Patient self-management of oral anticoagulation and external quality assessment procedures

被引:21
|
作者
Murray, ET [1 ]
Kitchen, DP
Kitchen, S
Jennings, I
Woods, TAL
Preston, FE
Fitzmaurice, DA
机构
[1] Univ Birmingham, Sch Med, Dept Primary Care & Gen Practice, Birmingham B15 2TT, W Midlands, England
[2] UK NEQAS Blood Coagulat, Sheffield, S Yorkshire, England
关键词
oral anticoagulation; patient self-management; external quality assessment; training; supervision;
D O I
10.1046/j.1365-2141.2003.04501.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The role of external quality assessment (EQA) is a contentious issue for patient self-management (PSM) of oral anticoagulation. Patients from general practices in the West Midlands undertaking PSM were recruited to compare efficacy of patients' and health professionals' EQA procedure using the UK National External Quality Assessment Scheme (NEQAS). Patients using Coaguchek (Roche Diagnostics) were trained to perform EQA as part of their PSM training. They undertook PSM for 26 weeks and were asked to perform EQA using material provided by the UK NEQAS twice at home without supervision and twice at the practice with supervision. Patients' results were compared with health care professional users of Coaguchek S. Twenty-three PSM patients were compared with 75 health care professional users of the NEQAS scheme. The PSM group international normalized ratio (INR) percentage time in range was 74%. There was no significant difference in the median results on NEQAS samples obtained by the patients and those obtained by professionals. Three patients were outwith consensus (results > 15% from the median INR) on more than one occasion. Patients were able to perform the EQA tests competently. The data show that good agreement can be achieved between patients analysing the same EQA samples, with coefficients of variation ranging from 22.3% to as low as 5.4%. Further study is required to determine how precision within these EQA schemes relates to the stability of treatment in patients' management of their own anticoagulation.
引用
收藏
页码:825 / 828
页数:4
相关论文
共 50 条
  • [1] Patient self-management of oral anticoagulation
    Li, QH
    Tosiello, L
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 281 (24): : 2283 - 2284
  • [2] Patient self-management of oral anticoagulation
    Bhavnani, M
    Shiach, CR
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 2002, 24 (04): : 253 - 257
  • [3] Patient self-management of oral anticoagulation
    Wheeler, BM
    [J]. ANNALS OF INTERNAL MEDICINE, 2005, 142 (12) : 1023 - 1023
  • [4] An evaluation of four methods of external quality assurance (EQA) for patient self-management of oral anticoagulation
    Murray, ET
    Fitzmaurice, DA
    Kitchen, D
    Jennings, I
    McCahon, D
    Kitchen, S
    Woods, T
    Preston, FE
    Hobbs, FDR
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2005, 129 : 12 - 12
  • [5] Patient self-management of oral anticoagulation - Reply
    Sawicki, PT
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1999, 281 (24): : 2284 - 2284
  • [6] Patient self-management of oral anticoagulation: A review
    Sunderji, R
    Fung, A
    Gin, K
    Shalansky, K
    Carter, C
    [J]. CANADIAN JOURNAL OF CARDIOLOGY, 2003, 19 (08) : 931 - 935
  • [7] Guidelines for "patient self-management of oral anticoagulation"
    Bernardo, A
    Völler, H
    [J]. DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2001, 126 (12) : 346 - 351
  • [8] Patient self-testing and self-management of oral anticoagulation
    Siegmund Braun
    Michael Spannagl
    Heinz Völler
    [J]. Analytical and Bioanalytical Chemistry, 2009, 393
  • [9] Patient self-testing and self-management of oral anticoagulation
    Braun, Siegmund
    Spannagl, Michael
    Voeller, Heinz
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 393 (05) : 1463 - 1471
  • [10] The quality of oral anticoagulation before, during and after a period of patient self-management
    Piso, B
    Jimenz-Boj, E
    Krinninger, B
    Watzke, HH
    [J]. THROMBOSIS RESEARCH, 2002, 106 (02) : 101 - 104