ADAPTIVE FEEDBACK CANCELLATION IN HEARING AIDS USING A SINUSOIDAL NEAR-END SIGNAL MODEL

被引:10
|
作者
Ngo, Kim [1 ]
van Waterschoot, Toon [1 ]
Christensen, Mads Graesboll [2 ]
Moonen, Marc [1 ]
Jensen, Soren Holdt [2 ]
Wouters, Jan [3 ]
机构
[1] Katholieke Univ Leuven, ESAT SCD, Kasteelpk Arenberg 10, B-3001 Louvain, Belgium
[2] Aalborg Univ, Dept Elect Syst, DK-9220 Aalborg, Denmark
[3] Katholieke Univ Leuven, ExpORL O & N2, B-3000 Louvain, Belgium
关键词
Adaptive Feedback Cancellation; Frequency Estimation; Decorrelation; Hearing Aids;
D O I
10.1109/ICASSP.2010.5496063
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Acoustic feedback is a well-known problem in hearing aids, which is caused by the undesired acoustic coupling between the loudspeaker and the microphone. Acoustic feedback limits the maximum amplification that can be used in the hearing aid without making it unstable. The goal of adaptive feedback cancellation (AFC) is to adaptively model the feedback path and estimate the feedback signal, which is then subtracted from the microphone signal. The main problem in identifying the feedback path model is the correlation between the near-end signal and the loudspeaker signal, which is caused by the closed signal loop. A possible solution to this problem is to use the prediction error method (PEM)-based AFC with a linear prediction (LP) model for the near-end signal. In this paper, a modification to the PEM-based AFC is presented where the LP model is replaced by a sinusoidal near-end signal model. More specifically, it is shown that using frequency estimation techniques to estimate the sinusoidal near-end signal model improves the performance of the PEM-based AFC compared to using a LP model. Simulation results for a hearing aid scenario indicate a significant improvement in terms of misadjustment and maximum stable gain increase.
引用
收藏
页码:181 / 184
页数:4
相关论文
共 50 条
  • [1] Adaptive feedback cancellation for audio signals using a warped all-pole near-end signal model
    van Waterschoot, Toon
    Moonen, Marc
    2008 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH AND SIGNAL PROCESSING, VOLS 1-12, 2008, : 269 - 272
  • [2] Analysis of an adaptive algorithm for feedback cancellation in hearing aids for sinusoidal signals
    Sakai, Hideaki
    2007 EUROPEAN CONFERENCE ON CIRCUIT THEORY AND DESIGN, VOLS 1-3, 2007, : 416 - 419
  • [3] Adaptive feedback cancellation in hearing aids
    Spriet, A.
    Rombouts, G.
    Moonen, M.
    Wouters, J.
    JOURNAL OF THE FRANKLIN INSTITUTE-ENGINEERING AND APPLIED MATHEMATICS, 2006, 343 (06): : 545 - 573
  • [4] Adaptive feedback cancellation in hearing aids with linear prediction of the desired signal
    Spriet, A
    Proudler, I
    Moonen, M
    Wouters, J
    IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2005, 53 (10) : 3749 - 3763
  • [5] ADAPTIVE FEEDBACK CANCELLATION IN HEARING AIDS USING THE IPLS ALGORITHM
    Plate, Randall S.
    Wang, Yongchao
    Luo, Zhi-Quan
    Gao, Chris
    2010 IEEE INTERNATIONAL CONFERENCE ON ACOUSTICS, SPEECH, AND SIGNAL PROCESSING, 2010, : 177 - 180
  • [6] An efficient adaptive feedback cancellation for hearing aids
    Park, YC
    Kim, IY
    Lee, SM
    PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 1647 - 1650
  • [7] An efficient adaptive feedback cancellation for hearing aids
    Lee, SM
    Kim, IY
    Park, YC
    IEICE TRANSACTIONS ON FUNDAMENTALS OF ELECTRONICS COMMUNICATIONS AND COMPUTER SCIENCES, 2005, E88A (09) : 2446 - 2450
  • [8] Adaptive feedback cancellation in hearing aids with clipping in the feedback path
    Freed, Daniel J.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2008, 123 (03): : 1618 - 1626
  • [9] Acoustic Feedback Noise Cancellation in Hearing Aids Using Adaptive Filter
    Kumar, D. N. S. Ravi
    Madhavi, K. Reddy
    Praveen, N.
    Kumar, Hari Hara P.
    Srinivas, Ganganagunta
    Raju, M., V
    INTERNATIONAL JOURNAL OF INTEGRATED ENGINEERING, 2022, 14 (07): : 45 - 55
  • [10] Using a signal cancellation technique to assess adaptive directivity of hearing aids
    Wu, Yu-Hsiang
    Bentler, Ruth A.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2007, 122 (01): : 496 - 511