Determination of copper(II) in wastewater by electroplating samples using a PVC-membrane copper(II)-selective electrode

被引:81
|
作者
Zamani, H. A. [1 ]
Rajabzadeh, G.
Firouz, A.
Ganjali, M. R.
机构
[1] Islamic Azad Univ, Quchan Branch, Dept Chem, Quchan, Iran
[2] Univ Tehran, Dept Chem, Tehran, Iran
关键词
D O I
10.1134/S1061934807110135
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A PVC membrane containing 4-amino-6-methyl-1,2,4-triazin-3,5-dithione (AMTD) as a suitable ionophore, exhibits a Nernstian response for Cu(2+) ions over a wide concentration range up to 1 x 10(-1) and 1 x 10(-6) M, with a detection limit of 7.5 x 10(-7) M in the pH range 3.0-7.5. It has a fast response time (< 15 s) and can be used for at least 12 weeks without any major deviation in the potential. The electrode revealed a very good selectivity with respect to all common alkali, alkaline-earth, transition, and heavy-metal ions. It was successfully applied to the recovery of copper ions from wastewater. The electrode was also used as an indicator electrode in the potentiometric titration of Cu(II) ions with EDTA.