Non-invasive measurement of pulse pressure variation and systolic pressure variation using a finger cuff corresponds with intra-arterial measurement

被引:39
|
作者
Lansdorp, B. [1 ,2 ]
Ouweneel, D. [1 ]
de Keijzer, A. [1 ]
van der Hoeven, J. G. [2 ]
Lemson, J. [2 ]
Pickkers, P. [2 ]
机构
[1] Univ Twente, MIRA Inst Biomed Technol & Tech Med, NL-7500 AE Enschede, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Intens Care, NL-6500 HB Nijmegen, Netherlands
关键词
blood pressure determination; diagnostic techniques and procedures; fluid therapy; haemodynamics; pulse pressure; systolic pressure; PREDICTING FLUID RESPONSIVENESS; VENA-CAVA DIAMETER; RESPIRATORY CHANGES; ARTERIAL PULSE; VENTILATED PATIENTS; SEPTIC PATIENTS; CRITICALLY-ILL; SURGERY; FINAPRES; NEXFIN;
D O I
10.1093/bja/aer187
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. Pulse pressure variation (PPV) and systolic pressure variation (SPV) are reliable predictors of fluid responsiveness in patients undergoing controlled mechanical ventilation. Currently, PPV and SPV are measured invasively and it is unknown if an arterial pressure (AP) signal obtained with a finger cuff can be used as an alternative. The aim of this study was to validate PPV and SPV measured using a finger cuff. Methods. Patients receiving mechanical ventilation under sedation after cardiac artery bypass graft (CABG) surgery were included after arrival on the intensive care unit. AP was measured invasively in the radial artery and non-invasively using the finger cuff of the Nexfin (TM) monitor. I.V. fluid challenges were administered according to clinical need. The mean value of PPV and SVV was calculated before and after administration of a fluid challenge. Agreement of the calculated PPV and SPV from both methods was assessed using the Bland-Altman analysis. Results. Nineteen patients were included and 28 volume challenges were analysed. Correlation between the two methods for PPV and SPV [mean (SD)=6.9 (4.3)% and 5.3 (2.6)%, respectively] was r=0.96 (P<0.0001) and r=0.95 (P<0.0001), respectively. The mean bias was 20.95% for PPV and -0.22% for SPV. Limits of agreement were -4.3% and 2.4% for PPV and -2.2% and 1.7% for SPV. The correlation between changes in PPV and SPV as a result of volume expansion measured by the two different methods was r=0.88 (P<0.0001) and r=0.87 (P<0.0001), respectively. Conclusions. In patients receiving controlled mechanical ventilation after CABG, PPV and SPV can be measured reliably non-invasively using the inflatable finger cuff of the Nexfin (TM) monitor.
引用
收藏
页码:540 / 545
页数:6
相关论文
共 50 条
  • [1] NONINVASIVE MEASUREMENT OF PULSE AND SYSTOLIC PRESSURE VARIATION USING A FINGER CUFF CORRESPOND WITH INTRA ARTERIAL MEASUREMENTS IN MECHANICALLY VENTILATED PATIENTS
    Lansdorp, B.
    Ouweneel, D.
    Lemson, J.
    Hoeven, J. V. D.
    Pickkers, P.
    [J]. INTENSIVE CARE MEDICINE, 2010, 36 : S96 - S96
  • [2] A Comparison of Non-Invasive Blood Pressure Measurement Strategies with Intra-Arterial Measurement
    Rebesco, Matthew R.
    Pinkston, M. Cornelia
    Smyrnios, Nicholas A.
    Weisberg, Stacy N.
    [J]. PREHOSPITAL AND DISASTER MEDICINE, 2020, 35 (05) : 516 - 523
  • [3] Non-invasive measurement of pulse pressure variation using a finger-cuff method in obese patients having laparoscopic bariatric surgery
    Flick, Moritz
    Schumann, Roman
    Hoppe, Phillip
    Bonney, Iwona
    Wesselink, Wilbert
    Saugel, Bernd
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2021, 35 (06) : 1341 - 1347
  • [4] Non-invasive measurement of pulse pressure variation using a finger-cuff method in obese patients having laparoscopic bariatric surgery
    Moritz Flick
    Roman Schumann
    Phillip Hoppe
    Iwona Bonney
    Wilbert Wesselink
    Bernd Saugel
    [J]. Journal of Clinical Monitoring and Computing, 2021, 35 : 1341 - 1347
  • [5] Non-invasive measurement of pulse pressure variation using a finger-cuff method (CNAP system): a validation study in patients having neurosurgery
    Moritz Flick
    Phillip Hoppe
    Jasmin Matin Mehr
    Luisa Briesenick
    Karim Kouz
    Gillis Greiwe
    Jürgen Fortin
    Bernd Saugel
    [J]. Journal of Clinical Monitoring and Computing, 2022, 36 : 429 - 436
  • [6] Non-invasive measurement of pulse pressure variation using a finger-cuff method (CNAP system): a validation study in patients having neurosurgery
    Flick, Moritz
    Hoppe, Phillip
    Matin Mehr, Jasmin
    Briesenick, Luisa
    Kouz, Karim
    Greiwe, Gillis
    Fortin, Juergen
    Saugel, Bernd
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2022, 36 (02) : 429 - 436
  • [7] Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement
    Harju, Jarkko
    Vehkaoja, Antti
    Kumpulainen, Pekka
    Campadello, Stefano
    Lindroos, Ville
    Yli-Hankala, Arvi
    Oksala, Niku
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2018, 32 (01) : 13 - 22
  • [8] EFFECT OF NON-INVASIVE BLOOD PRESSURE CUFF INFLATION ON INTRA-ARTERIAL BLOOD PRESSURE VALUES
    Sheshadri, V.
    Tiwari, A.
    Nagappa, M.
    Venkatraghavan, L.
    [J]. ANESTHESIA AND ANALGESIA, 2016, 122
  • [9] Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement
    Jarkko Harju
    Antti Vehkaoja
    Pekka Kumpulainen
    Stefano Campadello
    Ville Lindroos
    Arvi Yli-Hankala
    Niku Oksala
    [J]. Journal of Clinical Monitoring and Computing, 2018, 32 : 13 - 22
  • [10] A novel non-invasive cuff-less optoelectronic sensor to measure blood pressure: comparison against intra-arterial measurement
    Lobo, Melvin
    Tucker, Arthur
    Burton, Maudrian
    McDonnell, Andrea
    Shah, Nita
    Shah, Sandeep
    O'Brien, Benjamin
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2019, 33 : 5 - 5