An acquisition and image reconstruction scheme for reduced x-ray exposure dynamic 3D CTA

被引:0
|
作者
Supanich, M. [1 ]
Rowley, H. [2 ]
Turk, A. [2 ]
Speidel, M. [1 ]
Pulfer, K. [2 ]
Hsieh, J. [3 ]
Chen, G. [1 ,2 ]
Mistretta, C. [1 ,2 ]
机构
[1] UW Madison Dept Med Phys, 3000 N Grandview Bld, Waukesha, WI USA
[2] UW Hosp, Dept Radiol, Waukesha, WI USA
[3] GE Healthcare, Waukesha, WI USA
关键词
dose reduction; dynamic CTA; cerebral perfusion; HYPR; multi-detector CT; undersampled reconstruction;
D O I
10.1117/12.772656
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present Computed Tomography (CT) acquisition and reconstruction schemes for low-dose neuro-angiography based on the method of HighlY constrained back PRojection (HYPR). Simulated and experimental low X-ray radiation dose scans were prepared using the techniques of interleaved view angle under-sampling and tube current reduction. Dynamic CT Angiograms (CTAs) were produced for both standard and low dose images sets. The spatial correlation coefficient, r, between the two reconstruction approaches was determined for each time frame and the SNR and CNR values in arterial ROIs were calculated. The undersampled HYPR reconstructions produced r values of > 0.95 at undersampling and dose reduction factors-of 10 and SNR and CNR were more than doubled using HYPR techniques at a tube current of 25 mA. HYPR approaches to contrast enhanced neuro-imaging provide not only volumetric brain hemodynamics but also the ability to produce high quality maps of standard perfusion parameters. The synergy of volumetric hemodynamics and assessment of tissue function provides the medical imaging community with high quality diagnostic information at a fraction of the radiation dose in a single contrast-enhanced scan.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] 3D Image Reconstruction from X-Ray Measurements with Overlap
    Klodt, Maria
    Hauser, Raphael
    [J]. COMPUTER VISION - ECCV 2016, PT VI, 2016, 9910 : 19 - 33
  • [2] Reduced memory augmented Lagrangian algorithm for 3D iterative X-ray CT image reconstruction
    McGaffin, Madison G.
    Ramani, Sathish
    Fessler, Jeffrey A.
    [J]. MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [3] 3D IMAGE RECONSTRUCTION ON X-RAY MICRO-COMPUTED TOMOGRAPHY
    Louk, Andreas C.
    Suparta, Gede B.
    [J]. INTERNATIONAL CONFERENCE ON EXPERIMENTAL MECHANICS 2014, 2015, 9302
  • [4] 3D X-RAY IMAGE COMPOSITION
    Armeanu, Constantin Catalin
    [J]. UNIVERSITY POLITEHNICA OF BUCHAREST SCIENTIFIC BULLETIN-SERIES A-APPLIED MATHEMATICS AND PHYSICS, 2016, 78 (01): : 309 - 320
  • [5] Calibration of X-ray projections in 3D reconstruction
    Wang, P
    Mou, XQ
    Qin, ZY
    Cai, YL
    [J]. VISUALIZATION AND OPTIMIZATION TECHNIQUES, 2001, 4553 : 44 - 49
  • [6] The Study on the Implementation of the X-Ray CT System Using the Cone-Beam for the 3D Dynamic Image Acquisition
    Jeong, Chan-Woong
    Jun, Kyu-Suk
    [J]. JOURNAL OF THE ACOUSTICAL SOCIETY OF KOREA, 2009, 28 (04): : 370 - 374
  • [7] Full-color image reconstruction by holography for 3D X-ray CT
    Koike, Akifumi
    Morii, Hisashi
    Yomori, Mitsuhiro
    Neo, Yoichiro
    Aoki, Toru
    Mimura, Hidenori
    [J]. 8TH INTERNATIONAL CONFERENCE ON CORRELATION OPTICS, 2008, 7008
  • [8] 2D-to-3D: A Review for Computational 3D Image Reconstruction from X-ray Images
    Maken, Payal
    Gupta, Abhishek
    [J]. ARCHIVES OF COMPUTATIONAL METHODS IN ENGINEERING, 2023, 30 (01) : 85 - 114
  • [9] 2D-to-3D: A Review for Computational 3D Image Reconstruction from X-ray Images
    Payal Maken
    Abhishek Gupta
    [J]. Archives of Computational Methods in Engineering, 2023, 30 : 85 - 114
  • [10] 3D reconstruction with X-ray CT for nondestructive inspection
    Tanaka, Toshiyuki
    [J]. TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 2008, 94 (07): : 483 - 486