The impact of dose calculation algorithms on partial and whole breast radiation treatment plans

被引:9
|
作者
Basran, Parminder S. [1 ,2 ]
Zavgorodni, Sergei [1 ,2 ]
Berrang, Tanya [3 ,4 ]
Olivotto, Ivo A. [3 ,4 ]
Beckham, Wayne [1 ,2 ]
机构
[1] BC Canc Agcy Vancouver Isl Ctr, Dept Med Phys, Victoria, BC, Canada
[2] Univ Victoria, Dept Phys & Astron, Victoria, BC, Canada
[3] BC Canc Agcy Vancouver Isl Ctr, Dept Radiat Oncol, Victoria, BC, Canada
[4] Univ British Columbia, Dept Surg, Vancouver, BC V6T 1W5, Canada
关键词
MONTE-CARLO SIMULATIONS; IRRADIATION; RADIOTHERAPY; BEAMS;
D O I
10.1186/1748-717X-5-120
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: This paper compares the calculated dose to target and normal tissues when using pencil beam (PBC), superposition/convolution (AAA) and Monte Carlo (MC) algorithms for whole breast (WBI) and accelerated partial breast irradiation (APBI) treatment plans. Methods: Plans for 10 patients who met all dosimetry constraints on a prospective APBI protocol when using PBC calculations were recomputed with AAA and MC, keeping the monitor units and beam angles fixed. Similar calculations were performed for WBI plans on the same patients. Doses to target and normal tissue volumes were tested for significance using the paired Student's t-test. Results: For WBI plans the average dose to target volumes when using PBC calculations was not significantly different than AAA calculations, the average PBC dose to the ipsilateral breast was 10.5% higher than the AAA calculations and the average MC dose to the ipsilateral breast was 11.8% lower than the PBC calculations. For ABPI plans there were no differences in dose to the planning target volume, ipsilateral breast, heart, ipsilateral lung, or contra-lateral lung. Although not significant, the maximum PBC dose to the contra-lateral breast was 1.9% higher than AAA and the PBC dose to the clinical target volume was 2.1% higher than AAA. When WBI technique is switched to APBI, there was significant reduction in dose to the ipsilateral breast when using PBC, a significant reduction in dose to the ipsilateral lung when using AAA, and a significant reduction in dose to the ipsilateral breast and lung and contra-lateral lung when using MC. Conclusions: There is very good agreement between PBC, AAA and MC for all target and most normal tissues when treating with APBI and WBI and most of the differences in doses to target and normal tissues are not clinically significant. However, a commonly used dosimetry constraint, as recommended by the ASTRO consensus document for APBI, that no point in the contra-lateral breast volume should receive >3% of the prescribed dose needs to be relaxed to >5%.
引用
下载
收藏
页数:9
相关论文
共 50 条
  • [1] The impact of dose calculation algorithms on partial and whole breast radiation treatment plans
    Parminder S Basran
    Sergei Zavgorodni
    Tanya Berrang
    Ivo A Olivotto
    Wayne Beckham
    Radiation Oncology, 5
  • [2] Evaluation of tangential breast treatment plans using two different dose calculation algorithms
    King, S
    Jeswani, S
    Krasin, MJ
    Dogan, N
    Sethi, A
    Emami, B
    RADIOLOGY, 1999, 213P : 201 - 201
  • [3] Comparison of Dose Calculation Algorithms for Accelerated Partial Breast Intensity Modulated Radiation Therapy
    Kim, H.
    Kim, J. O.
    Peskorski, D.
    Vargo, J. A. A.
    Barry, P. N., Jr.
    Huq, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E578 - E579
  • [4] CALCULATION OF LUNG NTCP VALUES FOR BREAST CANCER TREATMENT PLANS USING TWO DIFFERENT DOSE CALCULATION ALGORITHMS
    Sassowsky, M.
    Loessl, K.
    Messer, P.
    Aebersold, D.
    Mini, R.
    Born, E.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (08) : 521 - 521
  • [5] LUNG DOSE CALCULATIONS FOR BREAST TREATMENT: IMPACT OF DOSE CALCULATION ALGORITHMS IN DIFFERENT RESPIRATORY PHASES
    de'Palma, E. Vanetti
    Fogliata, A.
    Nicolini, G.
    Clivio, A.
    Winkler, P.
    Cozzi, L.
    RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S35 - S35
  • [6] Radiobiological impact of dose calculation algorithms on biologically optimized IMRT lung stereotactic body radiation therapy plans
    X. Liang
    J. Penagaricano
    D. Zheng
    S. Morrill
    X. Zhang
    P. Corry
    R. J. Griffin
    E. Y. Han
    M. Hardee
    V. Ratanatharathom
    Radiation Oncology, 11
  • [7] Radiobiological impact of dose calculation algorithms on biologically optimized IMRT lung stereotactic body radiation therapy plans
    Liang, X.
    Penagaricano, J.
    Zheng, D.
    Morrill, S.
    Zhang, X.
    Corry, P.
    Griffin, R. J.
    Han, E. Y.
    Hardee, M.
    Ratanatharathom, V.
    RADIATION ONCOLOGY, 2016, 11
  • [8] Investigating the Impact of the AXB Dose Calculation Algorithm in Tangential Breast Plans
    Kejda, A.
    Nahar, N.
    Wang, T.
    Ashworth, S.
    Sykes, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E302 - E302
  • [9] Impact of Dose Calculation Algorithms On Biologically Optimized VMAT Plans for Esophageal Cancer
    Thiyagarajan, Rajesh
    Nambiraj, N. Arunai
    Vikraman, S.
    Karrthick, K. P.
    Sigamani, Ashokkumar
    Subbarao, Bargavan
    Ramu, M.
    Sambasivaselli, R.
    Senniandavar, V.
    Kataria, Tejinder
    MEDICAL PHYSICS, 2015, 42 (06) : 3439 - 3439
  • [10] Impact of Dose Calculation Algorithms and Radiobiological Parameters on Prediction of Cardiopulmonary Complications in Left Breast Radiation Therapy
    Kargar N.
    Zeinali A.
    Molazadeh M.
    Journal of Biomedical Physics and Engineering, 14 (02): : 129 - 140