MEMS-Based Nanospray-Ionization Mass Spectrometer

被引:22
|
作者
Wright, Steven [1 ]
Syms, Richard R. A. [2 ]
Moseley, Richard [1 ]
Hong, Guodong [1 ]
O'Prey, Shane [1 ]
Boxford, William E. [1 ]
Dash, Neil [1 ]
Edwards, Peter [1 ]
机构
[1] Microsa Syst Ltd, Surrey GU21 5BX, England
[2] Univ London Imperial Coll Sci Technol & Med, Dept Elect & Elect Engn, London SW7 2AZ, England
基金
英国工程与自然科学研究理事会;
关键词
Electrospray; mass spectrometry; microelectromechanical systems (MEMS); quadrupole filter; ELECTROSTATIC ION GUIDE; LIQUID-CHROMATOGRAPHY; ELECTROSPRAY INTERFACE; MOLECULAR-BEAMS; GAS-DYNAMICS; QUADRUPOLE; TRAP; PLASMA; NANOELECTROSPRAY; DEVICES;
D O I
10.1109/JMEMS.2010.2082501
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An electrospray-ionization mass spectrometer (ESI-MS) whose main components are all fabricated using silicon microelectromechanical systems (MEMS) techniques is demonstrated for the first time. The ion source consists of a microengineered alignment bench containing a V-groove mounting for a nanospray capillary, an ion-extraction electrode, and a pneumatic nebulizer. The vacuum interface consists of two plates, each carrying a 50-mu m-diameter capillary, that are selectively etched and bonded together to provide a differentially pumped internal cavity. The quadrupole filter consists of a microfabricated frame that provides mountings for stainless-steel rods measuring 650 mu m in diameter and 30 mm in length. Two different quadrupoles are compared: a first-generation bonded silicon device and a second-generation silicon-on-glass device with a Brubaker prefilter. Differential pumping of a MEMS component is demonstrated for the first time, atmospheric pressure ionization and ion transfer into vacuum are characterized, ESI-MS operation is demonstrated, and spectra are presented for a variety of compounds.
引用
收藏
页码:1430 / 1443
页数:14
相关论文
共 50 条
  • [1] A portable MEMS-based mass spectrometer
    Hauschild, Jan-Peter
    Wapelhorst, Eric
    Mueller, Joerg
    TM-TECHNISCHES MESSEN, 2008, 75 (02) : 129 - 134
  • [2] MEMS-based spedde spectrometer
    Sheinis, A. I.
    Nigra, L.
    Kuhlen, M. Q.
    ADVANCES IN ADAPTIVE OPTICS II, PRS 1-3, 2006, 6272 : U1589 - U1599
  • [3] Development of a MEMS-Based Raman Spectrometer
    Russin, Timothy
    Fralick, Mark
    Kerber, Max
    Wang, Andrew
    Waters, Richard
    2010 IEEE SENSORS, 2010, : 56 - 60
  • [4] Harmonic Analysis with a MEMS-based Raman Spectrometer
    Russin, Timothy J.
    Kerber, Maxwell
    Russin, Alicia M.
    Wang, Andrew
    Waters, Richard
    2011 IEEE SENSORS, 2011, : 1588 - 1591
  • [6] Infra-red MEMS-based lamellar grating spectrometer
    Manzardo, O
    Schädelin, F
    Noell, W
    de Rooij, N
    Herzig, HP
    MEMS, MOEMS, AND MICROMACHINING, 2004, 5455 : 1 - 8
  • [7] A MEMS-based thermal infrared emitter for an integrated NDIR spectrometer
    Calaza, C.
    Salleras, M.
    Sabate, N.
    Santander, J.
    Cane, C.
    Fonseca, L.
    MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS, 2012, 18 (7-8): : 1147 - 1154
  • [8] MEMS-based polarized FTIR spectrometer for polymer quality control
    Eltagoury, Yomna M.
    Sabry, Yasser M.
    Khalil, Diaa A.
    MOEMS AND MINIATURIZED SYSTEMS XX, 2021, 11697
  • [9] A MEMS-based thermal infrared emitter for an integrated NDIR spectrometer
    C. Calaza
    M. Salleras
    N. Sabaté
    J. Santander
    C. Cané
    L. Fonseca
    Microsystem Technologies, 2012, 18 : 1147 - 1154
  • [10] MEMS-Based QCL Spectrometer Aims at Hazardous Materials Detection
    不详
    PHOTONICS SPECTRA, 2015, : 5 - 5