HPLC determination of nimesulide in tablets by electrochemical detection

被引:0
|
作者
Alvarez-Lueje, A [1 ]
Vásquez, P [1 ]
Núñez-Vergara, LJ [1 ]
Squella, JA [1 ]
机构
[1] Univ Chile, Chem & Pharmaceut Sci Fac, Bioelectrochem Lab, Santiago 1, Chile
关键词
nimesulide; HPLC; electrochemical detection; pharmaceuticals;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical chromatographic method for the determination of nimesulide in pharmaceutical forms has been developed. The method is based on high performance liquid chromatographic (HPLC) and electrochemical detection. Chromatography was performed on a mu Bondapak/mu Porasil C-18 column (150 mm(x) 3.9 mm I.D.). The mobile phase consisted of pH 3 buffer phosphate-methanol (40:60, v/v) at a flow-rate of 1 mL/min The analytes were detected by electrochemical detection in the pulse mode at 1200 mV. For comparative purposes both an HPLC with diode array detector and a spectrophotometric method were employed. The results of the recovery study (mean 101.07, rsd 1.44) show that HPLC with an electrochemical detection method is adequately precise and accurate and can be recommended for the determination of nimesulide in tablets.