Inter-detector scattering in BGO crystal arrays for PET

被引:0
|
作者
Mesbahi, ME [1 ]
Spyrou, NM [1 ]
机构
[1] Univ Surrey, Dept Phys, Guildford GU2 5XH, Surrey, England
来源
PHYSICS OF MEDICAL IMAGING | 1998年 / 3336卷
关键词
PET; BGO; block detectors; crystal arrays; inter-detector scattering;
D O I
10.1117/12.317089
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Positron Emission Tomography instrumentation has been directed towards improving detector spatial resolution with an emphasis on reduction of detector size. From the early days better resolution has meant ability to easily visualize and delineate regions in brain images in addition to improved accuracy in the measurement of smaller volumes of radionuclide concentration. The major difficulty in achieving both high resolution and high detection efficiency is the requirement for a small detector to fully absorb the 511 keV photon within its volume, thus minimising Compton scattered photons and also escape into neighbouring detector crystals. The latter could be recorded as true events and accepted as part of the data thus leading to mispositioning and consequently loss in spatial resolution. In addition, under certain conditions, identifying the correct pair of detectors in coincidence is ambiguous and an event can be rejected, causing a loss in the efficiency and an increase in the dead time. As individual detector crystals are becoming smaller the contribution of inter-detector scattering needs to be addressed.
引用
收藏
页码:809 / 817
页数:9
相关论文
共 50 条
  • [1] Simulation of inter-detector scattering in BGO block detector
    Alghamdi, AA
    Spyrou, NN
    Mesbahi, ME
    MEDICAL IMAGING 2004: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2004, 5368 : 898 - 905
  • [2] Recovery of inter-detector and inter-crystal scattering in brain PET based on LSO and GAGG crystals
    Lee, Seungeun
    Kim, Kyeong Yun
    Lee, Min Sun
    Lee, Jae Sung
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (19):
  • [3] Analysis of Inter-Detector Scattering in PET with Various Energy Window Settings
    Chuang, K. S.
    Wei, Z. J.
    Lin, H. H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S376 - S377
  • [4] Performance improvement of GAGG PET by inter-detector and inter-crystal scatter recovery
    Lee, Seungeun
    Kim, Kyeong Yun
    Lee, Min Sun
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [5] Comparative Evaluation of Inter-Detector Scatter Recovery Methods for Small Animal PET
    Lee, Seungeun
    Lee, Min Sun
    Kim, Kyeong Yun
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [6] A method of evaluating inter-detector scattering in block detectors for positron emission tomography
    Spyrou, NM
    Mesbahi, ME
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3): : 672 - 676
  • [7] A method of evaluating inter-detector scattering in block detectors for positron emission tomography
    Mesbahi, M.E.
    Proceedings of SPIE - The International Society for Optical Engineering, 2001, 4320 : 927 - 930
  • [8] A method of evaluating inter-detector scattering in block detectors for Positron Emission Tomography.
    Mesbahi, ME
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 927 - 930
  • [9] PET Detector Block with DOI Capabilities Based on a Large Monolithic BGO Crystal
    Gonzalez-Montoro, Andrea
    Majewski, Stan
    Sanchez, Filomeno
    Benlloch, Jose M.
    Zanettini, Silvia
    Gonzalez, Antonio J.
    2017 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2017,
  • [10] Characterization of Inter-detector Effects in a 3-D Position-Sensitive Dual-CZT Detector Modules for PET
    Gu, Yi
    Levin, Craig S.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 4088 - 4090