High background in Luminex® assay for HLA antibody screening: Interest of Adsorb Out™

被引:5
|
作者
Zerrouki, Asmae [1 ]
Ouadghiri, Sanae [1 ,2 ]
Benseffaj, Nadia [1 ,2 ]
Razine, Rachid [3 ]
Essakalli, Malika [1 ,2 ]
机构
[1] Univ Mohammed 5, Fac Med & Pharm Rabat, UPR Immunol, Rabat, Morocco
[2] Ctr Hosp Ibn Sina, Serv Transfus Sanguine & Hemovigilance, Rabat, Morocco
[3] Univ Mohammed 5, Fac Med & Pharm Rabat, Lab Epidemiol & Rech Clin, Lab Med Sociale, Rabat, Morocco
关键词
Luminex; HLA antibodies; High negative control bead value; Adsorb Out (TM); ANTIGEN-SPECIFIC ALLOANTIBODIES; DONOR-SPECIFIC ANTIBODIES; KIDNEY-TRANSPLANTATION; CLINICAL-RELEVANCE; PREGNANCY; BEADS; REJECTION;
D O I
10.1016/j.trim.2016.03.001
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Introduction: The Luminex (R) technology has become an integral component of clinical decision-making and diagnosis of transplanted organ rejection. Despite the superior sensibility of this technology, it is not completely problem free. We have observed in these bead-based assays that sera of some patients give a high negative control bead (NC) value which makes assessing HLA antibodies difficult. Treatment of sera by the Adsorb Out (TM) reagent may reduce the high background. In this study, we want to evaluate the effect of the Adsorb Out on the NC's MFI value by comparing treated and untreated patients' sera. Methods: HIA antibody screening was performed on 3011 sera. These sera came from patients awaiting and undergoing renal transplant from different Moroccan hospitals. The sera were analyzed using the standard protocol for Luminex (R) antibody screening. Sera with high NC's value has been pre-incubated by the Adsorb Out (TM), and analyzed on Luminex (R). Results: 3% of studied samples have high NC's value. The Adsorb Out decreases the NC's value and brings it back to a normal range in 62.2% treated sera. It has no effect in 12.3%. The Adsorb Out (TM) effect depends only of NC's value, independently to age, storage date, sex and immunization. Conclusion: The Adsorb Out (TM) reagent has an important effect in decreasing NC value of sera. However, it has no effect in some patient's sera. In these cases we could try another treatment, as EDTA, DTT. The non-specific binding may be caused by multiple patient-specific factors, it would be important to search correlation between them and NC's values. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 50 条
  • [2] UTILITY OF THE LIFESCREEN DELUXE (LMX): AN HLA ANTIBODY SCREENING ASSAY FOR LUMINEX
    Vega, Renato M.
    Callender, Carly J.
    Sholander, Jeff T.
    Leffell, Mary S.
    Zachary, Andrea A.
    [J]. HUMAN IMMUNOLOGY, 2009, 70 : S46 - S46
  • [3] A comparison of luminex HLA antibody screening methods
    Fernando, R.
    Turakhia, G.
    Karasu, A.
    Shirling, G.
    Beo, J.
    Little, A-M
    Stephens, H. A. F.
    [J]. INTERNATIONAL JOURNAL OF IMMUNOGENETICS, 2006, 33 (04) : 318 - 318
  • [4] THE PROBLEM OF HIGH BACKGROUND ON LUMINEX HLA ANTIBODY TESTING CAN BE VENDOR SPECIFIC
    Maccagno, Giancarlo
    Jung, Achim
    Carbol, Alexander
    Marandiuc, Dana
    Conradi, Roland
    Hitzler, Walter
    [J]. HLA, 2019, 93 (05) : 306 - 306
  • [5] CORRELATION OF FLOW HLA ANTIBODY SCREENING AND LUMINEX SINGLE ANTIGEN BEAD ASSAY.
    Lopes, Karine E.
    Clark, Adella
    Wan, Liang
    Wegner, Wendy E.
    Chen, Dong-Feng
    [J]. HUMAN IMMUNOLOGY, 2014, 75 : 76 - 76
  • [6] IMPACT OF ANTICOAGULANT IN SAMPLES FOR HLA ANTIBODY SCREENING BY LUMINEX
    Book, Benita K.
    Higgins, Nancy G.
    Rosner, Kevin M.
    Pescovitz, Mark D.
    [J]. HUMAN IMMUNOLOGY, 2008, 69 : S35 - S35
  • [7] The difference Luminex HLA antibody screening can make
    Davies, M.
    Stewart, K.
    Lockhart, S.
    Lumsdaine, J.
    [J]. INTERNATIONAL JOURNAL OF IMMUNOGENETICS, 2006, 33 (04) : 319 - 319
  • [8] A NOVEL LUMINEX SCREENING ASSAY FOR THE DETECTION OF IGA HLA ANTIBODIES
    Car, Helena
    Kramer, Cynthia S. M.
    Karahan, Gonca E.
    Claas, Frans H. J.
    Roelen, Dave L.
    Heidt, Sebastiaan
    [J]. HLA, 2019, 93 (05) : 292 - 292
  • [9] The high-dose hook effect in the detection and monitoring of HLA specific antibody by Luminex assay
    Lowe, D.
    Hathaway, M.
    Briggs, D.
    [J]. INTERNATIONAL JOURNAL OF IMMUNOGENETICS, 2007, 34 (04) : 288 - 288
  • [10] IS MFI GOOD ENOUGH TO INDICATE HLA ANTIBODY STRENGTH IN LUMINEX ASSAY?
    Du, Zeying
    Herczyk, Walter
    Weidner, Jerome
    Zimmerman, Andrea
    Ramahi, Sana
    Marino, Susana
    [J]. HUMAN IMMUNOLOGY, 2009, 70 : S16 - S16