The GATE-RTion/IDEAL Independent Dose Calculation System for Light Ion Beam Therapy

被引:10
|
作者
Grevillot, L. [1 ]
Boersma, D. J. [1 ,2 ]
Fuchs, H. [1 ,3 ]
Bolsa-Ferruz, M. [1 ,3 ]
Scheuchenpflug, L. [1 ,4 ]
Georg, D. [3 ]
Kronreif, G. [2 ]
Stock, M. [1 ]
机构
[1] MedAustron Ion Therapy Ctr, Wiener Neustadt, Austria
[2] ACMIT Gmbh, Wiener Neustadt, Austria
[3] Med Univ Vienna, Dept Radiat Oncol, Vienna, Austria
[4] Univ Vienna, Fac Phys, Isotope Phys, Vienna, Austria
关键词
GATE; GEANT4; independent dose caculation; proton; carbon ion; light ion beam therapy; IDEAL; GATE-RTion; MONTE-CARLO CODE; QUALITY-ASSURANCE; PROTON; SOFTWARE; GEANT4; BENCHMARKING; SIMULATIONS; VALIDATION; PLATFORM; CT;
D O I
10.3389/fphy.2021.704760
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Patient specific quality assurance can be improved using an independent dose calculation system. In addition, the implementation of such a system may support light ion beam therapy facilities in reducing the needs for beam time, by substituting some of the experimental patient-specific quality assurance procedures by independent dose calculation. The GATE-RTion-based IDEAL system for light ion beam therapy was developed for this purpose. It was built in a DICOM-in, DICOM-out fashion, for easy integration into a state-of-the-art technology-based workflow for scanned ion beam therapy. This article describes the IDEAL system, followed by its clinical implementation at MedAustron for proton and carbon ion beams. Medical physics acceptance and commissioning steps are presented together with key results: for 3D proton and carbon ion reference boxes, 97% of the points agreed within 5% from the measurements. Experimental validation of stopping powers using real pig samples were between 1.8% and 3.8% for soft tissues. Finally, five clinical cases are described, i.e. two proton and three carbon ion treatments. Dosimetric benchmarking against TPS calculations are presented and discussed in details. As expected, the IDEAL software evidenced limitations arising from the pencil beam algorithm available in the TPS for carbon ions, especially in the presence of air cavities. The IDEAL system was found to satisfy the clinical requirements for independent dose calculation of scanned ion beam delivery systems and is being clinically implemented at MedAustron. The open-source code as well as the documentation was released on the OpenGATE collaboration website, thus allowing for long term maintenance and future upgrades based on a more widespread utilization.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Commissioning of IDEAL/GATE-RTion for Proton and Carbon ion Independent Dose Calculation (IDC)
    Grevillot, L.
    Boersma, D.
    Gleyzes, R. Gonzalo
    Scheuchenpflug, L.
    Carlino, A.
    Elia, A.
    Fuchs, H.
    Stock, M.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S109 - S110
  • [2] Technical Note: GATE-RTion: a GATE/Geant4 release for clinical applications in scanned ion beam therapy
    Grevillot, L.
    Boersma, D. J.
    Fuchs, H.
    Aitkenhead, A.
    Elia, A.
    Bolsa, M.
    Winterhalter, C.
    Vidal, M.
    Jan, S.
    Pietrzyk, U.
    Maigne, L.
    Sarrut, D.
    MEDICAL PHYSICS, 2020, 47 (08) : 3675 - 3681
  • [3] Monte Carlo computation of 3D distributions of stopping power ratios in light ion beam therapy using GATE-RTion
    Bolsa-Ferruz, Marta
    Palmans, Hugo
    Boersma, David
    Stock, Markus
    Grevillot, Loic
    MEDICAL PHYSICS, 2021, 48 (05) : 2580 - 2591
  • [4] Evaluation of GATE-RTion (GATE/Geant4) Monte Carlo simulation settings for proton pencil beam scanning quality assurance
    Winterhalter, Carla
    Taylor, Michael
    Boersma, David
    Elia, Alessio
    Guatelli, Susanna
    Mackay, Ranald
    Kirkby, Karen
    Maigne, Lydia
    Ivanchenko, Vladimir
    Resch, Andreas F.
    Sarrut, David
    Sitch, Peter
    Vidal, Marie
    Grevillot, Loic
    Aitkenhead, Adam
    MEDICAL PHYSICS, 2020, 47 (11) : 5817 - 5828
  • [5] Dose calculation on megavoltage CT in carbon ion beam therapy
    Ecker, S.
    Ellerbrock, M.
    Ackermann, B.
    Lissner, S.
    Schubert, K.
    Herfarth, K.
    Jaekel, O.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (10) : 681 - 681
  • [6] Fast optimization and dose calculation in scanned ion beam therapy
    Hild, S.
    Graeff, C.
    Trautmann, J.
    Kraemer, M.
    Zink, K.
    Durante, M.
    Bert, C.
    MEDICAL PHYSICS, 2014, 41 (07)
  • [7] A PENCIL BEAM ALGORITHM FOR HELIUM ION BEAM DOSE CALCULATION
    Fuchs, H.
    Stroebele, J.
    Schreiner, T.
    Georg, D.
    RADIOTHERAPY AND ONCOLOGY, 2012, 102 : S39 - S40
  • [8] FLUKA particle therapy tool for Monte Carlo independent calculation of scanned proton and carbon ion beam therapy
    Kozlowska, Wioletta S.
    Bohlen, Till T.
    Cuccagna, Caterina
    Ferrari, Alfredo
    Fracchiolla, Francesco
    Magro, Giuseppe
    Mairani, Andrea
    Schwarz, Marco
    Vlachoudis, Vasilis
    Georg, Dietmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (07):
  • [9] A COMPUTER APPROACH TO DOSE CALCULATION FOR SUPPLEMENTARY BEAM THERAPY
    JAMESON, DG
    TREVELYAN, A
    BRITISH JOURNAL OF RADIOLOGY, 1969, 42 (493): : 57 - +
  • [10] A new convolution/superposition dose calculation method for external photon beam radiation therapy using beam modifiers and/or independent collimators
    Liu, HH
    MEDICAL PHYSICS, 1997, 24 (10) : 1647 - 1647