Optical design and performance of the biological small-angle X-ray scattering beamline at the Taiwan Photon Source

被引:18
|
作者
Liu, D-G [1 ]
Chang, C-H [1 ]
Chiang, L-C [1 ]
Lee, M-H [1 ]
Chang, C-F [1 ]
Lin, C-Y [1 ]
Liang, C-C [1 ]
Lee, T-H [1 ]
Lin, S-W [1 ]
Liu, C-Y [1 ]
Hwang, C-S [1 ]
Huang, J-C [1 ]
Kuan, C-K [1 ]
Wang, H-S [1 ]
Liu, Y-C [1 ]
Tseng, F-H [1 ]
Chuang, J-Y [1 ]
Liao, W-R [1 ]
Li, H-C [1 ]
Su, C-J [1 ]
Liao, K-F [1 ]
Yeh, Y-Q [1 ]
Shih, O. [1 ]
Wu, W-R [1 ]
Wang, C-A [1 ]
Jeng, U. [1 ,2 ]
机构
[1] Natl Synchrotron Radiat Res Ctr, Hsinchu Sci Pk, Hsinchu 30076, Taiwan
[2] Natl Tsing Hua Univ, Dept Chem Engn, Hsinchu 30013, Taiwan
关键词
BioSAXS; high flux; ultra-SAXS; microbeam; beamline optics;
D O I
10.1107/S1600577521009565
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The optical design and performance of the recently opened 13A biological small-angle X-ray scattering (SAXS) beamline at the 3.0GeV Taiwan Photon Source of the National Synchrotron Radiation Research Center are reported. The beamline is designed for studies of biological structures and kinetics in a wide range of length and time scales, from angstrom to micrometre and from microsecond to minutes. A 4m IU24 undulator of the beamline provides high-flux X-rays in the energy range 4.0-23.0keV. MoB4C double-multilayer and Si(111) double-crystal monochromators (DMM/DCM) are combined on the same rotating platform for a smooth rotation transition from a high-flux beam of similar to 4 x 10(14)photonss(-1) to a high-energy-resolution beam of Delta E/E similar or equal to 1.5 x 10(-4); both modes share a constant beam exit. With a set of Kirkpatrick-Baez (KB) mirrors, the X-ray beam is focused to the farthest SAXS detector position, 52m from the source. A downstream four-bounce crystal collimator, comprising two sets of Si(311) double crystals arranged in a dispersive configuration, optionally collimate the DCM (vertically diffracted) beam in the horizontal direction for ultra-SAXS with a minimum scattering vector q down to 0.0004 angstrom(-1), which allows resolving ordered d-spacing up to 1 mu m. A microbeam, of 10-50 mu m beam size, is tailored by a combined set of high-heat-load slits followed by micrometre-precision slits situated at the front-end 15.5m position. The second set of KB mirrors then focus the beam to the 40m sample position, with a demagnification ratio of similar to 1.5. A detecting system comprising two in-vacuum X-ray pixel detectors is installed to perform synchronized small- and wide-angle X-ray scattering data collections. The observed beamline performance proves the feasibility of having compound features of high flux, microbeam and ultra-SAXS in one beamline.
引用
收藏
页码:1954 / 1965
页数:12
相关论文
共 50 条
  • [1] Design and characterization of an undulator beamline optimized for small-angle coherent X-ray scattering at the Advanced Photon Source
    Sandy, AR
    Lurio, LB
    Mochrie, SGJ
    Malik, A
    Stephenson, GB
    Pelletier, JF
    Sutton, M
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 1999, 6 : 1174 - 1184
  • [2] Design and characterization of an undulator beamline optimized for small-angle coherent X-ray scattering at the advanced photon source
    Sandy, A.R.
    Lurio, L.B.
    Mochrie, S.G.J.
    Malik, A.
    Stephenson, G.B.
    Pelletier, J.F.
    Sutton, M.
    [J]. Journal of Synchrotron Radiation, 1999, 6 (06): : 1174 - 1184
  • [3] Performance of the new biological small- and wide-angle X-ray scattering beamline 13A at the Taiwan Photon Source
    Shih, O.
    Liao, K-F
    Yeh, Y-Q
    Su, C-J
    Wang, C-A
    Chang, J-W
    Wu, W-R
    Liang, C-C
    Lin, C-Y
    Lee, T-H
    Chang, C-H
    Chiang, L-C
    Chang, C-F
    Liu, D-G
    Lee, M-H
    Liu, C-Y
    Hsu, T-W
    Mansel, B.
    Ho, M-C
    Shu, C-Y
    Lee, F.
    Yen, E.
    Lin, T-C
    Jeng, U.
    [J]. JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2022, 55 : 340 - 352
  • [4] Conceptual design of the coherent X-ray scattering beamline at the taiwan photon source
    Huang, Yu-Shan
    Huang, Chi-Yi
    Chang, Chien-Hung
    Peng, Wen-Yan
    Jeng, U-Ser
    Mou, Chung-Yuan
    Lin, Tsang-Lang
    Chen, Hsin-Lung
    [J]. ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2011, 67 : C442 - C442
  • [5] The Coherent X-ray Scattering Beamline at Taiwan Photon Source
    Lin, J.
    Huang, Y.
    Chen, C.
    Jeng, U.
    Mou, C.
    Lin, T.
    Chen, H.
    [J]. ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2014, 70 : C1747 - C1747
  • [6] Small-angle X-ray scattering instruments at the advanced photon source
    Jemian, Pete R.
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2005, 230 : U3536 - U3536
  • [7] Advanced Small- and Wide-angle X-ray Scattering Beamline for Frontier Research in Biological Structures at the Taiwan Photon Source
    Liu, Din-Goa
    Chang, Chien-Hung
    Lee, Ming-Han
    Liu, Chin-Yen
    Chang, Chia-Feng
    Chiang, Liang-Chih
    Hwang, Ching-Shiang
    Huang, Jui-Che
    Sheng, Albert
    Kuan, Chien-Kuang
    Yeh, Yi-Qi
    Su, Chun-Jen
    Liao, Kuei-Fen
    Wu, Wei-Ru
    Shih, Orion
    Jeng, U-Ser
    [J]. 13TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI2018), 2019, 2054
  • [8] First performance assessment of the small-angle X-ray scattering beamline at ELETTRA
    Amenitsch, H
    Rappolt, M
    Kriechbaum, M
    Mio, H
    Laggner, P
    Bernstorff, S
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 1998, 5 : 506 - 508
  • [9] Design of the X-ray Nanoprobe Beamline at the Taiwan Photon Source
    Chang, S. H.
    Yin, G. C.
    Wang, D. J.
    Chang, C. H.
    Juang, J. M.
    Huang, L. J.
    Liu, C. Y.
    Chang, C. F.
    Chu, C. H.
    Tang, M. T.
    [J]. 11TH INTERNATIONAL CONFERENCE ON SYNCHROTRON RADIATION INSTRUMENTATION (SRI 2012), 2013, 425
  • [10] A dedicated small-angle X-ray scattering beamline with a superconducting wiggler source at the NSRRC
    Liu, Din-Goa
    Chang, Chien-Hung
    Liu, Chin-Yen
    Chang, Shih-Hung
    Juang, Jwei-Ming
    Song, Yen-Fang
    Yu, Kuan-Li
    Liao, Kuei-Fen
    Hwang, Ching-Shiang
    Fung, Hok-Sum
    Tseng, Ping-Chung
    Huang, Chi-Yi
    Huang, Liang-Jen
    Chung, Shih-Chun
    Tang, Mau-Tsu
    Tsang, King-Long
    Huang, Yu-Shan
    Kuan, Chien-Kuang
    Liu, Yi-Chih
    Liang, Keng S.
    Jeng, U-Ser
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 2009, 16 : 97 - 104