共 50 条
Capillary electrophoresis of trace metals in highly saline physiological sample matrices
被引:32
|作者:
Riaz, A
[1
]
Kim, B
[1
]
Chung, DS
[1
]
机构:
[1] Seoul Natl Univ, Sch Chem, Seoul 151747, South Korea
关键词:
capillary electrophoresis;
trace metals;
transient isotachophoresis;
urine;
D O I:
10.1002/elps.200305537
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
Trace metal ions in highly saline samples such as urine were determined with capillary electrophoresis (CE) without desalting or off-line preconcentration. By mixing with a dye, 4-(2-pyridylazo) resorcinol (PAR), the metal ions were converted into anionic complexes having strong absorbance near 500 nm. A large volume of the metal-PAR complex sample solution injected into a coated capillary was stacked isotachophoretically and separated under a reverse potential. The salt anion (chloride) and PAR in the sample matrix acted as the leading and terminating electrolytes, respectively. In a sample containing a 250 mm NaCl matrix, more than 400-fold enhancement in the absorbance detector response was realized compared to the normal CE injection mode. Combination of the dye complexation and isotachophoretic stacking provided excellent detection limits (S/N = 3) for three trace metal ions in the low ppb range (Fe2+, 0.7 ppb, Ni2+, 0.4 ppb; Zn2+, 1.2 ppb) with absorbance detection. The migration time reproducibility was excellent (relative standard deviations: standard samples < 1%, urine samples similar to 1%). The proposed method is convenient and fast, and the sample analysis can be completed within 20 min.
引用
收藏
页码:2788 / 2795
页数:8
相关论文