Finite element simulation and statistical investigation of an orthodontic mini-implant's stability in a novel screw design

被引:7
|
作者
Pouyafar, Vahid [1 ]
Meshkabadi, Ramin [2 ]
Sadr Haghighi, Amir Hooman [3 ]
Navid, Ali [1 ]
机构
[1] Univ Tabriz, Dept Mfg Engn, Tabriz, Iran
[2] Univ Mohaghegh Ardabili, Fac Adv Technol, Dept Engn Sci, Namin, Iran
[3] Tabriz Univ Med Sci, Fac Dent, Dept Orthodont, Tabriz, Iran
关键词
Mini-implant's stability; screw geometry; finite element simulation; Taguchi method; variance analysis; MECHANICAL-PROPERTIES; TAPER SHAPE; THREAD; LENGTH; ANCHORAGE;
D O I
10.1177/09544119211023630
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
One of the essential aspects of the mini-implant's successful application is its stability after being installed in the bone. The stability of the mini-implant affected the most by its geometry. In the present research, the effect of the geometry-related parameters of the mini-implant on its lateral displacement is investigated by Finite Element (FE) modeling using ABAQUS software. The parameters studied include length, diameter, pitch, and depth of the screw threads; besides, length and angle of the conical section. The Taguchi method was used to prevent many experiments. The mesh convergence tests and experimental tests confirmed the FE model quantitatively and qualitatively. Mean of means and variance analysis determined the parameters significance and their contribution on the stability. The screw diameter and length have the most contribution to mini-implant' displacement. The effect of screw pitch was less than that for length and diameter. The conical section improved the initial stability by creating compressive stress and additional friction in its surrounding bone. No significant effects on the stability of the mini-implant have been observed for the non-threaded part. By examining the effect of thread depth on its stability by defining the ratio of thread depth to the internal diameter and to maintain the strength of the screw the optimal value for internal to external ratio is set at about 0.7.
引用
收藏
页码:1046 / 1057
页数:12
相关论文
共 50 条
  • [1] Experimental & FEM Analysis of Orthodontic Mini-Implant Design on Primary Stability
    Mesic, Elmedin
    Muratovic, Enis
    Redzepagic-Vrazalica, Lejla
    Pervan, Nedim
    Muminovic, Adis J.
    Delic, Muamer
    Glusac, Mirza
    APPLIED SCIENCES-BASEL, 2021, 11 (12):
  • [2] Effective en-masse retraction design with orthodontic mini-implant anchorage: A finite element analysis
    Sung, Sang-Jin
    Jang, Gang-Won
    Chun, Youn-Sic
    Moon, Yoon-Shik
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (05) : 648 - 657
  • [3] Mechanical stability and clinical applicability assessment of novel orthodontic mini-implant design Part 3
    Song, Ha Na
    Hong, Christine
    Banh, Robert
    Ohebsion, Tania
    Asatrian, Greg
    Leung, Ho-Yin
    Wu, Benjamin M.
    Moon, Won
    ANGLE ORTHODONTIST, 2013, 83 (05) : 832 - 841
  • [4] Finite element analysis of mono- and bicortical mini-implant stability
    Holberg, Christof
    Winterhalder, Philipp
    Rudzki-Janson, Ingrid
    Wichelhaus, Andrea
    EUROPEAN JOURNAL OF ORTHODONTICS, 2014, 36 (05) : 550 - 556
  • [5] Biomechanical effect of abutment on stability of orthodontic mini-implant
    Motoyoshi, M
    Yano, S
    Tsuruoka, T
    Shimizu, N
    CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) : 480 - 485
  • [6] Mechanical stability assessment of novel orthodontic mini-implant designs: Part 2
    Hong, Christine
    Peter Truong
    Ha Na Song
    Wu, Benjamin M.
    Moon, Won
    ANGLE ORTHODONTIST, 2011, 81 (06) : 1001 - 1009
  • [7] Orthodontic mini-implant stability and the ratio of pilot hole implant diameter
    Uemura, Miwa
    Motoyoshi, Mitsuru
    Yano, Shinya
    Sakaguchi, Masahito
    Igarashi, Yu
    Shimizu, Noriyoshi
    EUROPEAN JOURNAL OF ORTHODONTICS, 2012, 34 (01) : 52 - 56
  • [8] Mechanical influence of thread pitch on orthodontic mini-implant stability
    da Cunha, Amanda Carneiro
    Ayres de Freitas, Amanda Osorio
    Marquezan, Mariana
    Nojima, Lincoln Issamu
    BRAZILIAN ORAL RESEARCH, 2015, 29 (01): : 1 - 6
  • [9] A novel technique for measurement of orthodontic mini-implant stability using the Osstell ISQ device
    Hosein, Yara K.
    Dixon, S. Jeffrey
    Rizkalla, Amin S.
    Tassi, Ali
    ANGLE ORTHODONTIST, 2019, 89 (02) : 284 - 291
  • [10] Influence of bone density, screw size and surgical procedure on orthodontic mini-implant placement - part B: implant stability
    Moehlhenrich, S. C.
    Heussen, N.
    Modabber, A.
    Bock, A.
    Hoelzle, F.
    Wilmes, B.
    Danesh, G.
    Szalma, J.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2021, 50 (04) : 565 - 572