Micro-scale modeling of flow and oxygen transfer in hollow-fiber membrane bundle

被引:25
|
作者
Taskin, M. Ertan [1 ]
Fraser, Katharine H. [1 ]
Zhang, Tao [1 ]
Griffith, Bartley P. [1 ]
Wu, Zhongjun J. [1 ]
机构
[1] Univ Maryland, Sch Med, Dept Surg, Artificial Organs Lab, Baltimore, MD 21201 USA
基金
美国国家卫生研究院;
关键词
Artificial lung; Oxygenator; Hollow-fiber membrane; Crossflow; Micro-scale modeling; MASS-TRANSFER; MATHEMATICAL-MODEL; TRANSFER RATES; FLUID-FLOW; DESIGN; TRANSPORT;
D O I
10.1016/j.memsci.2010.06.034
中图分类号
TQ [化学工业];
学科分类号
0817 ;
摘要
The aim of this work was to develop a modeling approach to solve the flow and oxygen transfer when the blood passes through the hollow-fiber membrane bundle For this purpose, a "two-region" modeling approach was developed regarding the hollow-fiber and blood regions. The oxygen transfer in these regions was defined with separate diffusion processes. Two-dimensional single and multi-fiber geometries were created and flow solutions were obtained for a non-Newtonian fluid. The convection-diffusion-reaction equation was solved to produce the oxygen partial pressure distributions As a benefit of coupling the interstitial flow field Into the oxygen transfer through the hollow-fiber membrane bundle, the membrane resistance was taken into consideration Thus, varying oxygen partial pressures were observed on the outer fiber surface, which is contrary to the common simplifying assumptions of negligible membrane resistance and uniform oxygen content on the fiber surface (traditional approach) It was illustrated that, the current approach can be utilized to predict the mass transfer efficiencies without overestimating as compared to the predictions obtained with the traditional approach. Utilization of the current approach was found to be beneficial for the geometries with lower packing density which allows significant P-O2 variations on the fiber surfaces For the geometries with dense packings, the above simplifying assumptions could be applicable The model predictions were validated with the experimental measurements taken from a benchmark device. (C) 2010 Elsevier B V. All rights reserved
引用
收藏
页码:172 / 183
页数:12
相关论文
共 50 条
  • [1] Oxygen Transfer Model for a Flow-Through Hollow-Fiber Membrane Biofilm Reactor
    Gilmore, Kevin R.
    Little, John C.
    Smets, Barth F.
    Love, Nancy G.
    [J]. JOURNAL OF ENVIRONMENTAL ENGINEERING, 2009, 135 (09) : 806 - 814
  • [2] CFD Modeling for Water Transfer Analysis of Hollow-Fiber Membrane Humidifiers
    Choi, Yoora
    Nguyen, Xuan Linh
    Yu, Sangseok
    [J]. TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS B, 2023, 47 (04) : 187 - 194
  • [3] Modeling Ionic Strength Effects on Hollow-Fiber Nanofiltration Membrane Mass Transfer
    Yonge, David T.
    Biscardi, Paul G.
    Duranceau, Steven J.
    [J]. MEMBRANES, 2018, 8 (03)
  • [4] Computational flow modeling in hollow-fiber dialyzers
    Eloot, S
    De Wachter, D
    Van Tricht, I
    Verdonck, P
    [J]. ARTIFICIAL ORGANS, 2002, 26 (07) : 590 - 599
  • [5] Oxygen transfer characteristics of hollow-fiber, composite membranes
    Ahmed, T
    Semmens, MJ
    Voss, MA
    [J]. ADVANCES IN ENVIRONMENTAL RESEARCH, 2004, 8 (3-4): : 637 - 646
  • [6] Development of micro-scale hollow fiber ultrafiltration membranes
    Halaui, R.
    Moldavsky, A.
    Cohen, Y.
    Semiat, R.
    Zussman, E.
    [J]. JOURNAL OF MEMBRANE SCIENCE, 2011, 379 (1-2) : 370 - 377
  • [7] MODELING OF THALLIUM EXTRACTION IN A HOLLOW-FIBER MEMBRANE CONTACTOR
    Shirazian, S.
    Fadaei, F.
    Ashrafizadeh, S. N.
    [J]. SOLVENT EXTRACTION AND ION EXCHANGE, 2012, 30 (05) : 490 - 506
  • [8] Micro-scale modeling of axial flow through unidirectional disordered fiber arrays
    Chen, Xiaoming
    Papathanasiou, T. D.
    [J]. COMPOSITES SCIENCE AND TECHNOLOGY, 2007, 67 (7-8) : 1286 - 1293
  • [9] Simulation of external mass transfer in hollow-fiber membrane contactors
    V. A. Kirsch
    V. I. Roldugin
    S. D. Bazhenov
    T. V. Plisko
    [J]. Petroleum Chemistry, 2015, 55 : 776 - 782
  • [10] Simulation of external mass transfer in hollow-fiber membrane contactors
    Kirsch, V. A.
    Roldugin, V. I.
    Bazhenov, S. D.
    Plisko, T. V.
    [J]. PETROLEUM CHEMISTRY, 2015, 55 (10) : 776 - 782