Flow-injection electrogenerated chemiluminescence determination of isoniazid using luminol

被引:34
|
作者
Zheng, XW [1 ]
Guo, ZH [1 ]
Zhang, ZJ [1 ]
机构
[1] Shaanxi Normal Univ, Dept Chem, Xian 710062, Peoples R China
关键词
D O I
10.2116/analsci.17.1095
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Based on a new electrogenerated chemiluminescence (ECL) analytical idea, this paper explains a sensitive and selective flow-injection ECL method using luminol for the determination of isoniazid, based on the sensitizing effect of isoniazid for the weak ECL emission of electrochemically oxidized luminol. Under the optimum experimental conditions, the relative ECL intensity was linear with isoniazid concentration in the range of 4.0 x 10(-8) mol/L to 8.0 x 10(-6) mol/L and with a detecting limit of 2.8 x 10(-8) mol/L.
引用
收藏
页码:1095 / 1099
页数:5
相关论文
共 50 条