Time resolution of a 1.8-m long BC-420 plastic scintillator bar with metal-packaged H11934 photomultiplier tubes
被引:0
|
作者:
Ishikawa, Takatsugu
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Ishikawa, Takatsugu
[1
]
Miyabe, Manabu
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Miyabe, Manabu
[1
]
Noumi, Hiroyuki
论文数: 0引用数: 0
h-index: 0
机构:
Osaka Univ, Res Ctr Nucl Phys RCNP, Ibaraki 5670047, Japan
High Energy Accelerator Res Org KEK, Inst Particle & Nucl Studies, Tsukuba 3050801, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Noumi, Hiroyuki
[2
,3
]
Ohnishi, Hiroaki
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Ohnishi, Hiroaki
[1
]
Sada, Yuta
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Sada, Yuta
[1
]
Shirotori, Kotaro
论文数: 0引用数: 0
h-index: 0
机构:
Osaka Univ, Res Ctr Nucl Phys RCNP, Ibaraki 5670047, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Shirotori, Kotaro
[2
]
Tokiyasu, Atsushi O.
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Tokiyasu, Atsushi O.
[1
]
Yoshida, Chihiro
论文数: 0引用数: 0
h-index: 0
机构:
Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, JapanTohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
Yoshida, Chihiro
[1
]
机构:
[1] Tohoku Univ, Res Ctr Elect Photon Sci ELPH, Sendai 9820826, Japan
[2] Osaka Univ, Res Ctr Nucl Phys RCNP, Ibaraki 5670047, Japan
[3] High Energy Accelerator Res Org KEK, Inst Particle & Nucl Studies, Tsukuba 3050801, Japan
We tested a plastic-scintillator (PS) bar for the forward time-of-flight (TOF) detector of the J-PARC E50 spectrometer using a 740-MeV/c positron beam. The bar is composed of Saint-Gobain Crystals (BC-420), with a 30 mm x 40 mm cross section and 1800-mm length. Each side of the bar is connected to a metal-packaged photomultiplier tube (PMT), Hamamatsu H11934-200 with an ultra-bialkali (UBA) photoelectric surface. The PMTs have an advanced feature of a small spread of the electron transit time (0.27 ns) and high quantum efficiency (& AP; 43%). The intrinsic resolutions for the timing measurement using the two PMTs connected to both the ends are found to be 0.053 and 0.059 ns in the standard deviation for the 40-and 30-mm-thick penetrations at the center, respectively. The results show that metal-packaged PMTs enable TOF measurements using an ordinary hodoscope consisting of 1.8-m-long PS bars with an excellent time resolution (& SIM; 0.06 ns in the standard deviation).