Two-photon fluorescence microscopy signal formation in highly scattering media: theoretical and numerical simulation

被引:8
|
作者
Sergeeva, E. A. [1 ]
Katichev, A. R. [1 ]
Kirillin, M. Yu. [1 ]
机构
[1] Russian Acad Sci, Inst Appl Phys, Nizhnii Novgorod 603950, Russia
基金
俄罗斯基础研究基金会;
关键词
two-photon fluorescence microscopy; multiple scattering; radiative transfer theory; small-angle diffusion approximation; Monte Carlo simulation; focused Gaussian beam; IMAGING-DEPTH LIMIT; MONTE-CARLO; EXCITATION FLUORESCENCE; OPTICAL-PROPERTIES; PHASE FUNCTIONS; LIGHT; COLLECTION; SECTIONS;
D O I
10.1070/QE2010v040n12ABEH014506
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Using the radiative transfer theory and Monte Carlo simulations, we analyse the effect of scattering in a medium and of the size of the detector pinhole on the formation of the fluorescent signal in standard two-photon fluorescence microscopy (TPFM) systems. The theoretical analysis is based on a small-angle diffusion approximation of the radiative transfer equation, adapted to calculate the propagation of focused infrared radiation in media similar to the biological tissues in their optical properties. The accuracy of the model is evaluated by comparing the calculated excitation intensity in a highly scattering medium with the results of Monte Carlo simulations. To simulate a tightly focused Gaussian beam by the Monte Carlo method, the so called 'ray-optics' approach that correctly takes into account the finite size and shape of the beam waist is applied. It is shown that in the combined confocal and two-photon scanning microscopy systems not equipped with an external 'nondescanned' detector, the scattering significantly affects both the nonlinear excitation efficiency in the medium and the fluorescence collection efficiency of the system. In such systems, the rate of the useful TPFM signal in-depth decay is 1.5-2 times higher than in systems equipped with a 'nondescanned' detector.
引用
收藏
页码:1053 / 1061
页数:9
相关论文
共 50 条
  • [1] Out-of-focus fluorescence collection in two-photon microscopy of scattering media
    Leray, A.
    Odin, C.
    Le Grand, Y.
    OPTICS COMMUNICATIONS, 2008, 281 (24) : 6139 - 6144
  • [2] Highly nonlinear photodamage in two-photon fluorescence microscopy
    Hopt, A
    Neher, E
    BIOPHYSICAL JOURNAL, 2001, 80 (04) : 2029 - 2036
  • [3] Two-photon fluorescence microscopy imaging of biotissue: a theoretical model
    Sergeeva, EA
    SARATOV FALL MEETING 2002: OPTICAL TECHNOLOGIES IN BIOPHYSICS AND MEDICINE IV, 2002, 5068 : 10 - 15
  • [4] Spatial distribution of single-photon and two-photon fluorescence light in scattering media: Monte Carlo simulation
    Gan, XS
    Gu, M
    APPLIED OPTICS, 2000, 39 (10) : 1575 - 1579
  • [5] Theoretical model of two-photon fluorescence microscopy in a layered turbid medium
    Sergeeva, EA
    Dolin, LS
    Korytin, AI
    MULTIPHOTON MICROSCOPY IN THE BIOMEDICAL SCIENCES III, 2003, 4963 : 143 - 150
  • [6] Two-photon excitation fluorescence microscopy
    So, PTC
    Dong, CY
    Masters, BR
    Berland, KM
    ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2000, 2 : 399 - 429
  • [7] Scattering effect on the imaging depth limit in two-photon fluorescence microscopy
    Sergeeva, E. A.
    QUANTUM ELECTRONICS, 2010, 40 (05) : 411 - 417
  • [8] Two-photon fluorescence microscopy: High resolution imaging in scattering tissue
    Denk, W
    EUROPEAN JOURNAL OF NEUROSCIENCE, 1998, 10 : 267 - 267
  • [9] Measuring the scattering coefficient of turbid media from two-photon microscopy
    Sevrain, David
    Dubreuil, Matthieu
    Leray, Aymeric
    Odin, Christophe
    Le Grand, Yann
    OPTICS EXPRESS, 2013, 21 (21): : 25221 - 25235
  • [10] Adaptive optics two-photon fluorescence microscopy
    Zhou, Yaopeng
    Bifano, Thomas
    Lin, Charles
    MEMS ADAPTIVE OPTICS, 2007, 6467