Calibration of soft X-ray detection with synchrotron radiation

被引:0
|
作者
Sun, KX [1 ]
Yi, RQ
Jiang, SE
Yang, JM
Huang, TX
Cui, YL
Yang, GH
Ding, YK
Liu, SY
Zhang, BH
Wang, HB
Wen, SH
Zheng, ZJ
Jiang, WM
Zhao, YK
Cui, MQ
Zhu, PP
Zhao, YD
Li, G
Zhu, J
Zheng, L
Tang, ES
机构
[1] CAEP, Res Ctr Laser Fus, Natl Key Lab Laser Fus, Mianyang 621900, Peoples R China
[2] CAS, SR Lab, Inst High Energy Phys, Beijing 100039, Peoples R China
关键词
synchrotron radiation; soft X-ray; sensitivity; reflectivity; transmissivity;
D O I
暂无
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
In this paper the calibrations of different soft X-ray detector elements used in experiment diagnostics of Inertial Confinement Fusion (ICF) with Synchrotron Radiation is reported. Beijing Synchrotron Radiation Facility used special operation type with beam current 35-110mA, the storage ring electron energy 2GeV, photon energy region 50-1500eV. The performance of the light source is studied by transmission grating and face array CCD system. We divide 50-1500eV energy region in seven energy section through different material prefilter with monochromaticity than 95 % The energy response of different soft X-ray detector elements are calibrated. The calibration results are used in ICF data analysis, which secures the accuracy of data treatment.