Moscow University Chemistry Bulletin, 2008, 63 (2) : 71 - 74