Semi-automated system for measurement of nuchal translucency thickness

被引:49
|
作者
Moratalla, J. [1 ]
Pintoffl, K. [3 ]
Minekawa, R. [1 ]
Lachmann, R. [1 ]
Wright, D. [2 ]
Nicolaides, K. H. [1 ]
机构
[1] Kings Coll Hosp London, Harris Birthright Res Ctr Fetal Med, London SE5 9RS, England
[2] Univ Plymouth, Dept Math & Stat, Plymouth PL4 8AA, Devon, England
[3] GE Med Syst Kretztech GmbH & Co OHG, Zipf, Austria
关键词
automated measurement; first-trimester screening; nuchal translucency; trisomy; 21; ultrasound; MATERNAL AGE; REPEATABILITY; TRISOMY-21;
D O I
10.1002/uog.7737
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objective To estimate intersonographer and intrasonographer variance components of fetal nuchal translucency (NT) thickness measurement using the traditional manual approach and a new semi-automated system. Methods A semi-automated method was developed for measurement of the NT. In this method, the operator places an adjustable box over the relevant area at the back of the fetal neck. The system draws a line through the center of the nuchal membrane and another line at the edge of the soft tissue overlying the cervical spine. The system then identifies the largest vertical distance between the two lines. The images of 12 fetuses at 11-13 weeks of gestation satisfying the guidelines of The Fetal Medicine Foundation for measurement of NT were selected. They were exported in DICOM format from the ultrasound system, and four versions of each image were stored under different names. The resulting 48 images were presented in random order for electronic assessment. A total of 20 sonographers measured the NT in each set of 48 pictures, twice using the semi-automated system and twice using the manual system, according to a randomized block design. Within- and between-operator variance components were estimated. Relative biases were assessed by comparing the means from the two methods. Results The estimated between-operator SD using the semi-automated method was 0.0149 mm compared with 0.109 mm for the manual method. The respective within-operator SD values were 0.05 mm and 0.126 mm. The intraclass correlation coefficients for different sonographers measuring the same images were 0.98 and 0.85 for the semi-automated method and the manual method, respectively. Conclusion The measurement of fetal NT is more reliable when a semi-automatic approach is used rather than the traditional manual method. Copyright (C) 2010 ISUOG. Published by John Wiley & Sons, Ltd.
引用
收藏
页码:412 / 416
页数:5
相关论文
共 50 条
  • [1] Effect of image settings on nuchal translucency thickness measurement by a semi-automated system
    Chen, P. W.
    Chen, M.
    Leung, T. Y.
    Lau, T. K.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2012, 39 (02) : 169 - 174
  • [2] Semi-automated measurement of nuchal translucency thickness: blasphemy or oblation to quality?
    Ville, Y.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2010, 36 (04) : 400 - 403
  • [3] Semi-automated adjusted measurement of nuchal translucency: feasibility and reproducibility
    Grange, G.
    Althuser, M.
    Fresson, J.
    Bititi, A.
    Miyamoto, K.
    Tsatsaris, V.
    Morel, O.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2011, 37 (03) : 335 - 340
  • [4] Ultrasound semi-automated measurement of fetal nuchal translucency thickness based on principal direction estimation
    Yoon, Heechul
    Lee, Hyuntaek
    Jung, Haekyung
    Lee, Mi-Young
    Won, Hye-Sung
    [J]. MEDICAL IMAGING 2015: ULTRASONIC IMAGING AND TOMOGRAPHY, 2015, 9419
  • [5] A Semi-automated Method for the Measurement of the Fetal Nuchal Translucency in Ultrasound Images
    Catanzariti, Ezio
    Fusco, Giovanni
    Isgro, Francesco
    Masecchia, Salvatore
    Prevete, Roberto
    Santoro, Matteo
    [J]. IMAGE ANALYSIS AND PROCESSING - ICIAP 2009, PROCEEDINGS, 2009, 5716 : 613 - +
  • [6] Reliability of manual and semi-automated measurements of nuchal translucency by experienced operators
    Pruksanusak, Ninlapa
    Pranpanus, Savitree
    Suwanrath, Chitkasaem
    Kor-anantakul, Ounjai
    Suntharasaj, Thitima
    Hanprasertpong, Tharangrut
    Liabsuetrakul, Tippawan
    [J]. INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 2013, 121 (03) : 240 - 242
  • [7] The influence of image setting on intracranial translucency measurement by manual and semi-automated system
    Zhen, Li
    Yang, Xin
    Ting, Yuen Ha
    Chen, Min
    Leung, Tak Yeung
    [J]. PRENATAL DIAGNOSIS, 2013, 33 (09) : 889 - 893
  • [8] Intra- and interoperator reliability of manual and semi-automated measurement of fetal nuchal translucency by sonographers with different levels of experience
    Abele, H.
    Hoopmann, M.
    Wright, D.
    Hoffmann-Poell, B.
    Huettelmaier, M.
    Pintoffl, K.
    Wallwiener, D.
    Kagan, K. O.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2010, 36 (04) : 417 - 422
  • [9] Intraoperator and interoperator repeatability of manual and semi-automated measurement of increased fetal nuchal translucency according to the operator's experience
    Kagan, Karl Oliver
    Abele, Harald
    Yazdi, Britta
    Boeer, Bettina
    Pintoffl, Klaus
    Wright, Dave
    Hoopmann, Markus
    [J]. PRENATAL DIAGNOSIS, 2011, 31 (13) : 1229 - 1233
  • [10] Impact of nuchal cord on measurement of fetal nuchal translucency thickness
    Scheier, M.
    Egle, D.
    Himmel, I.
    Ramoni, A.
    Viertl, S.
    Huter, O.
    Marth, C.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2007, 30 (02) : 197 - 200