Backscatter correction factor for megavoltage photon beam

被引:6
|
作者
Hu, Yida [1 ]
Zhu, Timothy C. [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
关键词
backscatter correction factor; megavoltage; photon beam; convolution method; ENERGY; EXIT;
D O I
10.1118/1.3633903
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: For routine clinical dosimetry of photon beams, it is often necessary to know the minimum thickness of backscatter phantom material to ensure that full backscatter condition exists. Methods: In case of insufficient backscatter thickness, one can determine the backscatter correction factor, BCF(s,d,t), defined as the ratio of absorbed dose measured on the central-axis of a phantom with backscatter thickness of t to that with full backscatter for square field size s and forward depth d. Measurements were performed in SAD geometry for 6 and 15 MV photon beams using a 0.125 cc thimble chamber for field sizes between 10 x 10 and 30 x 30 cm at depths between d(max) (1.5 cm for 6 MV and 3 cm for 15 MV) and 20 cm. Results: A convolution method was used to calculate BCF using Monte-Carlo simulated point-spread kernels generated for clinical photon beams for energies between Co-60 and 24 MV. The convolution calculation agrees with the experimental measurements to within 0.8% with the same physical trend. The value of BCF deviates more from 1 for lower energies and larger field sizes. According to our convolution calculation, the minimum BCF occurs at forward depth d(max) and 40 x 40 cm field size, 0.970 for 6 MV and 0.983 for 15 MV. Conclusions: The authors concluded that backscatter thickness is 6.0 cm for 6 MV and 4.0 cm for 15 MV for field size up to 10 x 10 cm when BCF = 0.998. If 4 cm backscatter thickness is used, BCF is 0.997 and 0.983 for field size of 10 x 10 and 40 x 40 cm for 6 MV, and is 0.998 and 0.990 for 10 x 10 and 40 x 40 cm for 15 MV, respectively. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3633903]
引用
收藏
页码:5563 / 5568
页数:6
相关论文
共 50 条
  • [1] Backscatter Correction Factor for Megavoltage Photon Beam
    Hu, Y.
    Zhu, T.
    MEDICAL PHYSICS, 2009, 36 (06) : 2601 - +
  • [2] Backscatter Correction Factor for Megavoltage Photon Beam Using Divergent Point-Spread Kernels
    Liang, X.
    Zhu, T.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [3] DOSIMETRY OF ELECTRON BACKSCATTER AT MANDIBULAR RECONSTRUCTION IN MEGAVOLTAGE PHOTON-BEAM
    DAS, IJ
    BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (758): : 185 - 186
  • [4] SEPARATING OUTPUT FACTOR INTO COLLIMATOR FACTOR AND BACKSCATTER FACTOR FOR MEGAVOLTAGE PHOTON DOSAGE CALCULATIONS
    SPICKA, J
    HERRON, D
    ORTON, CG
    HARDYBALA, S
    MEDICAL PHYSICS, 1983, 10 (04) : 543 - 543
  • [5] FURTHER BACKSCATTER STUDIES AND THEIR APPLICATION IN MEGAVOLTAGE PHOTON THERAPY
    WEATHERBURN, H
    BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (692): : 783 - 785
  • [6] The physics of small megavoltage photon beam dosimetry
    Andreo, Pedro
    RADIOTHERAPY AND ONCOLOGY, 2018, 126 (02) : 205 - 213
  • [7] Alignment of multiradiation isocenters for megavoltage photon beam
    Zhang, Yin
    Ding, Kai
    Cowan, Garth
    Tryggestad, Erik
    Armour, Elwood
    Wang, Ken Kang-Hsin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (06): : 314 - 324
  • [8] Effect of contrast media on megavoltage photon beam dosimetry
    Rankine, Ashley W.
    Lanzon, Peter J.
    Spry, Nigel A.
    MEDICAL DOSIMETRY, 2008, 33 (03) : 169 - 174
  • [9] On beam quality and flatness of radiotherapy megavoltage photon beams
    Hossain, Murshed
    Rhoades, Jeffrey
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (01) : 135 - 145
  • [10] THE PROBLEMS AND PERSPECTIVES OF MEGAVOLTAGE PHOTON BEAM RADIOTHERAPY IN UKRAINE
    Starenkiy, V. P.
    Vasyliev, L. L.
    Samofalov, I. O.
    PROBLEMS OF ATOMIC SCIENCE AND TECHNOLOGY, 2016, (05): : 94 - 97