Flap Monitoring using Thermal Imaging Camera: A Contactless Method

被引:11
|
作者
Rabbani, Muhammad Jibran [1 ]
Bhatti, Asif Zubair [1 ]
Shahzad, Ahmed [1 ]
机构
[1] Shalamar Hosp, Dept Burn & Plast Surg, Lahore, Pakistan
关键词
Flir one; Flap monitoring; Thermal imaging camera; INFRARED THERMOGRAPHY; SALVAGE;
D O I
10.29271/jcpsp.2021.06.703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To determine the accuracy of the thermal imaging camera for the detection of vascular insult before the onset of permanent tissue damage. Study Design: Descriptive study. Place and Duration of Study: Department of Burn and Plastic Surgery, from July to December 2019. Methodology: The thermal image was used as an adjunct to clinical assessment in 84 flaps. Both pedicled and free flaps were performed in the study with a frequency of 68 (81%) and 16 (19%), respectively. Flap monitoring was performed regularly for five days by same team of surgeons and nurses. Flap vascularity was assessed by temperature gradient colour coding of thermal image taking the clinical assessment like colour, temperature, capillary refill, as a gold standard. Sensitivity and specificity of the thermal camera were calculated. Results: The age range of patients was from 5 years to 55 years. The diagnostic accuracy of thermal imaging camera was calculated to be 96.43% with a sensitivity of 98.7%, specificity of 75%. The positive predictive value was found to be 97.4%, negative predictive value was 85.7%, the false-negative value was 14.3%, and the false-positive value was 2.6%. Conclusion: Thermal imaging camera is an effective and useful adjunct for clinical assessment, as it is an easy and contactless technique.
引用
收藏
页码:703 / 706
页数:4
相关论文
共 50 条
  • [1] Fully Automated Contactless Respiration Monitoring Using a Camera
    Rocque, Mukul
    [J]. 2016 IEEE INTERNATIONAL CONFERENCE ON CONSUMER ELECTRONICS (ICCE), 2016,
  • [2] An Experimental and Clinical Study of Flap Monitoring with an Analysis of the Clinical Course of the Flap Using an Infrared Thermal Camera
    Kim, Hyun
    Kwak, Si Hyun
    Byeon, Je Yeon
    Lee, Da Woon
    Kim, Jun Hyuk
    Lim, Soomin
    Choi, Hwan Jun
    [J]. BIOENGINEERING-BASEL, 2024, 11 (07):
  • [3] Contactless method for monitoring the thermal properties of materials
    Mishchenko, SV
    Artyukhin, LS
    Dobrokhotov, SB
    [J]. INDUSTRIAL LABORATORY, 1998, 64 (02): : 98 - 102
  • [4] Contactless Monitoring Of Human Vitals: A Study With Simultaneous Measurements Using FMCW Radar And Thermal Camera
    Kundu, Soumitra
    Panda, Gargi
    Routray, Aurobinda
    Guha, Rajlakshmi
    Mohanty, Pravansu
    [J]. 2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [5] Contactless Heart Rate Monitoring Using A Standard RGB Camera
    Maji, Soumyajyoti
    Massaroni, Carlo
    Schena, Emiliano
    Silvestri, Sergio
    [J]. 2020 IEEE INTERNATIONAL WORKSHOP ON METROLOGY FOR INDUSTRY 4.0 & IOT (METROIND4.0&IOT), 2020, : 729 - 733
  • [6] Method for Measuring Weld Temperature Using an Infrared Thermal Imaging Camera
    Ro, Chan-Seung
    Kim, Kyeong-Suk
    Chang, Ho-Seob
    [J]. JOURNAL OF THE KOREAN SOCIETY FOR NONDESTRUCTIVE TESTING, 2014, 34 (04) : 299 - 304
  • [7] APPLICATIONS OF SMARTPHONE THERMAL CAMERA IMAGING SYSTEM IN MONITORING OF THE DEEP INFERIOR EPIGASTRIC PERFORATOR FLAP FOR BREAST RECONSTRUCTION
    Konczalik, Wojciech
    Nikkhah, Dariush
    Mosahebi, Afshin
    [J]. MICROSURGERY, 2017, 37 (05) : 457 - 458
  • [8] Contactless Interface Using Exhaled Breath and Thermal Imaging
    Lee, Kanghoon
    Park, Jong-Il
    [J]. SENSORS, 2023, 23 (07)
  • [9] Towards a Contactless Stress Classification Using Thermal Imaging
    Gioia, Federica
    Greco, Alberto
    Callara, Alejandro Luis
    Scilingo, Enzo Pasquale
    [J]. SENSORS, 2022, 22 (03)
  • [10] METHOD FOR THE CONTACTLESS MEASUREMENT OF PLANE MOTIONS USING A CCD-CAMERA
    DIETMAYER, K
    KRUGER, K